Image Quality of CT Angiography of Supra-Aortic Arteries Comparison Between Advanced Modelled Iterative Reconstruction (ADMIRE), Sinogram Affirmed Iterative Reconstruction (SAFIRE) and Filtered Back Projection (FBP) in One Patients' Group

被引:16
作者
Oergel, Anja [1 ]
Bier, Georg [1 ]
Hennersdorf, Florian [1 ]
Richter, Hardy [2 ]
Ernemann, Ulrike [1 ]
Hauser, Till-Karsten [1 ]
机构
[1] Univ Hosp Tuebingen, Dept Diagnost & Intervent Neuroradiol, Hoppe Seyler Str 3, D-72076 Tubingen, Germany
[2] Univ Hosp Tuebingen, Dept Neurol Focus Neurovasc Dis & Neurooncol, Hoppe Seyler Str 3, D-72076 Tubingen, Germany
关键词
Diagnostic accuracy; Radiation dose reduction; Image processing; Iterative reconstruction; Signal-to-noise ratio; RADIATION-DOSE REDUCTION; COMPUTED-TOMOGRAPHY; PERFORMANCE; ALGORITHMS;
D O I
10.1007/s00062-018-0740-y
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose To evaluate the potential benefit in image quality of the iterative reconstruction (IR) technique advanced modelled iterative reconstruction (ADMIRE) in CT angiography (CTA) of supra-aortic arteries compared to sinogram affirmed iterative reconstruction (SAFIRE) and standard filtered back projection (FBP) in one patients' group. Methods In this study 29 patients underwent standard CTA of supra-aortic arteries. Images were reconstructed using three different reconstruction algorithms, FBP, and IR techniques ADMIRE and SAFIRE. General image quality was assessed by two radiologists in different arterial segments using a 5-point Likert scale. Mean attenuation and noise were measured at different levels of each vessel and signal-to-noise ratio (SNR) was calculated. Interrater variability was determined. Results Regarding total image quality IR showed only excellent, very good and good results and was more often graded excellent and very good than FBP reconstruction. Image noise levels and SNR were significantly (p < 0.01) improved in IR at the bilateral subclavian arteries (only in SAFIRE), vertebral V1 and V3 segments, common carotid arteries and proximal and distal internal carotid arteries. No significant differences in image quality were found when comparing SAFIRE and ADMIRE reconstructions except for V1 and V3. In these regions ADMIRE showed significantly better image quality than SAFIRE (p < 0.001 and p < 0.01). Interrater agreement was almost perfect (kappa = 0.88) for different image quality parameters. Conclusion The CTA of supra-aortic arteries using the IR techniques SAFIRE and ADMIRE was superior to FBP when comparing the objective and subjective image quality and ADMIRE has the potential to overcome SAFIRE.
引用
收藏
页码:101 / 107
页数:7
相关论文
共 16 条
[1]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[2]   Potential of combining iterative reconstruction with noise efficient detector design: aggressive dose reduction in head CT [J].
Brodoefel, H. ;
Bender, B. ;
Schabel, C. ;
Fenchel, M. ;
Ernemann, U. ;
Korn, A. .
BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1050)
[3]   Dose reduction with iterative reconstruction for coronary CT angiography: a systematic review and meta-analysis [J].
Den Harder, Annemarie M. ;
Willemink, Martin J. ;
De Ruiter, Quirina M. B. ;
De Jong, Pim A. ;
Schilham, Arnold M. R. ;
Krestin, Gabriel P. ;
Leiner, Tim ;
Budde, Ricardo P. J. .
BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1058)
[4]   Performance evaluation of iterative reconstruction algorithms for achieving CT radiation dose reduction - a phantom study [J].
Dodge, Cristina T. ;
Tamm, Eric P. ;
Cody, Dianna D. ;
Liu, Xinming ;
Jensen, Corey T. ;
Wei, Wei ;
Kundra, Vikas ;
Rong, X. John .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02) :511-531
[5]   A Novel Pairwise Comparison-Based Method to Determine Radiation Dose Reduction Potentials of Iterative Reconstruction Algorithms, Exemplified Through Circle of Willis Computed Tomography Angiography [J].
Ellmann, Stephan ;
Kammerer, Ferdinand ;
Brand, Michael ;
Allmendinger, Thomas ;
May, Matthias S. ;
Uder, Michael ;
Lell, Michael M. ;
Kramer, Manuel .
INVESTIGATIVE RADIOLOGY, 2016, 51 (05) :331-339
[6]   Review of radiation issues for computed tomography [J].
Frush, DP .
SEMINARS IN ULTRASOUND CT AND MRI, 2004, 25 (01) :17-24
[7]   Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography [J].
Gordic, Sonja ;
Desbiolles, Lotus ;
Sedlmair, Martin ;
Manka, Robert ;
Plass, Andre ;
Schmidt, Bernhard ;
Husarik, Daniela B. ;
Maisano, Francesco ;
Wildermuth, Simon ;
Alkadhi, Hatem ;
Leschka, Sebastian .
EUROPEAN RADIOLOGY, 2016, 26 (02) :459-468
[8]   Sinogram affirmed iterative reconstruction in head CT: Improvement of objective and subjective image quality with concomitant radiation dose reduction [J].
Korn, A. ;
Bender, B. ;
Fenchel, M. ;
Spira, D. ;
Schabel, C. ;
Thomas, C. ;
Flohr, T. ;
Claussen, C. D. ;
Bhadelia, R. ;
Ernemann, U. ;
Brodoefel, H. .
EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (09) :1431-1435
[9]   High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose [J].
Korn, A. ;
Fenchel, M. ;
Bender, B. ;
Danz, S. ;
Thomas, C. ;
Ketelsen, D. ;
Claussen, C. D. ;
Moonis, G. ;
Krauss, B. ;
Heuschmid, M. ;
Ernemann, U. ;
Brodoefel, H. .
NEURORADIOLOGY, 2013, 55 (04) :423-430
[10]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174