Comparison of five automated hematology analyzers in a university hospital setting: Abbott Cell-Dyn Sapphire, Beckman Coulter DxH 800, Siemens Advia 2120i, Sysmex XE-5000, and Sysmex XN-2000

被引:65
作者
Bruegel, Mathias [1 ]
Nagel, Dorothea [1 ]
Funk, Manuela [1 ]
Fuhrmann, Petra [1 ]
Zander, Johannes [1 ]
Teupser, Daniel [1 ]
机构
[1] Univ Munich, Inst Lab Med, D-81377 Munich, Germany
关键词
complete blood count; differential; flagging quality; hematology analyzer; inter-instrument comparison; RED-BLOOD-CELLS; PERFORMANCE EVALUATION; SPECIMENS; COUNTS; FLAGS;
D O I
10.1515/cclm-2014-0945
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Various types of automated hematology analyzers are used in clinical laboratories. Here, we performed a side-by-side comparison of five current top of the range routine hematology analyzers in the setting of a university hospital central laboratory. Methods: Complete blood counts (CBC), differentials, reticulocyte and nucleated red blood cell (NRBC) counts of 349 patient samples, randomly taken out of routine diagnostics, were analyzed with Cell-Dyn Sapphire (Abbott), DxH 800 (Beckman Coulter), Advia 2120i (Siemens), XE-5000 and XN-2000 (Sysmex). Inter-instrument comparison of CBCs including reticulocyte and NRBC counts and investigation of flagging quality in relation to microscopy were performed with the complete set of samples. Inter-instrument comparison of five-part differential was performed using samples without atypical cells in blood smear (n = 292). Automated five-part differentials and NRBCs were additionally compared with microscopy. Results: The five analyzers showed a good concordance for basic blood count parameters. Correlations between instruments were less well for reticulocyte counts, NRBCs, and differentials. The poorest concordance for NRBCs with microscopy was observed for Advia 2120i (Kendall's tau(b) = 0.37). The highest flagging sensitivity for blasts was observed for XN-2000 (97% compared to 65%-76% for other analyzers), whereas overall specificity was comparable between different instruments. Conclusions: To the best of our knowledge, this is the most comprehensive side-by-side comparison of five current top of the range routine hematology analyzers. Variable analyzer quality and parameter specific limitations must be considered in defining laboratory algorithms in clinical practice.
引用
收藏
页码:1057 / 1071
页数:15
相关论文
共 26 条
[1]  
[Anonymous], CONS RUL CONS GUID P
[2]  
[Anonymous], 2010, H20A2 CLSI
[3]  
Barnes P W, 2010, Lab Hematol, V16, P23, DOI 10.1532/LH96.09015
[4]   Nucleated red blood cell counts in the first week of life: a critical appraisal of relationships with perinatal outcome in preterm growth-restricted neonates [J].
Baschat, Ahmet A. ;
Gungor, Sadettin ;
Kush, Michelle L. ;
Berg, Christoph ;
Gembruch, Ulrich ;
Harman, Christopher R. .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2007, 197 (03) :286.e1-286.e8
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting [J].
Briggs, C. ;
Culp, N. ;
Davis, B. ;
d'Onofrio, G. ;
Zini, G. ;
Machin, S. J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 (06) :613-627
[7]   Performance evaluation of the Sysmex haematology XN modular system [J].
Briggs, Carol ;
Longair, Ian ;
Kumar, Punamar ;
Singh, Deepak ;
Machin, Samuel J. .
JOURNAL OF CLINICAL PATHOLOGY, 2012, 65 (11) :1024-1030
[8]   Automated blood cell counts - State of the art [J].
Buttarello, Mauro ;
Plebani, Mario .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (01) :104-116
[9]   The Usefulness of Blast Flags on the Sysmex XE-5000 Is Questionable [J].
Eilertsen, Heidi ;
Vollestad, Nina K. ;
Hagve, Tor-Arne .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2013, 139 (05) :633-640
[10]   Evaluation of the monocyte counting by two automated haematology analysers compared with flow cytometry [J].
Grimaldi, E ;
Carandente, P ;
Scopacasa, F ;
Romano, MF ;
Pellegrino, M ;
Bisogni, R ;
De Caterina, M .
CLINICAL AND LABORATORY HAEMATOLOGY, 2005, 27 (02) :91-97