Objectives: The aim of this study was to determine occupational dose levels in interventional radiology and cardiology procedures. Methods: The study covered a sample of 25 procedures and monitored occupational dose for all laboratory personnel. Each individual wore eight thermoluminescent dosemeters next to the eyes, wrists, fingers and legs during each procedure. Radiation protection shields used in each procedure were recorded. Results: The highest doses per procedure were recorded for interventionists at the left wrist (average 485 mu Sv, maximum 5239 mu Sv) and left finger (average 324 mu Sv, maximum 2877 mu Sv), whereas lower doses were recorded for the legs (average 124 mu Sv, maximum 1959 mu Sv) and the eyes (average 64 mu Sv, maximum 1129 mu Sv). Doses to the assisting nurses during the intervention were considerably lower; the highest doses were recorded at the wrists (average 26 mu Sv, maximum 41 mu Sv) and legs (average 18 mu Sv, maximum 22 mu Sv), whereas doses to the eyes were minimal (average 4 mu Sv, maximum 16 mu Sv). Occupational doses normalised to kerma area product (KAP) ranged from 11.9 to 117.3 mu Sv/1000 cGy cm(2) and KAP was poorly correlated to the interventionists' extremity doses. Conclusion: Calculation of the dose burden for interventionists considering the actual number of procedures performed annually revealed that dose limits for the extremities and the lenses of the eyes were not exceeded. However, there are cases in which high doses have been recorded and this can lead to exceeding the dose limits when bad practices are followed and the radiation protection tools are not properly used.
引用
收藏
页码:70 / 77
页数:8
相关论文
共 50 条
  • [31] Monte Carlo simulation for assessment of occupational doses in pediatric cardiology procedures
    Santos, Carla J.
    Santos, William S.
    Neves, Lucio P.
    Galeano, Diego C.
    Perini, Ana P.
    RADIATION MEASUREMENTS, 2024, 172
  • [32] Occupational radiation procedures and doses in South Korean dentists
    Kim, Yoon-Ji
    Cha, Eun Shil
    Lee, Won Jin
    COMMUNITY DENTISTRY AND ORAL EPIDEMIOLOGY, 2016, 44 (05) : 476 - 484
  • [33] An audit of patient radiation doses during interventional cardiology procedures in Uttarakhand, India, and establishment of local diagnostic reference levels
    Uniyal, Satish C.
    Singh, Vikram
    Rawat, Anurag
    Gururani, Kunal
    Belwal, Chandra Mohan
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (02) : 476 - 487
  • [34] Assessment of the occupational eye lens dose for clinical staff in interventional radiology, cardiology and neuroradiology
    Omar, Artur
    Kadesjo, Nils
    Palmgren, Charlotta
    Marteinsdottir, Maria
    Segerdahl, Tony
    Fransson, Annette
    JOURNAL OF RADIOLOGICAL PROTECTION, 2017, 37 (01) : 145 - 159
  • [35] Occupational radiation dose and protection to the eye lens for interventional cardiology operators
    Lian, Meng-En
    Wu, Guan-Yi
    Chao, Min-Fang
    Wang, Chong-Sheng
    Miao, Hsiao-Chien
    Tsai, Ming-Sung
    Hsieh, Mu-Yang
    Chen, Yeh-Peng
    Chu, Chun-Yuan
    Chang, Szu-Li
    Tsai, Hui-Yu
    RADIATION MEASUREMENTS, 2022, 154
  • [36] An audit of patient radiation doses in interventional radiology at a South African hospital
    Slave, Oneile
    Mahomed, Nasreen
    SA JOURNAL OF RADIOLOGY, 2023, 27 (01):
  • [37] DOSIMETRIC DATA AND RADIATION RISK ANALYSIS FOR NEW PROCEDURES IN INTERVENTIONAL CARDIOLOGY
    Signorotto, P.
    del Vecchio, A.
    Montorfano, M.
    Maisano, F.
    Giagnorio, M.
    Bellanca, R.
    Colombo, A.
    Calandrino, R.
    RADIATION PROTECTION DOSIMETRY, 2010, 142 (2-4) : 201 - 208
  • [38] Monte Carlo study of the scattered radiation field near the eyes of the operator in interventional procedures
    Ferrari, Paolo
    Becker, Frank
    Carinou, Eleftheria
    Chumak, Vadim
    Farah, Jad
    Jovanovic, Zoran
    Krstic, Dragana
    Morgun, Artem
    Principi, Sara
    Teles, Pedro
    JOURNAL OF RADIOLOGICAL PROTECTION, 2016, 36 (04) : 902 - 921
  • [39] Radiation metrics for vascular and interventional radiology procedures in a tertiary care institution
    Arabi, Mohammad
    Rajeh, Ali S.
    Alhendi, Nasser
    Alotaibi, Khalid T.
    Yahyan, Talal A.
    Alyousef, Khalid
    Ardah, Husam
    SAUDI MEDICAL JOURNAL, 2022, 43 (09) : 1035 - 1042
  • [40] Occupational Doses to Medical Staff Performing or Assisting with Fluoroscopically Guided Interventional Procedures
    Borrego, David
    Kitahara, Cari M.
    Balter, Stephen
    Yoder, Craig
    RADIOLOGY, 2020, 294 (02) : 353 - 359