Regional deconvolution method for partial volume correction in brain PET

被引:0
|
作者
Rusinek, H [1 ]
Tsuia, WH [1 ]
de Leon, MJ [1 ]
机构
[1] Bellevue Hosp Ctr, Dept Radiol, New York, NY 10016 USA
关键词
functional imaging; brain metabolism; MRI; PET; partial volume correction;
D O I
10.1117/12.428166
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Correction of PET images for partial volume effects (PVE) is of particular utility in studies of metabolism in brain aging and brain disorders, PVE is commonly corrected using voxel-by-voxel factors obtained from a high resolution brain mask (obtained from the coregistered MR scan), after convolution with the point spread function (PSF) of the imaging system. In a recently proposed regional deconvolution (RD) method, the observed regional activity is expressed as linear combinations of the true metabolic activity. The weights are obtained by integrating the PSF over the geometric extent of the brain regions. We have analyzed the accuracy of RD and two other PVE correction algorithms under a variety of conditions using simulated PET scans. Each of the brain regions was assigned a distribution of metabolic activity, with gray matter/white matter contrast representative of subjects in several age categories. Simulations were performed over a wide range of PET resolutions. The influence of PET/MR misregistration and heterogeneity of brain metabolism were also evaluated. Our results demonstrate the importance of correcting PET metabolic images for PVE. Without such correction, the regional brain activity values are contaminated with 30-40% errors. Under most conditions studied, the accuracy of RD and of the three-compartmental method were superior to the accuracy of the two-compartmental method. Our study provides the first demonstration of the feasibility of RD algorithm to provide accurate correction for a large number (n=109) of brain compartments. PVE correction methods appear to be promising tools in studies of metabolism in normal brain, brain aging, and brain disorders.
引用
收藏
页码:409 / 418
页数:10
相关论文
共 50 条
  • [1] PARTIAL VOLUME CORRECTION OF BRAIN PET STUDIES USING ITERATIVE DECONVOLUTION IN COMBINATION WITH HYPR DENOISING
    Golla, S. V. S.
    Lubberink, M.
    Lammertsma, A. A.
    Boellaard, R.
    JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2016, 36 : 690 - 691
  • [2] Partial volume correction of brain PET studies using iterative deconvolution in combination with HYPR denoising
    Golla, Sandeep S. V.
    Lubberink, Mark
    van Berckel, Bart N. M.
    Lammertsma, Adriaan A.
    Boellaard, Ronald
    EJNMMI RESEARCH, 2017, 7
  • [3] Partial volume correction of brain PET studies using iterative deconvolution in combination with HYPR denoising
    Sandeep S. V. Golla
    Mark Lubberink
    Bart N. M. van Berckel
    Adriaan A. Lammertsma
    Ronald Boellaard
    EJNMMI Research, 7
  • [4] A Monte Carlo study of deconvolution algorithms for partial volume correction in quantitative PET
    Tohka, Jussi
    Reilhac, Anthonin
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 3339 - 3345
  • [5] Performance evaluation of PMOD integrated partial volume correction method for brain PET data
    Barret, Olivier
    Mularczyk, Pawel
    Mikolajczyk, Krzysztof
    Burger, Cyrill
    NEUROIMAGE, 2010, 52 : S10 - S10
  • [6] Partial volume correction in brain PET-FDG studies
    Prieto, E.
    Marti-Climent, J. M.
    Gomez-Fernandez, M.
    Lecumberri, P.
    Bustince, H.
    Arbizu, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 : S148 - S148
  • [7] Importance of partial-volume correction in brain PET studies
    Fazio, F
    Perani, D
    JOURNAL OF NUCLEAR MEDICINE, 2000, 41 (11) : 1849 - 1850
  • [8] The effects of iterative deconvolution partial-volume correction on beta-amyloid PET
    Thomas, B.
    Thurfjell, L.
    Rinne, J.
    Ourselin, S.
    Hutton, B.
    JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2009, 29 : S337 - S338
  • [9] Partial volume effect correction in PET image using iterative deconvolution and shearlet transform
    Jomaa, Hajer
    Mabrouk, Rostom
    Khlifa, Nawres
    2018 4TH INTERNATIONAL CONFERENCE ON ADVANCED TECHNOLOGIES FOR SIGNAL AND IMAGE PROCESSING (ATSIP), 2018,
  • [10] Analysis of partial volume correction on quantification and regional heterogeneity in cardiac PET
    Turco, A.
    Nuyts, J.
    Duchenne, J.
    Gheysens, O.
    Voigt, J. U.
    Claus, P.
    Vunckx, K.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2020, 27 (01) : 62 - 70