A novel energy mapping approach for CT-based attenuation correction in PET

被引:11
作者
Teimourian, B. [1 ,2 ]
Ay, M. R. [1 ,3 ,4 ]
Zafarghandi, M. Shamsaie [2 ]
Ghafarian, P. [5 ,6 ]
Ghadiri, H. [1 ,4 ]
Zaidi, H. [7 ,8 ,9 ]
机构
[1] Univ Tehran Med Sci, Res Ctr Sci & Technol Med, Tehran, Iran
[2] Amirkabir Univ Technol, Tehran Polytech, Fac Phys & Nucl Engn, Tehran, Iran
[3] Univ Tehran Med Sci, Res Inst Nucl Med, Tehran, Iran
[4] Univ Tehran Med Sci, Dept Med Phys & Biomed Engn, Tehran, Iran
[5] Shahid Beheshti Univ Med Sci, Telemed Res Ctr, NRITLD, Masih Daneshvari Hosp, Tehran, Iran
[6] Shahid Beheshti Univ Med Sci, Chron Resp Dis Res Ctr, NRITLD, Masih Daneshvari Hosp, Tehran, Iran
[7] Univ Hosp Geneva, Div Nucl Med & Mol Imaging, CH-1211 Geneva, Switzerland
[8] Univ Geneva, Geneva Neurosci Ctr, CH-1211 Geneva, Switzerland
[9] Univ Groningen, Univ Med Ctr Gronigen, Dept Nucl Med & Mol Imaging, NL-9700 RB Groningen, Netherlands
基金
瑞士国家科学基金会;
关键词
PET/CT; attenuation correction; attenuation map; energy-mapping; dual-energy CT; ARTIFACTS; IMAGES; TOMOGRAPHY;
D O I
10.1118/1.3694108
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Dual-energy CT (DECT) is arguably the most accurate energy mapping technique in CT-based attenuation correction (CTAC) implemented on hybrid PET/CT systems. However, this approach is not attractive for clinical use owing to increased patient dose. The authors propose a novel energy mapping approach referred to as virtual DECT (VDECT) taking advantage of the DECT formulation but using CT data acquired at a single energy (kV(P)). For this purpose, the CT image acquired at one energy is used to generate the CT image at a second energy using calculated kV(P) conversion curves derived from phantom studies. Methods: The attenuation map (mu-map) at 511 keV was generated for the XCAT phantom and clinical studies using the bilinear, DECT, and VDECT techniques. The generated mu-maps at 511 keV are compared to the reference derived from the XCAT phantom serving as ground truth. PET data generated from a predefined activity map for the XCAT phantom were then corrected for attenuation using mu-maps generated using the different energy mapping approaches. In addition, the generated mu-maps using the above described methods for a cylindrical polyethylene phantom containing different concentrations of K2HPO4 in water were compared to actual attenuation coefficients. Likewise, CT images of five clinical whole-body studies were used to generate mu-maps using the various energy-mapping approaches were compared with mu-maps acquired at 511 keV using Ge-68/Ga-68 rod sources for the clinical studies. Results: The results of phantom studies demonstrate that the proposed method is more accurate than the bilinear technique. All three mu-maps yielded almost similar results for soft and lung tissues whereas for bone tissues, the DECT and the VDECT methods produced a much smaller mean relative difference (3.0% and 2.8%, respectively) than the bilinear approach (11.8%). Likewise, the comparison of PET images corrected for attenuation using the various methods showed that the proposed method provides better accuracy (6.5%) than the bilinear method (13.4%). Clinical studies further demonstrated that, compared to the bilinear method, the VDECT approach has better agreement for bony structures with the DECT technique (1.5% versus 8.9%) and transmission scanning (8.8% versus 17.7%). Conclusions: It was concluded that the proposed method outperforms the bilinear method especially in bony structures. Further evaluation using a large clinical PET/CT database is underway to evaluate the potential of the technique in a clinical setting. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.3694108]
引用
收藏
页码:2078 / 2089
页数:12
相关论文
共 50 条
  • [41] Clinical Utility of CT-based Attenuation-correction in Myocardial Perfusion SPECT Imaging
    Hatipoglu, Filiz
    Cetin, Neslihan
    MOLECULAR IMAGING AND RADIONUCLIDE THERAPY, 2023, 32 (02) : 138 - 145
  • [42] A novel method for incorporating respiratory-matched attenuation correction in the motion correction of cardiac PET-CT studies
    McQuaid, Sarah J.
    Lambrou, Tryphon
    Hutton, Brian F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (10) : 2903 - 2915
  • [43] MR-based attenuation correction for cardiac FDG PET on a hybrid PET/MRI scanner: comparison with standard CT attenuation correction
    Vontobel, Jan
    Liga, Riccardo
    Possner, Mathias
    Clerc, Olivier F.
    Mikulicic, Fran
    Veit-Haibach, Patrick
    ter Voert, Edwin E. G. W.
    Fuchs, Tobias A.
    Stehli, Julia
    Pazhenkottil, Aju P.
    Benz, Dominik C.
    Graeni, Christoph
    Gaemperli, Oliver
    Herzog, Bernhard
    Buechel, Ronny R.
    Kaufmann, Philipp A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (10) : 1574 - 1580
  • [44] What is the benefit of CT-based attenuation correction in myocardial perfusion SPET?
    Apostolopoulos, Dimitrios J.
    Savvopoulos, Christos
    HELLENIC JOURNAL OF NUCLEAR MEDICINE, 2016, 19 (02): : 89 - 92
  • [45] Interpolated average CT for cardiac PET/CT attenuation correction
    Greta S. P. Mok
    Cobie Y. T. Ho
    Bang-Hung Yang
    Tung-Hsin Wu
    Journal of Nuclear Cardiology, 2016, 23 : 1072 - 1079
  • [46] Impact of X-ray tube settings and metallic leads on neurological PET imaging when using CT-based attenuation correction
    Ay, Mohammad Reza
    Zaidi, Habib
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) : 411 - 414
  • [47] CT-Based Attenuation Correction in I-123-Ioflupane SPECT
    Lange, Catharina
    Seese, Anita
    Schwarzenboeck, Sarah
    Steinhoff, Karen
    Umland-Seidler, Bert
    Krause, Bernd J.
    Brenner, Winfried
    Sabri, Osama
    Kurth, Jens
    Hesse, Swen
    Buchert, Ralph
    PLOS ONE, 2014, 9 (09):
  • [48] Low dose interpolated average CT for thoracic PET/CT attenuation correction using an active breathing controller
    Sun, Tao
    Wu, Tung-Hsin
    Wang, Shyh-Jen
    Yang, Bang-Hung
    Wu, Nien-Yun
    Mok, Greta S. P.
    MEDICAL PHYSICS, 2013, 40 (10)
  • [49] Impact of CT-Based and MRI-Based Attenuation Correction Methods on 18 F-FDG PET Quantification Using PET Phantoms
    Wu, Cheng-Han
    Tu, Chun-Yuan
    Shen, Li-Chen
    Lu, Yueh-Hsun
    Yang, Bang-Hung
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2022, 42 (03) : 374 - 381
  • [50] Interpolated average CT for PET attenuation correction in different lesion characteristics
    Ho, Cobie Y. T.
    Wu, Tung-Hsin
    Mok, Greta S. P.
    NUCLEAR MEDICINE COMMUNICATIONS, 2016, 37 (03) : 297 - 306