A new solid state tritium surface monitor

被引:4
作者
Willms, RS
Dogruel, D
Myers, R
Farrell, R
机构
[1] Los Alamos Natl Lab, Los Alamos, NM 87544 USA
[2] Radiat Monitoring Devices Inc, Watertown, MA 02472 USA
关键词
D O I
10.13182/FST05-A955
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Traditionally the amount of tritium on a surface is determined by swiping the surface with a material such as filter paper and counting the removed tritium by scintillation. While effective, this method can be time consuming, can alter the surface, only measures removable tritium and produces radioactive waste. For a given application each of these considerations may or may not be a disadvantage. A solid state monitor, on the other hand, has the potential to provide rapid analysis, not alter the surface, measure all tritium on a surface and produce little or not radioactive waste. This allure has promoted open wall ion chamber and PIN diode-based tritium surface monitor development, and these techniques have enjoyed certain success. Recently the first tests were performed with an avalanche photodiode (APD) for surface tritium measurement. While quite similar in concept to PIN diode based measurements, side-by-side testing showed that the APD provided substantially better counting efficiency. Considerations included count rate, background, sensitivity, stability and effect of ambient light. Of particular importance in the US, the APD was able to measure concentrations down to the "free release" limit, i.e., the concentration below which items can be removed from radiological control areas.
引用
收藏
页码:409 / 412
页数:4
相关论文
共 50 条
  • [21] THE HIGH-PRESSURE EQUATION OF STATE OF SOLID MOLECULAR TRITIUM
    DRIESSEN, A
    SILVERA, IF
    JOURNAL OF LOW TEMPERATURE PHYSICS, 1984, 54 (5-6) : 565 - 585
  • [22] AMBULATORY ACTIVITY MONITOR WITH SOLID-STATE MEMORY
    COLBURN, TR
    SMITH, BM
    GUARINI, JJ
    SIMMONS, NN
    ISA TRANSACTIONS, 1976, 15 (02) : 149 - 154
  • [23] MONITOR TEAMS SPARE GATES AND SOLID STATE LAMPS
    PROSSER, TF
    ELECTRONICS, 1970, 43 (20): : 79 - &
  • [24] PORTABLE, SOLID-STATE, PLUTONIUM WOUND MONITOR
    ONEAL, BL
    HEALTH PHYSICS, 1978, 35 (06): : 878 - 879
  • [25] Portable intelligent tritium in air monitor
    Purghel, L
    Calin, MR
    Bartos, D
    Serbia, L
    Lupu, A
    Lupu, F
    Lupu, D
    FUSION SCIENCE AND TECHNOLOGY, 2005, 48 (01) : 390 - 392
  • [26] New solid-state fluorescence sensor used to monitor photosynthetic parameters and algal biomass
    Donovan, Chelsea D.
    Ellison, Robert M.
    OCEANS 2005, VOLS 1-3, 2005, : 375 - 377
  • [27] A tritium activity monitor for the KATRIN Experiment
    Schmitt, Udo
    ASTROPARTICLE, PARTICLE AND SPACE PHYSICS, DETECTORS AND MEDICAL PHYSICS APPLICATIONS, 2008, 4 : 254 - 257
  • [28] Aqueous effluent tritium monitor development
    Hofstetter, K.J.
    Wilson, H.T.
    Fusion Technology, 1992, 21 (2 pt 2): : 446 - 451
  • [29] AN ONLINE TRITIUM-IN-WATER MONITOR
    SINGH, AN
    RATNAKARAN, M
    VOHRA, KG
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1985, 236 (01) : 159 - 164
  • [30] A PORTABLE TRITIUM-IN-AIR MONITOR
    OSBORNE, RV
    COVEART, AS
    HEALTH PHYSICS, 1970, 19 (02): : 354 - +