Experimental method to determine the absolute efficiency curve of a wavelength dispersive spectrometer

被引:17
|
作者
Trincavelli, Jorge [1 ,2 ]
Limandri, Silvina [1 ,2 ]
Carreras, Alejo [2 ,3 ]
Bonetto, Rita [2 ,4 ,5 ]
机构
[1] Ciudad Univ, Fac Matemat Astron & Fis, Univ Nacl Cordoba, RA-5000 Cordoba, Argentina
[2] Ciudad Univ, Consejo Nacl Invest Cient & Tecn Republ Argentina, RA-5000 Cordoba, Argentina
[3] Univ Nacl San Luis, Inst Invest & Tecnol Quim, RA-5700 San Luis, Argentina
[4] UNLP, Ctr Invest & Desarrollo Ciencias Aplicadas Dr Jor, Fac Ciencias Exactas, RA-1900 La Plata, Argentina
[5] UNLP, Fac Ingn, RA-1900 La Plata, Argentina
关键词
detection efficiency; wavelength dispersive X-ray spectrometer; electron probe microanalysis; standardless microanalysis;
D O I
10.1017/S1431927608080379
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A method for the experimental determination of the absolute efficiency of wavelength dispersive spectrometers was developed, based on the comparison of spectra measured with a wavelength dispersive system and with an energy dispersive spectrometer. The aim of studying this parameter arises because its knowledge is necessary to perform standardless analysis. A simple analytical expression was obtained for the efficiency curve for three crystals (TAP, PET, and LiF) of the spectrometer used, within an energy range from 0.77 to 10.83 keV. Although this expression is particular for the system used in this work, the method may be extended to other spectrometers and crystals for electron probe microanalysis and X-ray fluorescence.
引用
收藏
页码:306 / 314
页数:9
相关论文
共 6 条
  • [1] Enhancing the efficiency of a wavelength-dispersive spectrometer based on a slitless design using a single-bounce monocapillary
    Bzheumikhova, Karina
    Kayser, Y.
    Unterumsberger, R.
    Weser, J.
    Stadelhoff, C.
    Beckhoff, B.
    JOURNAL OF SYNCHROTRON RADIATION, 2025, 32 : 171 - 179
  • [2] Calibration of wavelength-dispersive X-ray spectrometer for standardless analysis
    Sitko, R
    Zawisza, B
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2005, 60 (01) : 95 - 100
  • [3] Research of Passive Efficiency Calibration Method and Software Development Based on HPGe γ Spectrometer
    Zhou Chunlin
    Shao Yong
    Lei Junniu
    Li Sufen
    Zhang Jiaoyu
    Xu Zhenhua
    PROCEEDINGS OF THE THIRD INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOLS 1 - 4, 2010, : 874 - 881
  • [4] Experimental characterization of alpha spectrometer for optimization of operational parameters affecting energy resolution and detection efficiency
    Ryu, Jeonghyeon
    Park, Ji-young
    Lee, Hyeon-Woo
    Kim, Hyuncheol
    Chae, Jung-Seok
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2021, 329 (02) : 959 - 967
  • [5] Experimental characterization of alpha spectrometer for optimization of operational parameters affecting energy resolution and detection efficiency
    Jeonghyeon Ryu
    Ji-young Park
    Hyeon-Woo Lee
    Hyuncheol Kim
    Jung-Seok Chae
    Journal of Radioanalytical and Nuclear Chemistry, 2021, 329 : 959 - 967
  • [6] Experimental validation of absolute full energy peak efficiency and energy resolution of NaI(Tl), CsI(Tl), BGO, YAP(Ce) and CeBr3 scintillation detectors modeled with Monte Carlo codes
    Mouhti, Issam
    Mcfee, John Elton
    El-Bouzaidi, Mohamed Drissi
    El Qars, J.
    Ouacha, El Houssaine
    Assakrar, M'hand
    Bellioua, Mohammed
    RADIATION PHYSICS AND CHEMISTRY, 2025, 231