Motion estimation for cardiac functional analysis using two x-ray computed tomography scans

被引:5
作者
Fung, George S. K. [1 ]
Ciuffo, Luisa [2 ]
Ashikaga, Hiroshi [2 ,3 ]
Taguchi, Katsuyuki [1 ]
机构
[1] Johns Hopkins Univ, Sch Med, Russell H Morgan Dept Radiol & Radiol Sci, Div Med Imaging Phys, Baltimore, MD 21287 USA
[2] Johns Hopkins Univ, Sch Med, Cardiac Arrhythmia Serv, Baltimore, MD 21287 USA
[3] Johns Hopkins Univ, Sch Med, Dept Biomed Engn, Baltimore, MD 21287 USA
关键词
cardiac; cardiac functional analysis; computed tomography; motion estimation; CT;
D O I
10.1002/mp.12425
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This work concerns computed tomography (CT)-based cardiac functional analysis (CFA) with a reduced radiation dose. As CT-CFA requires images over the entire heartbeat, the scans are often performed at 10-20% of the tube current settings that are typically used for coronary CT angiography. A large image noise then degrades the accuracy of motion estimation. Moreover, even if the scan was performed during the sinus rhythm, the cardiac motion observed in CT images may not be cyclic with patients with atrial fibrillation. In this study, we propose to use two CT scan data, one for CT angiography at a quiescent phase at a standard dose and the other for CFA over the entire heart beat at a lower dose. Methods: We have made the following four modifications to an image- based cardiac motion estimation method we have previously developed for a full-dose retrospectively gated coronary CT angiography: (a) a full-dose prospectively gated coronary CT angiography image acquired at the least motion phase was used as the reference image; (b) a three-dimensional median filter was applied to lower-dose retrospectively gated cardiac images acquired at 20 phases over one heartbeat in order to reduce image noise; (c) the strength of the temporal regularization term was made adaptive; and (d) a one-dimensional temporal filter was applied to the estimated motion vector field in order to decrease jaggy motion patterns. We describe the conventional method iME1 and the proposed method iME2 in this article. Five observers assessed the accuracy of the estimated motion vector field of iME2 and iME1 using a 4-point scale. The observers repeated the assessment with data presented in a new random order 1 week after the first assessment session. Results: The study confirmed that the proposed iME2 was robust against the mismatch of noise levels, contrast enhancement levels, and shapes of the chambers. There was a statistically significant difference between iME2 and iME1 (accuracy score, 2.08 +/- 0.81 versus 2.77 +/- 0.98, P < 0.01) and the improvement by the score of + 0.69 seemed clinically relevant. Inter-observer concordance was good: The inter-class correlation coefficient was 0.63 and Kendall's rank correlation coefficients were in the range of 0.41-0.67 (P < 0.01), respectively. Intra-observer reproducibility between sessions was good with the inter-class correlation coefficient of 0.76. Conclusion: We have proposed iME2 method for CT-CFA with two CT scans. The observer study verified the robustness and accuracy of iME2 method and its improved performance over iME1 method. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4677 / 4686
页数:10
相关论文
共 11 条
  • [1] Regional function analysis of left atrial appendage using motion estimation CT and risk of stroke in patients with atrial fibrillation
    Al-Issa, Abdullah
    Inoue, Yuko
    Cammin, Jochen
    Tang, Qiulin
    Nazarian, Saman
    Calkins, Hugh
    Fishman, Elliot K.
    Taguchi, Katsuyuki
    Ashikaga, Hiroshi
    [J]. EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2016, 17 (07) : 788 - 796
  • [2] Quantitative Assessment of Atrial Regional Function Using Motion Estimation Computed Tomography
    Ashikaga, Hiroshi
    Cammin, Jochen
    Tang, Qiulin
    Knudsen, Karly
    Inoue, Yuko
    Fishman, Elliot K.
    Taguchi, Katsuyuki
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2014, 38 (05) : 773 - 778
  • [3] A simple fixed-point approach to invert a deformation field
    Chen, Mingli
    Lu, Weiguo
    Chen, Quan
    Ruchala, Kenneth J.
    Olivera, Gustavo H.
    [J]. MEDICAL PHYSICS, 2008, 35 (01) : 81 - 88
  • [4] Coronary computed tomography and triple rule out CT in patients with acute chest pain and an intermediate cardiac risk profile. Part 1: Impact on patient management
    Gruettner, Joachim
    Fink, Christian
    Walter, Thomas
    Meyer, Mathias
    Apfaltrer, Paul
    Schoepf, U. Joseph
    Saur, Joachim
    Sueselbeck, Tim
    Traunwieser, Dominik
    Takx, Richard
    Kralev, Stefan
    Borggrefe, Martin
    Schoenberg, Stefan O.
    Henzler, Thomas
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (01) : 100 - 105
  • [5] Dual-energy CT angiography of the lungs: Comparison of test bolus and bolus tracking techniques for the determination of scan delay
    Henzler, Thomas
    Meyer, Mathias
    Reichert, Miriam
    Krissak, Radko
    Nance, John W., Jr.
    Haneder, Stefan
    Schoenberg, Stefan O.
    Fink, Christian
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (01) : 132 - 138
  • [6] Peryat JM, 2008, MED IMAGING COMPUTIN
  • [7] A New Method for Cardiac Computed Tomography Regional Function Assessment Stretch Quantifier for Endocardial Engraved Zones (SQUEEZ)
    Pourmorteza, Amir
    Schuleri, Karl H.
    Herzka, Daniel A.
    Lardo, Albert C.
    McVeigh, Elliot R.
    [J]. CIRCULATION-CARDIOVASCULAR IMAGING, 2012, 5 (02) : 243 - 250
  • [8] Rasband W.S., IMAGE J
  • [9] Tang Q, 2012, SPIE MED IMAGING 201, V8314, P8314
  • [10] A fully four-dimensional, iterative motion estimation and compensation method for cardiac CT
    Tang, Qiulin
    Cammin, Jochen
    Srivastava, Somesh
    Taguchi, Katsuyuki
    [J]. MEDICAL PHYSICS, 2012, 39 (07) : 4291 - 4305