Hyperpolarized 3Helium gas for functional magnetic resonance imaging of the lung

被引:6
作者
Heussel, CP
Gast, KK
Dahmen, A
Morbach, AE
Kauczor, HU
Schreiber, WG
Thelen, M
van Beek, EJR
机构
[1] Johannes Gutenberg Univ Mainz, Klin & Poliklin Radiol, D-55131 Mainz, Germany
[2] Deutsch Krebsforschungszentrum, D-6900 Heidelberg, Germany
[3] Univ Iowa Hosp & Clin, Carver Coll Med, Dept Radiol, Iowa City, IA 52242 USA
关键词
lung diseases; pulmonary function; hyperpolarized helium-3; magnetic resonance imaging (MRI);
D O I
10.1007/s00063-005-1048-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Lung imaging is traditionally done using X-ray-based methods, since MRI is limited by low proton density as well as inherent magnetic field inhomogeneities of the lung tissue. After introduction of MRI using hyperpolarized noble gases, a totally new field of MRI of the chest has rapidly evolved. These techniques reveal new functional information of the lungs, which could not be obtained before. The first part of this review describes the underlying MR technology explaining distribution of static ventilation, dynamic distribution of ventilation lung microstructure (apparent diffusion coefficient [ADC]), measurement of oxygen partial pressure (pO(2)), and safety. The clinical potential is afterwards demonstrated in the second part. Therefore, the effort in normal lungs and the mainly focused diseases chronic obstructive pulmonary disease (COPD), smoker's lung, cystic fibrosis, asthma, lung transplantation, and pulmonary embolism are reported.
引用
收藏
页码:413 / 424
页数:12
相关论文
共 110 条
[1]  
ALTES T, 2003, P 11 ANN M ISMRM TOR
[2]   Hyperpolarized 3He MR lung ventilation imaging in asthmatics:: Preliminary findings [J].
Altes, TA ;
Powers, PL ;
Knight-Scott, J ;
Rakes, G ;
Platts-Mills, TAE ;
de Lange, EE ;
Alford, BA ;
Mugler, JP ;
Brookeman, JR .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (03) :378-384
[3]   STUDY OF MECHANICAL COMPRESSION OF SPIN-POLARIZED HE-3 GAS [J].
BECKER, J ;
HEIL, W ;
KRUG, B ;
LEDUC, M ;
MEYERHOFF, M ;
NACHER, PJ ;
OTTEN, EW ;
PROKSCHA, T ;
SCHEARER, LD ;
SURKAU, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 346 (1-2) :45-51
[4]   CYSTIC-FIBROSIS - SCORING SYSTEM WITH THIN-SECTION CT [J].
BHALLA, M ;
TURCIOS, N ;
APONTE, V ;
JENKINS, M ;
LEITMAN, BS ;
MCCAULEY, DI ;
NAIDICH, DP .
RADIOLOGY, 1991, 179 (03) :783-788
[5]   In vivo NMR of hyperpolarized 3He in the human lung at very low magnetic fields [J].
Bidinosti, CP ;
Choukeife, J ;
Nacher, PJ ;
Tastevin, G .
JOURNAL OF MAGNETIC RESONANCE, 2003, 162 (01) :122-132
[6]   In vivo He-3 MR images of guinea pig lungs [J].
Black, RD ;
Middleton, HL ;
Cates, GD ;
Cofer, GP ;
Driehuys, B ;
Happer, W ;
Hedlund, LW ;
Johnson, GA ;
Shattuck, MD ;
Swartz, JC .
RADIOLOGY, 1996, 199 (03) :867-870
[7]   Quantification of atelectases in artificial respiration: Spiral-CT versus dynamic single-slice CT [J].
Bletz, C ;
Markstaller, K ;
Karmrodt, J ;
Herweling, A ;
Melvan, M ;
Goetz, R ;
Stepniak, A ;
Eberle, B ;
Kauczor, HU ;
Heussel, CP ;
Thelen, M .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (03) :409-416
[8]   NUCLEAR POLARIZATION IN HE-3 GAS INDUCED BY OPTICAL PUMPING AND DIPOLAR EXCHANGE [J].
BOUCHIAT, MA ;
CARVER, TR ;
VARNUM, CM .
PHYSICAL REVIEW LETTERS, 1960, 5 (08) :373-375
[9]   STEADY-STATE PRODUCTION OF HIGH NUCLEAR-POLARIZATION IN HE-3-HE-4 MIXTURES [J].
CANDELA, D ;
HAYDEN, ME ;
NACHER, PJ .
PHYSICAL REVIEW LETTERS, 1994, 73 (19) :2587-2590
[10]  
Chen XJ, 1999, MAGNET RESON MED, V42, P721, DOI 10.1002/(SICI)1522-2594(199910)42:4<721::AID-MRM14>3.0.CO