Geometric Optimization of an Extracorporeal Centrifugal Blood Pump with an Unshrouded Impeller Concerning Both Hydraulic Performance and Shear Stress

被引:5
|
作者
Huang, Bo [1 ]
Guo, Miao [1 ]
Lu, Bin [2 ]
Wu, Qingyu [3 ]
Zuo, Zhigang [1 ]
Liu, Shuhong [1 ]
机构
[1] Tsinghua Univ, Dept Energy & Power Engn, State Key Lab Hydro Sci & Engn, Beijing 100084, Peoples R China
[2] Zhejiang Qingke Pumping Med Technol Ltd, Taizhou 318010, Peoples R China
[3] Tsinghua Univ, Sch Clin Med, Beijing 100084, Peoples R China
基金
中国国家自然科学基金; 国家重点研发计划;
关键词
centrifugal blood pump; geometric optimization; hydraulic performance; maximum scalar shear stress; VENTRICULAR ASSIST DEVICE; FLUID-DYNAMICS; HEMOLYSIS; FLOW; DAMAGE; DESIGN; MODEL; TERM;
D O I
10.3390/pr9071211
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Centrifugal blood pumps have provided a powerful artificial support system for patients with vascular diseases. In the design process, geometrical optimization is usually needed to acquire a more biocompatible model for clinical uses. In the current paper, we propose a method for multi-objective optimization concerning both the hydraulic and the hemolytic performances of the pump based on the near-orthogonal array in which the traditional hemolysis index (HI) is replaced with the maximum scalar shear stress criteria to reduce the computation load. The method is demonstrated with the optimization of an extracorporeal centrifugal blood pump with an unshrouded impeller. CFD studies on the original and nine modified pump models are carried out. The calculated hydraulic performances of the optimized model are also compared against the experiments for validation of the numeric method, with an error of 3.6% at the original design point. The resulting blood pump with low maximum scalar shear stress (132.2 Pa) shows a low degree of calculated HI (1.69 x 10(-3)).
引用
收藏
页数:15
相关论文
共 25 条
  • [1] THE EFFECT OF SHEAR STRESS ON THE IMPELLER DESIGN OF CENTRIFUGAL BLOOD PUMP
    Mondee, Kamonwan
    Foojinphan, Nutchanant
    Wannawat, Praemai
    Naiyanetr, Phornphop
    2016 9TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON), 2016,
  • [2] Hydraulic performance and physiological compatibility of an extracorporeal centrifugal blood pump during startup
    Geng, Da
    Han, Huan
    Huang, Bo
    Liu, Shuhong
    Zuo, Zhigang
    PHYSICS OF FLUIDS, 2025, 37 (03)
  • [3] Numerical investigation on the effect of impeller axial position on hemodynamics of an extracorporeal centrifugal blood pump
    Lv, Shen
    He, Zhi-Peng
    Liu, Guang-Mao
    Hu, Sheng-Shou
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2024, 27 (13) : 1744 - 1755
  • [4] Hydraulic Analysis and Optimization in impeller of Magnetically Levitated Centrifugal Blood Pump
    Zhang, Huimin
    Zhu, Qi
    FRONTIERS OF MANUFACTURING AND DESIGN SCIENCE II, PTS 1-6, 2012, 121-126 : 1204 - 1208
  • [5] Impeller (straight blade) design variations and their influence on the performance of a centrifugal blood pump
    Fang, Peng
    Du, Jianjun
    Yu, Shunzhou
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2020, 43 (12): : 782 - 795
  • [6] Hydraulic Design Optimization and Performance Analysis of a Centrifugal Blood Pump
    Park, Moo Ryong
    Yoo, Seong Yeon
    Oh, Hyoung Woo
    Yoon, Eui Soo
    TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS B, 2006, 30 (01) : 87 - 94
  • [7] Improvement of Internal Flow Performance of a Centrifugal Pump-As-Turbine (PAT) by Impeller Geometric Optimization
    Xu, Jian
    Wang, Longyan
    Asomani, Stephen Ntiri
    Luo, Wei
    Lu, Rong
    MATHEMATICS, 2020, 8 (10) : 1 - 23
  • [8] Investigation of flow pattern and hydraulic performance of a centrifugal pump impeller through the PIV method
    Li, Xiaojun
    Chen, Hui
    Chen, Bo
    Luo, Xianwu
    Yang, Baofeng
    Zhu, Zuchao
    RENEWABLE ENERGY, 2020, 162 : 561 - 574
  • [9] Evaluation of hydraulic radial forces on the impeller by the volute in a centrifugal rotary blood pump
    Boehning F.
    Timms D.L.
    Amaral F.
    Oliveira L.
    Graefe R.
    Hsu P.-L.
    Schmitz-Rode T.
    Steinseifer U.
    Artificial Organs, 2011, 35 (08) : 818 - 825
  • [10] Optimization of a centrifugal blood pump in terms of hemolysis index and hydraulic efficiency
    Incebay, Omer
    Yapici, Rafet
    JOURNAL OF THE BRAZILIAN SOCIETY OF MECHANICAL SCIENCES AND ENGINEERING, 2024, 46 (10)