Finite element analysis of a condylar support prosthesis to replace the temporomandibular joint

被引:25
作者
Abel, Eric W. [1 ]
Hilgers, Andre [1 ]
McLoughlin, Philip M. [2 ]
机构
[1] Univ Dundee, Sch Engn Phys & Math, Dundee DD1 4HN, Scotland
[2] Ninewells Hosp & Med Sch, Dundee DD1 9SY, Scotland
关键词
Temporomandibular joint prosthesis; TMJ implant design; Condylar support; Finite element analysis; TMJ IMPLANT; BONE; MECHANOSTAT; GEOMETRY; STRAIN; MODEL; LOAD;
D O I
10.1016/j.bjoms.2015.01.016
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
This paper presents a finite element study of a temporomandibular joint (TMJ) prosthesis in which the mandibular component sits on the condyle after removal of only the diseased articular surface and minimal amount of condylar bone. The condylar support prosthesis (CSP) is customised to fit the patient and allows a large part of the joint force to be transmitted through the condyle to the ramus, rather than relying only on transfer of the load by the screws that fix the prosthesis to the ramus. The 3-dimensional structural finite element analysis compared a design of CSP with a standard commercial prosthesis and one that was modified to fit the ramus, to relate the findings to the different designs and geometrical features. The models simulated an incisal bite under high loading. In the CSP and in its fixation screws, the stresses were much lower than those in the other 2 prostheses and the bone strains were at physiological levels. The CSP gives a more physiological form of load transfer than is possible without the condylar contact, and considerably reduces the amount of strain on the bone around the screws. (C) 2015 The British Association of Oral and Maxillofacial Surgeons. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:352 / 357
页数:6
相关论文
共 23 条
[1]   Influence of Thickness and Contact Surface Geometry of Condylar Stem of TMJ Implant on Its Stability [J].
Arabshahi, Zohreh ;
Kashani, Jamal ;
Kadir, Mohammed Rafiq Abdul ;
Azari, Abbas .
2011 INTERNATIONAL CONFERENCE ON PHYSICS SCIENCE AND TECHNOLOGY (ICPST), 2011, 22 :414-419
[2]  
Beaupre GS, 1987, DEV BIOMECH, V3
[3]   A 3-DIMENSIONAL INVESTIGATION OF TEMPOROMANDIBULAR-JOINT LOADING [J].
FAULKNER, MG ;
HATCHER, DC ;
HAY, A .
JOURNAL OF BIOMECHANICS, 1987, 20 (10) :997-1002
[4]  
Field Clarice, 2010, J Dent Biomech, V2010, P902537, DOI 10.4061/2010/902537
[5]   BONE MASS AND THE MECHANOSTAT - A PROPOSAL [J].
FROST, HM .
ANATOMICAL RECORD, 1987, 219 (01) :1-9
[6]  
FROST HM, 1983, CLIN ORTHOP RELAT R, P286
[7]   Effect of Screw Fixation on Temporomandibular Joint Condylar Prosthesis [J].
Hsu, Jui-Ting ;
Huang, Heng-Li ;
Tsai, Ming-Tzu ;
Fuh, Lib-Jyh ;
Tu, Ming-Gene .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 69 (05) :1320-1328
[8]   Functional significance of strain distribution in the human mandible under masticatory load: Numerical predictions [J].
Ichim, I. ;
Kieser, J. A. ;
Swain, M. V. .
ARCHIVES OF ORAL BIOLOGY, 2007, 52 (05) :465-473
[9]   Finite Element Analysis of a TMJ Implant [J].
Kashi, A. ;
Chowdhury, A. Roy ;
Saha, S. .
JOURNAL OF DENTAL RESEARCH, 2010, 89 (03) :241-245
[10]  
Keyak JH, 1998, J BIOMECH, V31, P125, DOI 10.1016/S0021-9290(97)00123-1