Three-Dimensional Echocardiography in Paravalvular Aortic Regurgitation Assessment after Transcatheter Aortic Valve Implantation

被引:75
|
作者
Goncalves, Alexandra [1 ,2 ]
Almeria, Carlos [1 ]
Marcos-Alberca, Pedro [1 ]
Feltes, Gisela [1 ]
Hernandez-Antolin, Rosana [1 ]
Rodriguez, Enrique [1 ]
Silva Cardoso, Jose C. [2 ]
Macaya, Carlos [1 ]
Luis Zamorano, Jose [1 ]
机构
[1] Univ Complutense, Hosp Clin San Carlos, Cardiovasc Inst, E-28040 Madrid, Spain
[2] Univ Porto, Sch Med, Dept Cardiol, P-4100 Oporto, Portugal
关键词
3D transthoracic echocardiography; Paravalvular aortic regurgitation; Transcatheter aortic valve implantation; Vena contracta; TRANSESOPHAGEAL ECHOCARDIOGRAPHY; VALVULAR REGURGITATION; EUROPEAN-ASSOCIATION; FOLLOW-UP; STENOSIS; AREA; RECOMMENDATIONS; FEASIBILITY; REPLACEMENT; EXPERIENCE;
D O I
10.1016/j.echo.2011.08.019
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Paravalvular aortic regurgitation (AR) after transcatheter aortic valve implantation (TAVI) is common, but the evaluation of its severity by two-dimensional (2D) transthoracic echocardiography (TTE) presents several constrains. The aim of this study was to assess the usefulness of a new methodology, using three-dimensional (3D) TTE, for better assessment of paravalvular AR after TAVI. Methods: Two-dimensional and 3D TTE was performed in 72 patients, 5 months after TAVI, using the X5-1 PureWave microbeamforming xMATRIX probe. The position and severity of the paravalvular AR jets were described using 2D and 3D TTE, and a model was designed for paravalvular AR systematic location description. Vena contracta width was measured using 2D transthoracic echocardiographic views, and the planimetry of the vena contracta was assessed after the perfect alignment plane was obtained using the multiplanar 3D transthoracic echocardiographic reconstruction tool. AR volume was calculated as the difference between 3D TTE-derived total left ventricular stroke volume and right ventricular stroke volume estimated using 2D TTE. Diagnostic efficiency for moderate AR was assessed using receiver operating characteristic curve analysis. Results: Forty-three patients (57.4%) presented with AR; 10 (13.3%) had central AR, and 33 (44.0%) had paravalvular AR jets. Vena contracta widths were similar between patients with moderate and mild AR (2.1 +/- 0.53 vs 1.9 +/- 0.16 mm, P = .16), but vena contracta planimetry was larger in patients with moderate AR than in those with mild AR (0.30 +/- 0.12 vs 0.09 +/- 0.07 cm(2), P = .001). Vena contracta planimetry on 3D TTE was better correlated with AR volume than vena contracta width on 2D TTE (Kendall's tau = 0.82 [P < .001] vs 0.66 [P < .001]). The areas under the receiver operating characteristic curves were 0.96 for vena contracta planimetry and 0.35 for vena contracta width. Conclusions: This study proposes an alternative methodology for paravalvular AR assessment after TAVI. Using vena contracta planimetry on 3D TTE, an accurate methodology for paravalvular AR jet evaluation and moderate AR classification is described. (J Am Soc Echocardiogr 2012;25:47-55.)
