Magnetic Resonance Driven Electrical Impedance Tomography: A Simulation Study

被引:4
|
作者
Negishi, Michiro [1 ]
Tong, Tangji
Constable, R. Todd [1 ,2 ]
机构
[1] Yale Univ, Sch Med, Dept Diagnost Radiol, Dept Biomed Engn, New Haven, CT 06520 USA
[2] Yale Univ, Sch Med, Dept Neurosurg, New Haven, CT 06520 USA
关键词
Conductivity image; current density imaging (CDI); electrical impedance tomography (EIT); magnetic resonance electrical impedance tomography (MREIT); CONDUCTIVITY; MREIT;
D O I
10.1109/TMI.2010.2098035
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Magnetic resonance electrical impedance tomography (MREIT) is a method for reconstructing a three-dimensional image of the conductivity distribution in a target volume using magnetic resonance (MR). In MREIT, currents are applied to the volume through surface electrodes and their effects on the MR induced magnetic fields are analyzed to produce the conductance image. However, current injection through surface electrodes poses technical problems such as the limitation on the safely applicable currents. In this paper, we present a new method called magnetic resonance driven electrical impedance tomography (MRDEIT), where the magnetic resonance in each voxel is used as the applied magnetic field source, and the resultant electromagnetic field is measured through surface electrodes or radio-frequency (RF) detectors placed near the surface. Because the applied magnetic field is at the RF frequency and eddy currents are the integral components in the method, a vector wave equation for the electric field is used as the basis of the analysis instead of a quasi-static approximation. Using computer simulations, it is shown that complex permittivity images can be reconstructed using MRDEIT, but that improvements in signal detection are necessary for detecting moderate complex permittivity changes.
引用
收藏
页码:828 / 837
页数:10
相关论文
共 50 条
  • [1] A feasibility study of magnetic resonance driven electrical impedance tomography using a phantom
    Wan, Yuqing
    Negishi, Michiro
    Constable, R. Todd
    PHYSIOLOGICAL MEASUREMENT, 2013, 34 (06) : 623 - 644
  • [2] Magnetic resonance electrical impedance tomography of the breast: A simulation study on basic imaging setup
    Lee, B. I.
    Oh, S. H.
    Kim, T. -S.
    Woo, E. J.
    Lee, S. Y.
    Kwon, O.
    Seo, J. K.
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 4377 - 4380
  • [3] Magnetic resonance electrical impedance tomography (MREIT):: Simulation study of J-substitution algorithm
    Kwon, O
    Woo, EJ
    Yoon, JR
    Seo, JK
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (02) : 160 - 167
  • [4] Multi-frequency Electrical Impedance Tomography and Magnetic Resonance Electrical Impedance Tomography
    Seo, Jin Keun
    Woo, Eung Je
    MATHEMATICAL MODELING IN BIOMEDICAL IMAGING I: ELECTRICAL AND ULTRASOUND TOMOGRAPHIES, ANOMALY DETECTION, AND BRAIN IMAGING, 2009, 1983 : 1 - 71
  • [5] A Simulation Study on Induced Current Magnetic Resonance Electrical Impedance Tomography of Anisotropic Brain Tissues
    Liu, Yang
    Dai, Songshi
    2010 3RD INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2010), VOLS 1-7, 2010, : 75 - 77
  • [6] Magnetic Resonance Electrical Impedance Tomography (MREIT)
    Seo, Jin Keun
    Woo, Eung Je
    SIAM REVIEW, 2011, 53 (01) : 40 - 68
  • [7] Fast imaging for magnetic resonance electrical impedance tomography
    Hamamura, Mark J.
    Muftuler, L. Tugan
    MAGNETIC RESONANCE IMAGING, 2008, 26 (06) : 739 - 745
  • [8] On Denoising Technology in Magnetic Resonance Electrical Impedance Tomography
    Xie, Jiaye
    Liu, Jijun
    PROCEEDINGS OF FIRST INTERNATIONAL CONFERENCE OF MODELLING AND SIMULATION, VOL III: MODELLING AND SIMULATION IN ELECTRONICS, COMPUTING, AND BIO-MEDICINE, 2008, : 346 - 351
  • [9] Coregistration of electrical impedance tomography and magnetic resonance imaging
    Halter, R. J.
    Manwaring, P.
    Hartov, A.
    Paulsen, K. D.
    13TH INTERNATIONAL CONFERENCE ON ELECTRICAL BIOIMPEDANCE AND THE 8TH CONFERENCE ON ELECTRICAL IMPEDANCE TOMOGRAPHY 2007, 2007, 17 : 416 - 419
  • [10] A feasibility study of magnetic resonance electrical impedance tomography for prostate cancer detection
    Liu, Yang
    Zhang, Yingchun
    PHYSIOLOGICAL MEASUREMENT, 2014, 35 (04) : 567 - 581