Library-based scatter correction for dedicated cone beam breast CT: a feasibility study

被引:3
|
作者
Shi, Linxi [1 ,2 ]
Vedantham, Srinivasan [3 ]
Karellas, Andrew [3 ]
Zhu, Lei [1 ,2 ]
机构
[1] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Nucl & Radiol Engn Program, Atlanta, GA 30332 USA
[2] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Med Phys Program, Atlanta, GA 30332 USA
[3] Univ Massachusetts, Sch Med, Dept Radiol, Worcester, MA USA
来源
MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING | 2016年 / 9783卷
关键词
cone-beam breast CT; scatter correction; scatter library; Monte Carlo simulation; X-RAY SCATTER; COMPUTED-TOMOGRAPHY;
D O I
10.1117/12.2217327
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Purpose: Scatter errors are detrimental to cone-beam breast CT (CBBCT) accuracy and obscure the visibility of calcifications and soft-tissue lesions. In this work, we propose practical yet effective scatter correction for CBBCT using a library-based method and investigate its feasibility via small-group patient studies. Method: Based on a simplified breast model with varying breast sizes, we generate a scatter library using Monte-Carlo (MC) simulation. Breasts are approximated as semi-ellipsoids with homogeneous glandular/adipose tissue mixture. On each patient CBBCT projection dataset, an initial estimate of scatter distribution is selected from the pre-computed scatter library by measuring the corresponding breast size on raw projections and the glandular fraction on a first-pass CBBCT reconstruction. Then the selected scatter distribution is modified by estimating the spatial translation of the breast between MC simulation and the clinical scan. Scatter correction is finally performed by subtracting the estimated scatter from raw projections. Results: On two sets of clinical patient CBBCT data with different breast sizes, the proposed method effectively reduces cupping artifact and improves the image contrast by an average factor of 2, with an efficient processing time of 200ms per cone beam projection. Conclusion: Compared with existing scatter correction approaches on CBBCT, the proposed library-based method is clinically advantageous in that it requires no additional scans or hardware modifications. As the MC simulations are pre-computed, our method achieves a high computational efficiency on each patient dataset The library-based method has shown great promise as a practical tool for effective scatter correction on clinical CBBCT.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Library-Based X-Ray Scatter Correction for Dedicated Cone-Beam Breast CT: Clinical Validation
    Shi, L.
    Vedantham, S.
    Karellas, A.
    Zhu, L.
    MEDICAL PHYSICS, 2016, 43 (06) : 3819 - 3819
  • [2] Library based x-ray scatter correction for dedicated cone beam breast CT
    Shi, Linxi
    Vedantham, Srinivasan
    Karellas, Andrew
    Zhu, Lei
    MEDICAL PHYSICS, 2016, 43 (08) : 4529 - 4544
  • [3] Scatter Correction for Dedicated Cone Beam Breast CT Based On a Forward Projection Model
    Shi, L.
    Vedantham, S.
    Karellas, A.
    Zhu, L.
    MEDICAL PHYSICS, 2016, 43 (06) : 3820 - 3820
  • [4] The role of off-focus radiation in scatter correction for dedicated cone beam breast CT
    Shi, Linxi
    Vedantham, Srinivasan
    Karellas, Andrew
    Zhu, Lei
    MEDICAL PHYSICS, 2018, 45 (01) : 191 - 201
  • [5] Scatter correction for clinical cone beam CT breast imaging based on breast phantom studies
    Cai, Weixing
    Ning, Ruola
    Conover, David
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2011, 19 (01) : 91 - 109
  • [6] Dedicated cone-beam breast CT: Feasibility study with surgical mastectomy specimens
    Yang, Wei Tse
    Carkaci, Selin
    Chen, Lingyun
    Lai, Chao-Jen
    Sahin, Aysegul
    Whitman, Gary J.
    Shaw, Chris C.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (06) : 1312 - 1315
  • [7] X-ray scatter correction for dedicated cone beam breast CT using a forward-projection model
    Shi, Linxi
    Vedantham, Srinivasan
    Karellas, Andrew
    Zhu, Lei
    MEDICAL PHYSICS, 2017, 44 (06) : 2312 - 2320
  • [8] Feasibility of Improving Cone Beam CT Number Consistency with a Scatter-Correction Algorithm
    Li, J.
    Yao, W.
    Xiao, Y.
    Yu, Y.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [9] A model-based scatter artifacts correction for cone beam CT
    Zhao, Wei
    Vernekohl, Don
    Zhu, Jun
    Wang, Luyao
    Xing, Lei
    MEDICAL PHYSICS, 2016, 43 (04) : 1736 - 1753
  • [10] Correction of scatter in megavoltage cone-beam CT
    Spies, L
    Ebert, M
    Groh, BA
    Hesse, BM
    Bortfeld, T
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03): : 821 - 833