In Vitro Identification of Four-Element Windkessel Models Based on Iterated Unscented Kalman Filter

被引:22
作者
Huang, Huan [1 ]
Yang, Ming [1 ]
Zang, Wangfu [2 ]
Wu, Shunjie [1 ]
Pang, Yafei [1 ]
机构
[1] Shanghai Jiao Tong Univ, Dept Instrument Sci & Engn, Shanghai 200240, Peoples R China
[2] Shanghai Jiao Tong Univ, Dept Cardiac Surg, Ruijin Hosp, Sch Med, Shanghai 200240, Peoples R China
基金
中国国家自然科学基金;
关键词
Mock circulatory loop (MCL); system identification; total arterial inertance; unscented Kalman filter (UKF); Windkessel model; MOCK CIRCULATION; ARTERIAL; IMPEDANCE; LOOP; BED;
D O I
10.1109/TBME.2011.2161477
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mock circulatory loops (MCLs) have been widely used to test left ventricular assist devices. The hydraulic properties of the mock systemic arterial system are usually described by two alternative four-element windkessel (W4) models. Compared with three-element windkessel model, their parameters, especially the inertial term, are much more difficult to estimate. In this paper, an estimator based on the iterated unscented Kalman filter (IUKF) algorithm is proposed to identify model parameters. Identifiability of these parameters for different measurements is described. Performance of the estimator for different model structures is first evaluated using numerical simulation data contaminated with artificial noise. An MCL is developed to test the proposed algorithm. Parameter estimates for different models are compared with the calculated values derived from the mechanical and hydraulic properties of the MCL to validate model structures. In conclusion, the W4 model with an inertance and an aortic characteristic resistance arranged in series is proposed to represent the mock systemic arterial system. Once model structure is appropriately selected, IUKF can provide reasonable estimation accuracy in a limited time and may be helpful for future clinical applications.
引用
收藏
页码:2672 / 2680
页数:9
相关论文
共 29 条
  • [11] Unscented filtering and nonlinear estimation
    Julier, SJ
    Uhlmann, JK
    [J]. PROCEEDINGS OF THE IEEE, 2004, 92 (03) : 401 - 422
  • [12] Estimation of Three- and Four-Element Windkessel Parameters Using Subspace Model Identification
    Kind, Taco
    Faes, Theo J. C.
    Lankhaar, Jan-Willem
    Vonk-Noordegraaf, Anton
    Verhaegen, Michel
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (07) : 1531 - 1538
  • [13] In Vitro Validation of Finite-Element Model of AAA Hemodynamics Incorporating Realistic Outlet Boundary Conditions
    Kung, Ethan O.
    Les, Andrea S.
    Medina, Francisco
    Wicker, Ryan B.
    McConnell, Michael V.
    Taylor, Charles A.
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2011, 133 (04):
  • [14] Li J. K., 2004, Dynamics of the vascular system
  • [15] Design and initial testing of a mock human circulatory loop for left ventricular assist device performance testing
    Liu, Y
    Allaire, PJ
    Wood, H
    Olsen, D
    [J]. ARTIFICIAL ORGANS, 2005, 29 (04) : 341 - 345
  • [16] Liu Yingjie, 2006, Cardiovasc Eng, V6, P151, DOI 10.1007/s10558-006-9019-z
  • [17] ASYMPTOTIC-BEHAVIOR OF THE EXTENDED KALMAN FILTER AS A PARAMETER ESTIMATOR FOR LINEAR-SYSTEMS
    LJUNG, L
    [J]. IEEE TRANSACTIONS ON AUTOMATIC CONTROL, 1979, 24 (01) : 36 - 50
  • [18] Characterization of an adult mock circulation for testing cardiac support devices
    Pantalos, GM
    Koenig, SC
    Gillars, KJ
    Giridharan, GA
    Ewert, DL
    [J]. ASAIO JOURNAL, 2004, 50 (01) : 37 - 46
  • [19] Aortic input impedance in infants and children
    Sharp, MK
    Pantalos, GM
    Minich, L
    Tani, LY
    McGough, EC
    Hawkins, JA
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2000, 88 (06) : 2227 - 2239
  • [20] A Dynamical State Space Representation and Performance Analysis of a Feedback-Controlled Rotary Left Ventricular Assist Device
    Simaan, Marwan A.
    Ferreira, Antonio
    Chen, Shaohi
    Antaki, James F.
    Galati, David G.