Comparative prospective randomized efficacy testing of different guiding catheters for coronary sinus cannulation in heart failure patients

被引:12
作者
Butter, C
Gras, D
Ritter, P
Stellbrink, C
Fleck, E
Tockman, B
Schubert, B
Pochet, T
deVoogt, W
机构
[1] German Heart Inst, Dept Internal Med Cardiol, D-13353 Berlin, Germany
[2] St Lucas Andreas Hosp, Amsterdam, Netherlands
[3] Guidant Res, Brussels, Belgium
[4] Guidant Res, St Paul, MN USA
[5] Univ Aachen Klinikum, Aachen, Germany
[6] Clin Bizet, Paris, France
[7] Polyclin St Henri, Nantes, France
关键词
coronary sinus; cannulation; heart failure patients;
D O I
10.1023/A:1027439309485
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The efficacy of different shapes of guiding catheters for coronary sinus (CS) cannulation was evaluated at implant of a cardiac resynchronization therapy device that required transvenous placement of a pacing lead in a coronary vein on the left ventricle. Methods and Results: Comparative testing of 5 different guiding catheter shapes was attempted in 29 consecutive patients. Four newer guiding catheters were tested in a prospective randomized manner with a reference standard catheter always tested in the last place. All of the new catheters included a distal shape that consisted of a single curve or a compound set of curves at the catheter tip, followed by a catheter shaft design which fell into one of two classes: (1) straight shaft catheters and (2) curved shaft catheters. The curved shaft class of catheters, two new and one standard, achieved higher (16/23 (70%), 17/23 (74%) and 17/22 (77%)) CS cannulation success rates than the straight shaft class catheters ((13/23 (56%) and 12/23 (52%)). The pairing of two catheters, one from each class in a single patient, or both from the curved shaft class resulted in a combined CS cannulation success rate of up to 96%. Two of the catheters were more effective from a left hand insertion than from a right hand insertion site. Conclusions: The ability to switch between guiding catheters with different shapes and compound curves can improve the CS cannulation success rate. The selection of guiding catheters should be influenced by the intended use of the catheter either from the left or from the right hand side.
引用
收藏
页码:343 / 351
页数:9
相关论文
共 12 条
  • [1] Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure
    Auricchio, A
    Stellbrink, C
    Block, M
    Sack, S
    Vogt, J
    Bakker, P
    Klein, H
    Kramer, A
    Ding, J
    Salo, R
    Tockman, B
    Pochet, T
    Spinelli, J
    [J]. CIRCULATION, 1999, 99 (23) : 2993 - 3001
  • [2] Human experience with transvenous biventricular defibrillation using an electrode in a left ventricular vein
    Butter, C
    Meisel, E
    Engelmann, L
    Eldar, M
    Fleck, E
    Schubert, B
    Hahn, S
    Michel, U
    Pfeiffer, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (03): : 324 - 331
  • [3] Transvenous biventricular defibrillation halves energy requirements in patients
    Butter, C
    Meisel, E
    Tebbenjohanns, J
    Engelmann, L
    Fleck, E
    Schubert, B
    Hahn, S
    Pfeiffer, D
    [J]. CIRCULATION, 2001, 104 (21) : 2533 - 2538
  • [4] Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay.
    Cazeau, S
    Leclercq, C
    Lavergne, T
    Walker, S
    Varma, C
    Linde, C
    Garrigue, S
    Kappenberger, L
    Haywood, GA
    Santini, M
    Bailleul, C
    Daubert, JC
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2001, 344 (12) : 873 - 880
  • [5] Cardiac resynchronization therapy in advanced heart failure the multicenter InSync clinical study
    Gras, D
    Leclercq, C
    Tang, ASL
    Bucknall, C
    Luttikhuis, HO
    Kirstein-Pedersen, A
    [J]. EUROPEAN JOURNAL OF HEART FAILURE, 2002, 4 (03) : 311 - 320
  • [6] Gurley J, 2002, EUR HEART J, V23, P51
  • [7] KOCOVIC D, 2002, JACC SA, V39
  • [8] Maertens SLJJ, 2001, EUR HEART J, V22, P131
  • [9] Effectiveness of resynchronization therapy in patients with end-stage heart failure
    Molhoek, SG
    Bax, JJ
    van Erven, L
    Bootsma, M
    Boersma, E
    Steendijk, P
    van der Wall, EE
    Schalij, MJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (04) : 379 - 383
  • [10] PROCLEMER A, 2002, PACE 2, V25, P690