Fabric tears as a new cause of type III endoleak with Ancure endograft

被引:38
作者
Teutelink, A [1 ]
van der Laan, MJ [1 ]
Milner, R [1 ]
Blankensteijn, JD [1 ]
机构
[1] Univ Utrecht, Ctr Med, Dept Vasc Surg, NL-3508 GA Utrecht, Netherlands
关键词
D O I
10.1016/S0741-5214(03)00416-6
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: We present two case reports of type IIIb endoleak. One was due to fabric erosion caused by placement of a stent (Wallstent; W. L. Gore & Associates, Flagstaff, Ariz) after endovascular aneurysm repair; the other arose spontaneously. In both cases, an Ancure endograft (Guidant/EVT, Menlo Park, Calif) was placed. Case reports: In case 1, a large endoleak developed 36 months after uncomplicated endovascular treatment of an abdominal aortic aneurysm. In case 2, endoleak developed 30 months after a complicated procedure. In both cases, two Wallstents were used to treat type I endoleak and limb kinking in the first postoperative months. One type 111 endoleak was within the endograft at the level of the stents. Conclusion: To our knowledge, these are the first type Ill endoleaks reported in association with Ancure endografts. Placement of Wallstents inside endografts is of concern, and another indication for close monitoring during follow-up.
引用
收藏
页码:843 / 846
页数:4
相关论文
共 20 条
[1]   Treatment of a failed bifurcated abdominal aortic stent graft with thrombolysis and wallstent placement [J].
Amesur, NB ;
Zajko, AB ;
Makaroun, MS .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 8 (05) :795-798
[2]  
Becquemin JP, 2002, J ENDOVASC THER, V9, P203, DOI 10.1583/1545-1550(2002)009<0203:EFDSAT>2.0.CO
[3]  
2
[4]  
Beebe HG, 2001, J VASC SURG, V33, pS55
[5]  
Boyle JR, 1998, J ENDOVASC SURG, V5, P216, DOI 10.1583/1074-6218(1998)005<0216:TAKOUA>2.0.CO
[6]  
2
[7]   Failure of endovascular abdominal aortic aneurysm graft limbs [J].
Carpenter, JP ;
Neschis, DG ;
Fairman, RM ;
Barker, CF ;
Golden, MA ;
Velazquez, OC ;
Mitchell, ME ;
Baum, RA .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (02) :296-302
[8]   Clinical experience with a bifurcated endovascular graft for abdominal aortic aneurysm repair [J].
Chuter, TAM ;
Risberg, B ;
Hopkinson, BR ;
Wendt, G ;
Scott, RAP ;
Walker, PJ ;
Viscomi, S ;
White, G .
JOURNAL OF VASCULAR SURGERY, 1996, 24 (04) :655-666
[9]   Stent-graft design: the good, the bad and the ugly [J].
Chuter, TAM .
CARDIOVASCULAR SURGERY, 2002, 10 (01) :7-13
[10]  
Dias NV, 2001, J ENDOVASC THER, V8, P268, DOI 10.1583/1545-1550(2001)008<0268:IPEDAS>2.0.CO