Analyte solvent and injection volume as variables affecting method development in semipreparative reversed-phase liquid chromatography

被引:11
作者
Kozlowski, Edward S. [1 ]
Dalterlo, Richard A. [1 ]
机构
[1] Bristol Myers Squibb Co, Bioanalyt & Discovery Analyt Sci, Wallingford, CT 06492 USA
关键词
analyte solvent; injection volume; reversed phase chromatography; solvent strength; solvent viscosity;
D O I
10.1002/jssc.200700099
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Analyte solvent and injection volume were examined as parameters that affect peak elution during method development for semipreparative RP HPLC purification. Analytical and sernipreparative scale HPLC with gradient elution were used to analyze a mixture of three standard compounds with significantly different retention factors (k). This mixture was analyzed after (i) dissolution in solvents of varied compositions, and (ii) with progressively larger injection volumes. As analyte solvent composition and injection volume were changed, the most notable effects on peak shape were observed for the compounds with the smallest k values. Overall changes in peak shape were less pronounced when analyte solvent composition was similar to the starting mobile phase regardless of injection volume. Scale-up to sernipreparative conditions yielded results consistent with those observed at the analytical scale. These data show that peak shape is greatly affected by analyte solvent composition and injection volume, and that these effects can be ameliorated by the dissolution of analytes in solvent that closely resembles that of the mobile phase used for initial run conditions. The following study addresses the concepts of peak elution in RP HPLC and how they factor into sernipreparative purification.
引用
收藏
页码:2286 / 2292
页数:7
相关论文
共 14 条
[1]   Peak distortion in reversed-phase liquid chromatography as a consequence of viscosity differences between sample solvent and mobile phase [J].
Castells, CB ;
Castells, RC .
JOURNAL OF CHROMATOGRAPHY A, 1998, 805 (1-2) :55-61
[2]   Influence of differences between sample and mobile phase viscosities on the shape of chromatographic elution profiles [J].
Castells, RC ;
Castells, CB ;
Castillo, MA .
JOURNAL OF CHROMATOGRAPHY A, 1997, 775 (1-2) :73-79
[3]   The bandwidth in gradient elution chromatography with a retained organic modifier [J].
Gritti, Fabrice ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1145 (1-2) :67-82
[4]  
HOFFMAN NE, 1991, J LIQ CHROMATOGR, V14, P651, DOI 10.1080/01483919108049277
[5]   INJECTION OF ELUITES IN SOLVENTS STRONGER THAN THE MOBILE PHASE IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
HOFFMAN, NE ;
PAN, SL ;
RUSTUM, AM .
JOURNAL OF CHROMATOGRAPHY, 1989, 465 (02) :189-200
[6]   Effect of sample solvent on the chromatographic peak shape of analytes eluted under reversed-phase liquid chromatogaphic conditions [J].
Keunchkarian, Sonia ;
Reta, Mario ;
Romero, Lilian ;
Castells, Cecilia .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1119 (1-2) :20-28
[7]   Volume-load capacity in fast-gradient liquid chromatography - Effect of sample solvent composition and injection volume on chromatographic performance [J].
Layne, J ;
Farcas, T ;
Rustamov, I ;
Ahmed, F .
JOURNAL OF CHROMATOGRAPHY A, 2001, 913 (1-2) :233-242
[8]   Viscous fingering induced flow instability in multidimensional liquid chromatography [J].
Mayfield, KJ ;
Shalliker, RA ;
Catchpoole, HJ ;
Sweeney, AP ;
Wong, V ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1080 (02) :124-131
[9]   Assessment of injection volume limits when using on-column focusing with microbore liquid chromatography [J].
Mills, MJ ;
Maltas, J ;
Lough, WJ .
JOURNAL OF CHROMATOGRAPHY A, 1997, 759 (1-2) :1-11
[10]   Viscous fingering in packed chromatographic columns: Linear stability analysis [J].
Rousseaux, G. ;
De Wit, A. ;
Martin, M. .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1149 (02) :254-273