Accuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes

被引:98
作者
Garg, Satish K. [1 ]
Kipnes, Mark [2 ]
Castorino, Kristin [3 ]
Bailey, Timothy S. [4 ]
Akturk, Halis Kaan [1 ]
Welsh, John B. [5 ]
Christiansen, Mark P. [6 ]
Balo, Andrew K. [5 ]
Brown, Sue A. [7 ]
Reid, Jennifer L. [5 ]
Beck, Stayce E. [5 ]
机构
[1] Univ Colorado, Barbara Davis Ctr Diabet, Dept Med & Pediat, Aurora, CO USA
[2] DGD Clin, San Antonio, TX USA
[3] Sansum Diabet Res Inst, Santa Barbara, CA USA
[4] AMCR Inst, Escondido, CA USA
[5] Dexcom Inc, San Diego, CA USA
[6] Diablo Clin Res, Walnut Creek, CA USA
[7] Univ Virginia, Charlottesville, VA USA
关键词
Continuous glucose monitoring; Dexcom; Accuracy; G7; MARD; PERFORMANCE; SYSTEM;
D O I
10.1089/dia.2022.0011
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: We evaluated the accuracy and safety of a seventh generation (G7) Dexcom continuous glucose monitor (CGM) during 10.5 days of use in adults with diabetes.Methods: Adults with either type 1 or type 2 diabetes (on intensive insulin therapy or not) participated at 12 investigational sites in the United States. In-clinic visits were conducted on days 1 or 2, 4 or 7, and on the second half of day 10 or the first half of day 11 for frequent comparisons with comparator blood glucose measurements obtained with the YSI 2300 Stat Plus glucose analyzer. Participants wore sensors concurrently on the upper arm and abdomen. Accuracy evaluation included the proportion of CGM values within 15% of comparator glucose levels >100 mg/dL or within 15 mg/dL of comparator levels <= 100 mg/dL (%15/15), along with the %20/20 and %30/30 agreement rates. The mean absolute relative difference (MARD) between temporally matched CGM and comparator values was also calculated.Results: Data from 316 participants (619 sensors, 77,774 matched pairs) were analyzed. For arm- and abdomen-placed sensors, overall MARDs were 8.2% and 9.1%, respectively. Overall %15/15, %20/20, and %30/30 agreement rates were 89.6%, 95.3%, and 98.8% for arm-placed sensors and were 85.5%, 93.2%, and 98.1% for abdomen-placed sensors. Across days of wear, glucose concentration ranges, and rates of change, %20/20 agreement rates varied by no more than 9% from the overall %20/20. No serious adverse events were reported.Conclusions: The G7 CGM provides accurate glucose readings with single-digit MARD with arm or abdomen placement in adults with diabetes. Clinicaltrials.gov: NCT04794478
引用
收藏
页码:373 / 380
页数:8
相关论文
共 21 条
[1]  
Abbott, ART44417001
[2]   New Features and Performance of a Next-Generation SEVEN-Day Continuous Glucose Monitoring System with Short Lag Time [J].
Bailey, Timothy ;
Zisser, Howard ;
Chang, Anna .
DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 (12) :749-755
[3]   Effect of Continuous Glucose Monitoring on Glycemic Control in Adults With Type 1 Diabetes Using Insulin Injections The DIAMOND Randomized Clinical Trial [J].
Beck, Roy W. ;
Riddlesworth, Tonya ;
Ruedy, Katrina ;
Ahmann, Andrew ;
Bergenstal, Richard ;
Haller, Stacie ;
Kollman, Craig ;
Kruger, Davida ;
McGill, Janet B. ;
Polonsky, William ;
Toschi, Elena ;
Wolpert, Howard ;
Price, David .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2017, 317 (04) :371-378
[4]  
Calhoun P., 2018, J Diabetes Sci Technol, V13, P308
[5]   Performance of the Dexcom G6 Continuous Glucose Monitoring System in Pregnant Women with Diabetes [J].
Castorino, Kristin ;
Polsky, Sarit ;
O'Malley, Grenye ;
Levister, Camilla ;
Nelson, Kristen ;
Farfan, Christian ;
Brackett, Scott ;
Puhr, Sarah ;
Levy, Carol J. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2020, 22 (12) :943-947
[6]  
dexcom, Dexcom G6 Continuous Glucose Monitoring System User Guide
[7]   State of Type 1 Diabetes Management and Outcomes from the T1D Exchange in 2016-2018 [J].
Foster, Nicole C. ;
Beck, Roy W. ;
Miller, Kellee M. ;
Clements, Mark A. ;
Rickels, Michael R. ;
DiMeglio, Linda A. ;
Maahs, David M. ;
Tamborlane, William V. ;
Bergenstal, Richard ;
Smith, Elizabeth ;
Olson, Beth A. ;
Garg, Satish K. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2019, 21 (02) :66-72
[8]  
Freckmann Guido, 2019, J Diabetes Sci Technol, V13, P890, DOI [10.1177/1932296819826965, 10.1177/1932296819826965]
[9]   Accuracy of the Dexcom G6 Glucose Sensor during Aerobic, Resistance, and Interval Exercise in Adults with Type 1 Diabetes [J].
Guillot, Florian H. ;
Jacobs, Peter G. ;
Wilson, Leah M. ;
Youssef, Joseph El ;
Gabo, Virginia B. ;
Branigan, Deborah L. ;
Tyler, Nichole S. ;
Ramsey, Katrina ;
Riddell, Michael C. ;
Castle, Jessica R. .
BIOSENSORS-BASEL, 2020, 10 (10)
[10]   Association of Real-time Continuous Glucose Monitoring With Glycemic Control and Acute Metabolic Events Among Patients With Insulin-Treated Diabetes [J].
Karter, Andrew J. ;
Parker, Melissa M. ;
Moffet, Howard H. ;
Gilliam, Lisa K. ;
Dlott, Richard .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2021, 325 (22) :2273-2284