Development and testing of a portable vocal accumulator

被引:102
作者
Cheyne, HA
Hanson, HM
Genereux, RP
Stevens, KN
Hillman, RE
机构
[1] Massachusetts Eye & Ear Infirm, Voice & Speech Lab, Boston, MA 02114 USA
[2] MIT, Cambridge, MA 02139 USA
[3] Sensimetr Corp, Somerville, NJ USA
来源
JOURNAL OF SPEECH LANGUAGE AND HEARING RESEARCH | 2003年 / 46卷 / 06期
关键词
ambulatory monitoring; voice assessment; vocal function; voice disorders;
D O I
10.1044/1092-4388(2003/113)
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
This research note describes the design and testing of a device for unobtrusive, long-term ambulatory monitoring of voice use, named the Portable Vocal Accumulator (PVA). The PVA contains a digital signal processor for analyzing input from a neck-placed miniature accelerometer. During its development, accelerometer recordings were obtained from 99 participants with normal or dysphonic voices. The recordings were used to (a) test the specifications and capabilities of the PVA for monitoring normal and dysphonic voices and (b) explore potentially useful displays for the large quantity of data generated by long-term monitoring. The current prototype PVA is pocket-sized (12 x 8.5 x 2 cm), lightweight (200 g), and capable of sampling 11 hr of voice-use data, including estimates of fundamental frequency, sound pressure level, and phonation duration.
引用
收藏
页码:1457 / 1467
页数:11
相关论文
共 10 条
[1]   VOCAL LOAD AS MEASURED BY THE VOICE ACCUMULATOR [J].
BUEKERS, R ;
BIERENS, E ;
KINGMA, H ;
MARRES, EHMA .
FOLIA PHONIATRICA ET LOGOPAEDICA, 1995, 47 (05) :252-261
[2]  
Fairbanks G, 1960, Voice and articulation drillbook, V2nd
[3]   TREATMENT OF VOCAL ABUSE DISORDERS USING A VOCAL INTENSITY CONTROLLER [J].
HOLBROOK, A ;
ROLNICK, MI ;
BAILEY, CW .
JOURNAL OF SPEECH AND HEARING DISORDERS, 1974, 39 (03) :298-303
[4]   ANALYSIS OF VOCAL ABUSE - FLUCTUATIONS IN PHONATION TIME AND INTENSITY IN 4 GROUPS OF SPEAKERS [J].
MASUDA, T ;
IKEDA, Y ;
MANAKO, H ;
KOMIYAMA, S .
ACTA OTO-LARYNGOLOGICA, 1993, 113 (04) :547-552
[5]   SIMPLE BIOFEEDBACK DEVICE TO REDUCE EXCESSIVE VOCAL INTENSITY [J].
MCGILLIVRAY, R ;
PROCTORWILLIAMS, K ;
MCLISTER, B .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1994, 32 (03) :348-350
[6]   A VOICE ACCUMULATOR-VALIDATION AND APPLICATION [J].
OHLSSON, AC ;
BRINK, O ;
LOFQVIST, A .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1989, 32 (02) :451-457
[7]   A NEWLY DEVISED SPEECH ACCUMULATOR [J].
RYU, S ;
KOMIYAMA, S ;
KANNAE, S ;
WATANABE, H .
ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY AND ITS RELATED SPECIALTIES, 1983, 45 (02) :108-114
[8]  
Stevens Kenneth N., 1998, Acoustic phonetics
[9]  
Szabo A, 2001, Logoped Phoniatr Vocol, V26, P102
[10]   A PORTABLE MICROPROCESSOR-BASED BIOFEEDBACK TRAINING DEVICE [J].
ZICKER, JE ;
TOMPKINS, WJ ;
RUBOW, RT ;
ABBS, JH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (09) :509-515