A Novel Indirect Reduction Technique in Ankle Syndesmotic Injuries: A Cadaveric Study

被引:18
作者
Cosgrove, Christopher T. [1 ]
Spraggs-Hughes, Amanda G. [1 ]
Putnam, Sara M. [1 ]
Ricci, William M. [1 ]
Miller, Anna N. [1 ]
McAndrew, Christopher M. [1 ]
Gardner, Michael J. [2 ]
机构
[1] Washington Univ, Sch Med, Dept Orthopaed Surg, St Louis, MO USA
[2] Stanford Univ, Dept Orthopaed Surg, Sch Med, Redwood City, CA USA
关键词
syndesmosis; cadaver study; computed tomography; TIBIOFIBULAR SYNDESMOSIS; RADIOGRAPHIC EVALUATION; SCREW FIXATION; MALREDUCTION; CT; FRACTURES; ACCURACY; CLAMP; LINE;
D O I
10.1097/BOT.0000000000001169
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: To describe a novel technique using preoperative computed tomography (CT) to plan clamp tine placement along the trans-syndesmotic axis (TSA). We hypothesized that preoperative CT imaging provides a reliable template on which to plan optimal clamp tine positioning along the TSA, reducing malreduction rates compared with other described techniques. Methods: CT images of 48 cadaveric through-knee specimens were obtained, and the TSA was measured as well as the optimal position of the medial clamp tine. The syndesmosis was then fully destabilized. Indirect clamp reductions were performed with the medial clamp tine placed at positions 10 degrees anterior to the TSA, along the TSA, and at both 10 and 20 degrees posterior to the TSA. The specimens were then separately reduced using manual digital pressure and palpation alone. CT was performed after each clamp and manual reduction. Results: On average, reduction clamp tines were within 3 +/- 2 degrees of the desired angle and within 5% +/- 4% of the templated location along the tibial line for all clamp reduction attempts. Palpation and direct visualization produced the overall lowest malreduction rates in all measurements: 4.9% and 3.0%, respectively. Off-axis clamping 10 degrees anterior or 20 degrees posterior to the patient-specific TSA demonstrated an increased overall malreduction rate: 15.8% and 11.3%, respectively. Significantly more over-compression occurred when a reduction clamp was used versus manual digital reduction alone (8.6% vs. 0%). Conclusions: Reduction clamp placement directly along an optimal clamping vector can be facilitated by preoperative CT measurements of the uninjured ankle. However, even in this setting, the use of reduction clamps increases the risk for syndesmotic malreduction and over-compression compared with manual digital reduction or direct visualization.
引用
收藏
页码:361 / 367
页数:7
相关论文
共 44 条
[1]   30-YEAR FOLLOW-UP OF ANKLE FRACTURES [J].
BAUER, M ;
JONSSON, K ;
NILSSON, B .
ACTA ORTHOPAEDICA SCANDINAVICA, 1985, 56 (02) :103-106
[2]   Radiographic measurement of the distal tibiofibular syndesmosis has limited use [J].
Beamer, A ;
van Hemert, WLW ;
Niesing, R ;
Entius, CAC ;
Ginai, AZ ;
Mulder, PGH ;
Swierstra, BA .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (423) :227-234
[3]   Instability of the tibio-fibular syndesmosis: have we been pulling in the wrong direction? [J].
Candal-Couto, JJ ;
Burrow, D ;
Bromage, S ;
Briggs, PJ .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2004, 35 (08) :814-818
[4]   Incisura Morphology as a Risk Factor for Syndesmotic Malreduction [J].
Cherney, Steven M. ;
Spraggs-Hughes, Amanda G. ;
McAndrew, Christopher M. ;
Ricci, William M. ;
Gardner, Michael J. .
FOOT & ANKLE INTERNATIONAL, 2016, 37 (07) :748-754
[5]   In Vivo Syndesmotic Overcompression After Fixation of Ankle Fractures With a Syndesmotic Injury [J].
Cherney, Steven M. ;
Haynes, Jacob A. ;
Spraggs-Hughes, Amanda G. ;
McAndrew, Christopher M. ;
Ricci, William M. ;
Gardner, Michael J. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (09) :414-419
[6]   Medial Clamp Tine Positioning Affects Ankle Syndesmosis Malreduction [J].
Cosgrove, Christopher T. ;
Putnam, Sara M. ;
Cherney, Steven M. ;
Ricci, William M. ;
Spraggs-Hughes, Amanda ;
McAndrew, Christopher M. ;
Gardner, Michael J. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (08) :440-446
[7]   Intraoperative Syndesmotic Reduction: Three-Dimensional Versus Standard Fluoroscopic Imaging [J].
Davidovitch, Roy I. ;
Weil, Yoram ;
Karia, Raj ;
Forman, Jordanna ;
Looze, Christopher ;
Liebergall, Meir ;
Egol, Kenneth .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (20) :1838-1843
[8]   Normal Tibiofibular Relationships at the Syndesmosis on Axial CT Imaging [J].
Dikos, Gregory D. ;
Heisler, Jason ;
Choplin, Robert H. ;
Weber, Timothy G. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (07) :433-438
[9]   Syndesmosis reduction by computer-assisted orthopaedic surgery with navigation: Feasibility and accuracy in a cadaveric study [J].
Dubois-Ferriere, Victor ;
Gamulin, Axel ;
Chowdhary, Ashwin ;
Fasel, Jean ;
Stern, Richard ;
Assal, Mathieu .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (12) :2694-2699
[10]   Radiographic and CT evaluation of tibiofibular syndesmotic diastasis: A cadaver study [J].
Ebraheim, NA ;
Lu, JK ;
Yang, H ;
Mekhail, AO ;
Yeasting, RA .
FOOT & ANKLE INTERNATIONAL, 1997, 18 (11) :693-698