Planimetric and continuity equation assessment of aortic valve area: Head to head comparison between cardiac magnetic resonance and echocardiography

被引:42
作者
Pouleur, Anne-Catherine [1 ]
de Waroux, Jean-Benoit le Polain [1 ]
Pasquet, Agnes [1 ]
Vancraeynest, David [1 ]
Vanoverschelde, Jean-Louis J. [1 ]
Gerber, Bernhard L. [1 ]
机构
[1] Univ Catholique Louvain, Dept Cardiovasc Dis, Clin Univ St Luc, B-1200 Brussels, Belgium
关键词
cardiac magnetic resonance; echocardiography; aortic stenosis; planimetry; continuity equation; comparative studies;
D O I
10.1002/jmri.21182
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the accuracy of planimetric and continuitv equation measurements of aortic valve area (AVA) by cardiac MR (cMR) to each other and against transthoracic (TTE) and transesophageal (TEE) echocardiography. Materials and Methods: A total of 31 patients (21 men, mean age = 67 +/- 13 years) with aortic stenosis (AS) and 16 controls (12 men, mean age = 57 +/- 14 years) underwent measurement of AVA by planimetric and continuity equation cMR. Measurements were compared to TEE planimetry and continuity equation TTE. Results: AVA by continuity equation cMR correlated highly to continuity equation TTE (r = 0.98) and was not significantly different (1.8 +/- 1.3 cm(2) VS. 1.8 +/- 1.4 cm(2), p = 0.62). Similarly. AVA by cMR planimetry was not statistically different from TEE planimetry (2.1 +/- 1.7 cm(2) vs. 2.1 +/- 1.6 cm(2), p = 0.34) and correlated highly (r = 0.98). Yet planimetric measurements of AVA by cMR and TEE were significantly higher than AVA by continuity equation cMR (P < 0.001 and P < 0.001, respectively) and TTE (P < 0.001 and P < 0.001. respectively). Conclusion: Both planimetry and continuity equation-based measurements of AVA by cMR are equally accurate. However. similar to TEE, cMR AVA is larger by planimetry than by continuity equation. This is consistent with the contention that the anatomical maximum opening of a stenotic aortic valve is larger than the size of the functional vena contracta.
引用
收藏
页码:1436 / 1443
页数:8
相关论文
共 26 条
[1]   Variation of anatomic valve area during ejection in patients with valvular aortic stenosis evaluated by two-dimensional echocardiographic planimetry: Comparison with traditional Doppler data [J].
Arsenault, M ;
Masani, N ;
Magni, G ;
Yao, JF ;
Deras, L ;
Pandian, N .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (07) :1931-1937
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Quantification of aortic stenosis in mechanically ventilated patients using multiplane transesophageal Doppler echocardiography [J].
Blumberg, FC ;
Pfeifer, M ;
Holmer, SR ;
Kromer, EP ;
Riegger, GAJ ;
Elsner, D .
CHEST, 1998, 114 (01) :94-97
[4]   DEPENDENCE OF GORLIN FORMULA AND CONTINUITY EQUATION VALVE AREAS ON TRANSVALVULAR VOLUME FLOW-RATE IN VALVULAR AORTIC-STENOSIS [J].
BURWASH, IG ;
THOMAS, DD ;
SADAHIRO, M ;
PEARLMAN, AS ;
VERRIER, ED ;
THOMAS, R ;
KRAFT, CD ;
OTTO, CM .
CIRCULATION, 1994, 89 (02) :827-835
[5]   Practical value of cardiac magnetic resonance imaging for clinical quantification of aortic valve stenosis comparison with echocardiography [J].
Caruthers, SD ;
Lin, SJ ;
Brown, P ;
Watkins, MP ;
Williams, TA ;
Lehr, KA ;
Wickline, SA .
CIRCULATION, 2003, 108 (18) :2236-2243
[6]   Planimetry of aortic valve area in aortic stenosis by magnetic resonance imaging [J].
Debl, K ;
Djavidani, B ;
Seitz, J ;
Nitz, WF ;
Schmid, FX ;
Muders, F ;
Buchner, S ;
Feuerbach, S ;
Riegger, G ;
Luchner, A .
INVESTIGATIVE RADIOLOGY, 2005, 40 (10) :631-636
[7]   Analysis of the effect of flow rate on the Doppler continuity equation for stenotic orifice area calculations - A numerical study [J].
DeGroff, CG ;
Shandas, R ;
Valdes-Cruz, L .
CIRCULATION, 1998, 97 (16) :1597-1605
[8]   Quantification of valvular aortic stenosis by magnetic resonance imaging [J].
Friedrich, MG ;
Schulz-Menger, J ;
Poetsch, T ;
Pilz, B ;
Uhlich, F ;
Dietz, R .
AMERICAN HEART JOURNAL, 2002, 144 (02) :329-334
[9]   Effect of three-dimensional valve shape on the hemodynamics of aortic stenosis: Three-dimensional echocardiographic stereolithography and patient studies [J].
Gilon, D ;
Cape, EG ;
Handschumacher, MD ;
Song, JK ;
Solheim, J ;
VanAuker, M ;
King, MEE ;
Levine, RA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 40 (08) :1479-1486
[10]   PIVOTAL ROLE OF AORTIC-VALVE AREA CALCULATION BY THE CONTINUITY EQUATION FOR DOPPLER ASSESSMENT OF AORTIC-STENOSIS IN PATIENTS WITH COMBINED AORTIC-STENOSIS AND REGURGITATION [J].
GRAYBURN, PA ;
SMITH, MD ;
HARRISON, MR ;
GURLEY, JC ;
DEMARIA, AN .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (04) :376-381