Extracting a Cardiac Signal From the Extracorporeal Pressure Sensors of a Hemodialysis Machine

被引:23
作者
Holmer, Mattias [1 ,2 ]
Sandberg, Frida [3 ,5 ]
Solem, Kristian [6 ]
Grigonyte, Egle [4 ,5 ]
Olde, Bo [6 ]
Sornmo, Leif [3 ,5 ]
机构
[1] Ctr Integrat Electrocardiol CIEL, Dept Biomed Engn, Lund, Sweden
[2] Gambro Lundia AB, SE-22100 Lund, Sweden
[3] Lund Univ, Dept Biomed Engn, Signal Proc Grp, S-22100 Lund, Sweden
[4] Lund Univ, Dept Biomed Engn, S-22100 Lund, Sweden
[5] Lund Univ, Ctr Integrat Electrocardiol CIEL, S-22100 Lund, Sweden
[6] Gambro Lundia AB, Res Dept, SE-22100 Lund, Sweden
基金
瑞典研究理事会;
关键词
Blood access pressure simulator; cardiac pressure signal; extracorporeal pressure sensors; fistula; heart rate; hemodialysis; DIALYSIS;
D O I
10.1109/TBME.2014.2385964
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Although patients undergoing hemodialysis treatment often suffer from cardiovascular disease, monitoring of cardiac rhythm is not performed on a routine basis. Without requiring any extra sensor, this study proposes a method for extracting a cardiac signal from the built-in extracorporeal venous pressure sensor of the hemodialysis machine. The extraction is challenged by the fact that the cardiac component is much weaker than the pressure component caused by the peristaltic blood pump. To further complicate the extraction problem, the cardiac component is difficult to separate when the pump and heart rates coincide. The proposed method estimates a cardiac signal by subtracting an iteratively refined blood pump model signal from the signal measured at the extracorporeal venous pressure sensor. The method was developed based on simulated pressure signals, and evaluated on clinical pressure signals acquired during hemodialysis treatment. The heart rate estimated from the clinical pressure signal was compared to that derived from a photoplethysmographic reference signal, resulting in a difference of 0.07 +/- 0.84 beats/min. The accuracy of the heartbeat occurrence times was studied for different strengths of the cardiac component, using both clinical and simulated signals. The results suggest that the accuracy is sufficient for analysis of heart rate and certain arrhythmias.
引用
收藏
页码:1305 / 1315
页数:11
相关论文
共 25 条
  • [1] Electrocardiographic abnormalities in patients receiving hemodialysis
    Abe, S
    Yoshizawa, M
    Nakanishi, N
    Yazawa, T
    Yokota, K
    Honda, M
    Sloman, G
    [J]. AMERICAN HEART JOURNAL, 1996, 131 (06) : 1137 - 1144
  • [2] Photoplethysmography and its application in clinical physiological measurement
    Allen, John
    [J]. PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) : R1 - R39
  • [3] [Anonymous], 2008, Time Series Analysis, DOI DOI 10.1029/2008WR007288
  • [4] Hemodialysis-induced left ventricular dysfunction is associated with an increase in ventricular arrhythmias
    Burton, James O.
    Korsheed, Shvan
    Grundy, Ben J.
    McIntyre, Christopher W.
    [J]. RENAL FAILURE, 2008, 30 (07) : 701 - 709
  • [5] Time Delay Estimation: A New Insight Into the Woody's Method
    Cabasson, Aline
    Meste, Olivier
    [J]. IEEE SIGNAL PROCESSING LETTERS, 2008, 15 : 573 - 576
  • [6] Automatic Real Time Detection of Atrial Fibrillation
    Dash, S.
    Chon, K. H.
    Lu, S.
    Raeder, E. A.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2009, 37 (09) : 1701 - 1709
  • [7] Hertz T., 2000, U.S. Patent, Patent No. [6 090 048, 6090048]
  • [8] Holmer M, 2013, COMPUT CARDIOL CONF, V40, P217
  • [9] A Novel Method for Detection of the Transition Between Atrial Fibrillation and Sinus Rhythm
    Huang, Chao
    Ye, Shuming
    Chen, Hang
    Li, Dingli
    He, Fangtian
    Tu, Yuewen
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (04) : 1113 - 1119
  • [10] Pulse Rate Variability Analysis for Discrimination of Sleep-Apnea-Related Decreases in the Amplitude Fluctuations of Pulse Photoplethysmographic Signal in Children
    Lazaro, Jesus
    Gil, Eduardo
    Maria Vergara, Jose
    Laguna, Pablo
    [J]. IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2014, 18 (01) : 240 - 246