Clinical experience with a shape memory polymer peripheral vascular embolisation plug: a case series

被引:14
作者
Morgan, Robert A. [1 ]
Loftus, Ian [2 ]
Ratnam, Lakshmi [1 ]
Das, Raj [1 ]
Mailli, Leto [3 ]
Hamady, Mohamad S. [4 ]
Lobotesis, Kyriakos [5 ]
机构
[1] St Georges Univ Hosp NHS Fdn Trust, Dept Radiol, London, England
[2] St Georges Healthcare NHS Trust, Vasc Inst, London, England
[3] Guys & St Thomas NHS Fdn Trust, Intervent Radiol, London, England
[4] Imperial Coll London, Intervent Radiol, St Marys Campus, London, England
[5] Charing Cross Hosp, Neuroradiol, London, England
关键词
Shape memory polymer; Peripheral embolisation; Vascular plug;
D O I
10.1186/s42155-021-00214-w
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundShape memory polymers are materials that are manufactured in a certain shape, can be stored in a temporary deformed shape, and then return to - or remember - their original shape upon exposure to external stimuli such as temperature and moisture. This property lends itself to application in endovascular medical devices. Peripheral vasculature embolisation devices incorporating this novel technology have become commercially available and this case series, where the data were collected as part of a post market registry, outlines initial clinical experience with these novel devices.ResultsEight cases are described in this series. The disease state/conditions for which embolisation was indicated were right common iliac artery aneurysms (n =3), a type II endoleak into the thoracic aorta following thoracic endovascular aneurysm repair (n =1), a left inferior gluteal artery aneurysm (n =1), left internal iliac artery aneurysms (n =2), and a case of splenomegaly, where splenectomy was planned after the embolisation procedure (n =1). Target arteries were 5-10mm in diameter. In each case, at least one IMPEDE Embolization Plug (IMP-Device) of an appropriate diameter was used. All procedures were technically successful and target vessel thrombosis was achieved in all cases. Follow-up imaging available during the 45-90-day data collection timeframe showed sustained vessel occlusion. This case series includes examples of situations commonly encountered when embolising the peripheral vasculature, namely, the use of one or multiple devices in a single vessel and in combination with the use of other embolic devices (e.g., microcoils, gelatin sponge, and PVA particles) in the same case. There were no adverse events related to the specific use of the device.ConclusionsThis small series illustrates the safety and efficacy of this novel sponge-based embolic device for the embolisation of small and medium sized arteries and further experience will demonstrate the utility of the shape memory polymer devices.
引用
收藏
页数:9
相关论文
共 2 条
[1]   Microscopic Assessment of Healing and Effectiveness of a Foam-Based Peripheral Occlusion Device [J].
Jessen, Staci L. ;
Friedemann, Molly C. ;
Ginn-Hedman, Anne-Marie ;
Graul, Lance M. ;
Jokerst, Steven ;
Robinson, Cedric B. ;
Landsman, Todd L. ;
Clubb, Fred J., Jr. ;
Maitland, Duncan J. .
ACS BIOMATERIALS SCIENCE & ENGINEERING, 2020, 6 (05) :2588-2599
[2]   Design and verification of a shape memory polymer peripheral occlusion device [J].
Landsman, Todd L. ;
Bush, Ruth L. ;
Glowczwski, Alan ;
Horn, John ;
Jessen, Staci L. ;
Ungchusri, Ethan ;
Diguette, Katelin ;
Smith, Harrison R. ;
Hasan, Sayyeda M. ;
Nash, Daniel ;
Clubb, Fred J., Jr. ;
Maitland, Duncan J. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 63 :195-206