Orthopaedic and non-orthopaedic applications of a single-energy iterative metal artefact reduction technique and other metal artefact reduction techniques explained

被引:2
作者
Khor, Robert [1 ]
Buchan, Kevin [2 ]
Kuganesan, Ahilan [1 ]
Ardley, Nicholas [1 ]
Lau, Kenneth K. [3 ]
机构
[1] Monash Hlth, Monash Med Ctr, Dept Diagnost Imaging, Melbourne, Vic, Australia
[2] Philips Healthcare, CT Clin Sci, Macquarie Pk, NSW, Australia
[3] Monash Univ, Fac Med Nursing & Hlth Sci, Clayton, Vic, Australia
关键词
COMPUTED-TOMOGRAPHY; UNITED-STATES; CT; ALGORITHM; MRI; HIP;
D O I
10.1259/bjr.20160171
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Metal within the CT field of view causes artefact that degrades the diagnostic quality of the processed images. This is related to the high atomic number of most metals and is due to a combination of beam hardening, scatter, edge effects and photon starvation. Both software and hardware metal artefact reduction (MAR) techniques have been developed. Iterative reconstruction software MAR techniques can be applied on raw CT data sets and show improved image quality in the setting of sparse projection data when compared with filtered back-projection methods. Recently, a novel single-energy iterative metal artefact reduction technique (IMART) was released for use with large orthopaedic devices. The aim of this pictorial essay was to demonstrate the usefulness of IMART in the setting of both orthopaedic and non-orthopaedic metallic objects and devices.
引用
收藏
页数:7
相关论文
共 22 条
  • [1] Artifacts in CT: Recognition and avoidance
    Barrett, JF
    Keat, N
    [J]. RADIOGRAPHICS, 2004, 24 (06) : 1679 - 1691
  • [2] Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography
    Boas, F. Edward
    Fleischmann, Dominik
    [J]. RADIOLOGY, 2011, 259 (03) : 894 - 902
  • [3] Trends in the incidence and prevalence of cardiac pacemaker insertions in an ageing population
    Bradshaw, Pamela J.
    Stobie, Paul
    Knuiman, Matthew W.
    Briffa, Thomas G.
    Hobbs, Michael S. T.
    [J]. OPEN HEART, 2014, 1 (01):
  • [4] Brown JH, 1999, AM J NEURORADIOL, V20, P694
  • [5] Trends in Permanent Pacemaker Implantation in the United States From 1993 to 2009 Increasing Complexity of Patients and Procedures
    Greenspon, Arnold J.
    Patel, Jasmine D.
    Lau, Edmund
    Ochoa, Jorge A.
    Frisch, Daniel R.
    Ho, Reginald T.
    Pavri, Behzad B.
    Kurtz, Steven M.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (16) : 1540 - 1545
  • [6] Artifact reduction strategies for prosthetic heart valve CT imaging
    Habets, Jesse
    Symersky, Petr
    Leiner, Tim
    de Mol, Bas A. J. M.
    Mali, Willem P. Th M.
    Budde, Ricardo P. J.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (08) : 2099 - 2108
  • [7] Metal-Induced Artifacts in MRI
    Hargreaves, Brian A.
    Worters, Pauline W.
    Pauly, Kim Butts
    Pauly, John M.
    Koch, Kevin M.
    Gold, Garry E.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (03) : 547 - 555
  • [8] Adaptive streak artifact reduction in computed tomography resulting from excessive x-ray photon noise
    Hsieh, J
    [J]. MEDICAL PHYSICS, 1998, 25 (11) : 2139 - 2147
  • [9] An evaluation of three commercially available metal artifact reduction methods for CT imaging
    Huang, Jessie Y.
    Kerns, James R.
    Nute, Jessica L.
    Liu, Xinming
    Balter, Peter A.
    Stingo, Francesco C.
    Followill, David S.
    Mirkovic, Dragan
    Howell, Rebecca M.
    Kry, Stephen F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (03) : 1047 - 1067
  • [10] Usefulness of a Metal Artifact Reduction Algorithm for Orthopedic Implants in Abdominal CT: Phantom and Clinical Study Results
    Jeong, Seonji
    Kim, Se Hyung
    Hwang, Eui Jin
    Shin, Cheong-Il
    Han, Joon Koo
    Choi, Byung Ihn
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (02) : 307 - 317