Correlation between gamma index passing rate and clinical dosimetric difference for pre-treatment 2D and 3D volumetric modulated arc therapy dosimetric verification

被引:46
|
作者
Jin, X. [1 ]
Yan, H. [1 ]
Han, C. [1 ]
Zhou, Y. [1 ]
Yi, J. [1 ]
Xie, C. [1 ]
机构
[1] Wenzhou Med Univ, Affiliated Hosp 1, Dept Radiotherapy & Chemotherapy, Wenzhou, Peoples R China
来源
BRITISH JOURNAL OF RADIOLOGY | 2015年 / 88卷 / 1047期
关键词
QUALITY-ASSURANCE; PER-BEAM; IMRT; VMAT; QA; TOMOTHERAPY; PHANTOM; CANCER; BOOST;
D O I
10.1259/bjr.20140577
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To investigate comparatively the percentage gamma passing rate (% GP) of two-dimensional (2D) and three-dimensional (3D) pre-treatment volumetric modulated arc therapy (VMAT) dosimetric verification and their correlation and sensitivity with percentage dosimetric errors (% DE). Methods: % GP of 2D and 3D pre-treatment VMAT quality assurance (QA) with different acceptance criteria was obtained by ArcCHECK (R) (Sun Nuclear Corporation, Melbourne, FL) for 20 patients with nasopharyngeal cancer (NPC) and 20 patients with oesophageal cancer. % DE were calculated from planned dose-volume histogram (DVH) and patients' predicted DVH calculated by 3DVH (R) software (Sun Nuclear Corporation). Correlation and sensitivity between % GP and % DE were investigated using Pearson's correlation coefficient (r) and receiver operating characteristics (ROCs). Results: Relatively higher % DE on some DVH-basedmetrics were observed for both patients with NPC and oesophageal cancer. Except for 2%/2mm criterion, the average % GPs for all patients undergoing VMAT were acceptablewith average rates of 97.11% +/- 1.54% and 97.39% +/- 1.37% for 2D and 3D 3%/3mm criteria, respectively. The number of correlations for 3D was higher than that for 2D (21 vs 8). However, the general correlation was still poor for all the analysed metrics (9 out of 26 for 3D 3%/3mm criterion). The average area under the curve (AUC) of ROCs was 0.66 +/- 0.12 and 0.71 +/- 0.21 for 2D and 3D evaluations, respectively. Conclusions: There is a lack of correlation between % GP and % DE for both 2D and 3D pre-treatment VMAT dosimetric evaluation. DVH-based dose metrics evaluation obtained from 3DVH will provide more useful analysis. Advances in knowledge: Correlation and sensitivity of % GP with % DE for VMAT QA were studied for the first time.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Pre-treatment 2D and 3D dosimetric verification of volumetric arc therapy. A correlation study between gamma index passing rate and clinical dose volume histogram
    Szczurek, Lukasz
    Juszkat, Robert
    Szczurek, Jolanta
    Turek, Ilona
    Sosnowski, Piotr
    PLOS ONE, 2019, 14 (08):
  • [2] DOSIMETRIC PERFORMANCE OF A 2D DIODE ARRAY FOR VERIFICATION OF VOLUMETRIC MODULATED ARC THERAPY
    Nakada, R.
    Tatsumi, D.
    Sakamoto, M.
    Inoue, M.
    Ichida, T.
    Ishii, K.
    Tsutsumi, S.
    Ogino, R.
    Hosono, M.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S212 - S212
  • [3] 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry
    Mans, Anton
    Remeijer, Peter
    Olaciregui-Ruiz, Igor
    Wendling, Markus
    Sonke, Jan-Jakob
    Mijnheer, Ben
    van Herk, Marcel
    Stroom, Joep C.
    RADIOTHERAPY AND ONCOLOGY, 2010, 94 (02) : 181 - 187
  • [4] Evaluation of 3D pre-treatment verification for volumetric modulated arc therapy plan in head region
    Ruangchan, S.
    Oonsiri, S.
    Suriyapee, S.
    13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [5] A Dosimetric Comparison Between Volumetric Modulated Arc and 3D Treatment Plans for Synchronous Bilateral Breast Patients
    Levin, D.
    Shekel, E.
    Epstein, D.
    Tova, Y.
    Zalmanov-Faermann, S.
    Pfeffer, R.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S750 - S751
  • [6] Dosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques
    Clemente-Gutierrez, Francisco
    Perez-Vara, Consuelo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (02): : 198 - 217
  • [7] Dosimetric precision of 3D gel dosimetry compared with radiochromic films in volumetric modulated arc therapy
    Skyt, P. S.
    Hoye, E. M.
    Yates, E. S.
    Schmidt, M. L.
    Petersen, J. B. B.
    Balling, P.
    Muren, L. P.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S409 - S410
  • [8] From 2D to 3D gamma passing rate tolerance and action limits for patient-specific quality assurance in volumetric-modulated arc therapy
    Zarros, Christos
    Patatoukas, George
    Kollaros, Nikos
    Chalkia, Marina
    Kougioumtzopoulou, Andromachi
    Kouloulias, Vasilios
    Platoni, Kalliopi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025,
  • [9] Correlation analysis between 2D and 3D patient-specific quality assurance for volumetric modulated arc therapy
    Sharma, S.
    Sharma, D.
    Subramani, V.
    Gopishankar, N.
    Bhaskar, S.
    Pathy, S.
    Kumar, P.
    Thulkar, S.
    Chander, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (04): : 685 - 691
  • [10] Correlation analysis between 2D and quasi-3D gamma evaluations for both intensity-modulated radiation therapy and volumetric modulated arc therapy
    Kim, Jung-in
    Choi, Chang Heon
    Wu, Hong-Gyun
    Kim, Jin Ho
    Kim, Kyubo
    Park, Jong Min
    ONCOTARGET, 2017, 8 (03) : 5449 - 5459