In Vitro Identification of Four-Element Windkessel Models Based on Iterated Unscented Kalman Filter

被引:22
作者
Huang, Huan [1 ]
Yang, Ming [1 ]
Zang, Wangfu [2 ]
Wu, Shunjie [1 ]
Pang, Yafei [1 ]
机构
[1] Shanghai Jiao Tong Univ, Dept Instrument Sci & Engn, Shanghai 200240, Peoples R China
[2] Shanghai Jiao Tong Univ, Dept Cardiac Surg, Ruijin Hosp, Sch Med, Shanghai 200240, Peoples R China
基金
中国国家自然科学基金;
关键词
Mock circulatory loop (MCL); system identification; total arterial inertance; unscented Kalman filter (UKF); Windkessel model; MOCK CIRCULATION; ARTERIAL; IMPEDANCE; LOOP; BED;
D O I
10.1109/TBME.2011.2161477
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mock circulatory loops (MCLs) have been widely used to test left ventricular assist devices. The hydraulic properties of the mock systemic arterial system are usually described by two alternative four-element windkessel (W4) models. Compared with three-element windkessel model, their parameters, especially the inertial term, are much more difficult to estimate. In this paper, an estimator based on the iterated unscented Kalman filter (IUKF) algorithm is proposed to identify model parameters. Identifiability of these parameters for different measurements is described. Performance of the estimator for different model structures is first evaluated using numerical simulation data contaminated with artificial noise. An MCL is developed to test the proposed algorithm. Parameter estimates for different models are compared with the calculated values derived from the mechanical and hydraulic properties of the MCL to validate model structures. In conclusion, the W4 model with an inertance and an aortic characteristic resistance arranged in series is proposed to represent the mock systemic arterial system. Once model structure is appropriately selected, IUKF can provide reasonable estimation accuracy in a limited time and may be helpful for future clinical applications.
引用
收藏
页码:2672 / 2680
页数:9
相关论文
共 29 条
  • [1] [Anonymous], 1999, SYSTEM IDENTIFICATIO
  • [2] ASSOUDI E, 2002, INT J CONTROL, V75, P784
  • [3] TRACKING TIME-VARYING PROPERTIES OF THE SYSTEMIC VASCULAR BED
    AVANZOLINI, G
    BARBINI, P
    CAPPELLO, A
    CEVENINI, G
    POHL, V
    SIKORA, T
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (03) : 373 - 381
  • [4] THE ITERATED KALMAN FILTER UPDATE AS A GAUSS-NEWTON METHOD
    BELL, BM
    CATHEY, FW
    [J]. IEEE TRANSACTIONS ON AUTOMATIC CONTROL, 1993, 38 (02) : 294 - 297
  • [5] Development of systemic arterial mechanical properties from infancy to adulthood interpreted by four-element windkessel models
    Burattini, Roberto
    Di Salvia, Paola Oriana
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2007, 103 (01) : 66 - 79
  • [6] QUANTITATIVE-EVALUATION OF THE SYSTEMIC ARTERIAL BED BY PARAMETER-ESTIMATION OF A SIMPLE-MODEL
    DESWYSEN, B
    CHARLIER, AA
    GEVERS, M
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1980, 18 (02) : 153 - 166
  • [7] Ceiling radiant cooling: Comparison of ARMAX and subspace identification modelling methods
    Ferkl, Lukas
    Siroky, Jan
    [J]. BUILDING AND ENVIRONMENT, 2010, 45 (01) : 205 - 212
  • [8] Frank O., 1895, Z. Biol. (Mnch.), V32, P370
  • [9] A physiologically representative in vitro model of the coronary circulation
    Geven, MCF
    Bohté, VN
    Aarnoudse, WH
    van den Berg, PMJ
    Rutten, MCM
    Pijls, NHJ
    van de Vosse, FN
    [J]. PHYSIOLOGICAL MEASUREMENT, 2004, 25 (04) : 891 - 904
  • [10] Experimental Verification of the Feasibility of the Cardiovascular Impedance Simulator
    Gwak, Kwan-Woong
    Paden, Brad E.
    Antaki, James F.
    Ahn, Ihn-Seok
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (05) : 1176 - 1183