引用
收藏
页码:47 / 55
页数:9
相关论文
共 50 条
  • [1] Prosthesis/annulus discongruence assessed by three-dimensional transoesophageal echocardiography: A predictor of significant paravalvular aortic regurgitation after transcatheter aortic valve implantation
    Santos, Ninel
    Alberto de Agustin, Jose
    Almeria, Carlos
    Goncalves, Alejandra
    Marcos-Alberca, Pedro
    Fernandez-Golfin, Covadonga
    Garcia, Eulogio
    Hernandez-Antolin, Rosana
    Perez de Isla, Leopoldo
    Macaya, Carlos
    Zamorano, Jose
    EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2012, 13 (11) : 931 - 937
  • [2] Three-dimensional printing for assessment of paravalvular leak in transcatheter aortic valve implantation
    Thorburn, Casey
    Abdel-Razek, Omar
    Fagan, Susan
    Pearce, Neil
    Furey, Michael
    Harris, Scott
    Bartellas, Michael
    Adams, Corey
    JOURNAL OF CARDIOTHORACIC SURGERY, 2020, 15 (01)
  • [3] Three-dimensional printing for assessment of paravalvular leak in transcatheter aortic valve implantation
    Casey Thorburn
    Omar Abdel-Razek
    Susan Fagan
    Neil Pearce
    Michael Furey
    Scott Harris
    Michael Bartellas
    Corey Adams
    Journal of Cardiothoracic Surgery, 15
  • [4] Comparison of Two- and Three-Dimensional Transthoracic Echocardiography to Cardiac Magnetic Resonance Imaging for Assessment of Paravalvular Regurgitation After Transcatheter Aortic Valve Implantation
    Altiok, Ertunc
    Frick, Michael
    Meyer, Christian G.
    Al Ateah, Ghazi
    Napp, Andreas
    Kirschfink, Annemarie
    Almalla, Mohammad
    Lotfi, Shahran
    Becker, Michael
    Herich, Lena
    Lehmacher, Walter
    Hoffmann, Rainer
    AMERICAN JOURNAL OF CARDIOLOGY, 2014, 113 (11): : 1859 - 1866
  • [5] Impact of paravalvular aortic regurgitation after transcatheter aortic valve implantation on survival
    Takagi, Hisato
    Umemoto, Takuya
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2016, 221 : 46 - 51
  • [6] Aortic valve calcium scoring is a predictor of paravalvular aortic regurgitation after transcatheter aortic valve implantation
    Colli, Andrea
    Gallo, Michele
    Bernabeu, Eduardo
    D'Onofrio, Augusto
    Tarzia, Vincenzo
    Gerosa, Gino
    ANNALS OF CARDIOTHORACIC SURGERY, 2012, 1 (02) : 156 - 159
  • [7] Impact of Aortic Valve and Annular Calcification on Paravalvular Aortic Regurgitation After Transcatheter Aortic Valve Replacement: A Real-Time Three-Dimensional Transesophageal Echocardiography Study
    Mihara, Hirotsugu
    Shibayama, Kentaro
    Harada, Kenji
    Berdejo, Javier
    Garudevan, Swaminatha V.
    Siegel, Robert
    Jilaihawi, Hasan
    Makkar, Raj R.
    Shiota, Takahiro
    CIRCULATION, 2013, 128 (22)
  • [8] Prosthesis/annulus discongruence by three-dimensional transesophageal echocardiography: predictor of significant aortic regurgitation after transcatheter aortic valve implantation
    Santos, N.
    Olmos, C.
    De Agustin Loeches, J. A.
    Marcos-Alberca, P.
    Fernandez-Golfin, C.
    Almeria, C.
    Rodrigo, J. L.
    Zamorano, J.
    Macaya, C.
    Perez De Isla, L.
    EUROPEAN HEART JOURNAL, 2012, 33 : 715 - 715
  • [9] New methodology for paravalvular aortic regurgitation assessment in the era of transcatheter aortic valve implantation
    Goncalves, A.
    Almeria, C.
    Marcos-Alberca, P.
    Feltes, G.
    Hernandez-Antolin, R.
    Rodriguez, E.
    Garcia, E.
    Silva Cardoso, J.
    Macaya, C.
    Zamorano, J.
    EUROPEAN HEART JOURNAL, 2011, 32 : 186 - 186
  • [10] Progression and Prognosis of Paravalvular Regurgitation After Transcatheter Aortic Valve Implantation
    Meneguz-Moreno, Rafael Alexandre
    de Castro-Filho, Antonio
    de Oliveira Ramos, Auristela Isabel
    Zumarraga, Mayra
    Le Bihan, David
    Barretto, Rodrigo
    de Alvim Siqueira, Dimytri Alexandre
    Cunha Abizaid, Alexandre Antonio
    de Moraes Rego Sousa, Amanda Guerra
    Eduardo Sousa, J.
    ARQUIVOS BRASILEIROS DE CARDIOLOGIA, 2017, 109 (06) : 590 - 598