Determination of chromium in whole blood by DRC-ICP-MS:: spectral and non-spectral interferences

被引:25
作者
Bonnefoy, C
Menudier, A
Moesch, C
Lachâtre, G
Mermet, JM
机构
[1] Univ Lyon 1, Analyt Sci Lab, F-69622 Villeurbanne, France
[2] Hop Univ Dupuytren, Dept Pharmacol Toxicol, F-87042 Limoges, France
关键词
inductively coupled plasma-mass spectrometry; dynamic reaction cell; chromium; whole blood; standard additions; matrix effects;
D O I
10.1007/s00216-005-3403-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Quantification of chromium in whole blood has been performed by ICP-quadrupole MS. The spectrometer was equipped with a dynamic reaction cell (DRC) with ammonia as reaction gas. The rejection parameter q (RPq) of the DRC and the flow rate of ammonia (NH3) were optimized and set at 0.7 and 0.6 mL min(-1), respectively. Blood was diluted 1:51 (v/v) with an aqueous solution containing 0.1 mg L-1 NH4OH, 0.1 g L-1 EDTA, 5 mg L-1 n-butanol, and 0.1 (0)/(00). Triton X100. Non-spectral matrix effects observed when using the DRC were confirmed by use of vanadium. External calibration with blank and standard Solutions prepared in purified water led to biased results for quality control samples. Standard addition calibration was therefore used and its validity verified. By comparing the slopes and calculating residues, it was proved that the plot obtained with standard additions and the plot obtained from blood samples of different concentrations were aligned down to 0.05 mu g L-1 after dilution.
引用
收藏
页码:167 / 173
页数:7
相关论文
共 10 条
  • [1] Validation of the determination of lead in whole blood by ICP-MS
    Bonnefoy, C
    Menudier, A
    Moesch, C
    Lachâtre, G
    Mermet, JM
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (09) : 1161 - 1165
  • [2] THE ASSESSMENT OF REFERENCE VALUES FOR ELEMENTS IN HUMAN BIOLOGICAL TISSUES AND FLUIDS - A SYSTEMATIC REVIEW
    CAROLI, S
    ALIMONTI, A
    CONI, E
    PETRUCCI, F
    SENOFONTE, O
    VIOLANTE, N
    [J]. CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1994, 24 (5-6) : 363 - 398
  • [3] Case CP, 2001, CLIN CHEM, V47, P275
  • [4] Determination of chromium in water and urine by reaction cell inductively coupled plasma mass spectrometry
    Chang, YL
    Jiang, SJ
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (12) : 1434 - 1438
  • [5] LAUWERYS RR, 2001, IND CHEM EXPOSURE GU, P77
  • [6] TRACE-ELEMENT REFERENCE VALUES IN TISSUES FROM INHABITANTS OF THE EUROPEAN COMMUNITY .1. A STUDY OF 46 ELEMENTS IN URINE, BLOOD AND SERUM OF ITALIAN SUBJECTS
    MINOIA, C
    SABBIONI, E
    APOSTOLI, P
    PIETRA, R
    POZZOLI, L
    GALLORINI, M
    NICOLAOU, G
    ALESSIO, L
    CAPODAGLIO, E
    [J]. SCIENCE OF THE TOTAL ENVIRONMENT, 1990, 95 : 89 - 105
  • [7] Evaluation of a tunable bandpass reaction cell for an inductively coupled plasma mass spectrometer for the determination of chromium and vanadium in serum and urine
    Nixon, DE
    Neubauer, KR
    Eckdahl, SJ
    Butz, JA
    Burritt, MF
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (05) : 951 - 966
  • [8] STEILER HG, 1994, HDB METALS CLIN ANAL, P753
  • [9] Tanner SD, 1999, ATOM SPECTROSC, V20, P45
  • [10] Reaction cells and collision cells for ICP-MS: a tutorial review
    Tanner, SD
    Baranov, VI
    Bandura, DR
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (09) : 1361 - 1452