Phantom based qualitative and quantitative evaluation of artifacts in MR images of metallic hip prostheses

被引:18
作者
Mansson, Sven [1 ]
Muller, Gunilla M. [1 ]
Wellman, Fredrik [1 ]
Nittka, Mathias [2 ]
Lundin, Bjorn [3 ]
机构
[1] Lund Univ, Skane Univ Hosp Malmo, Dept Clin Sci Malmo, S-22100 Lund, Sweden
[2] Siemens AG, Healthcare Sect, Imaging & Therapy Div, Erlangen, Germany
[3] Lund Univ, Skane Univ Hosp Lund, Dept Clin Sci Lund, S-22100 Lund, Sweden
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2015年 / 31卷 / 02期
关键词
MR-imaging; Artifacts; Phantom studies; Prostheses; Musculoskeletal imaging; REDUCTION; ARTHROPLASTY; SEQUENCE;
D O I
10.1016/j.ejmp.2014.12.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop methods for qualitative and quantitative evaluation of MRI artifacts near metallic prostheses, and to compare the efficiency of different artifact suppression techniques with different types of hip prostheses. Methods: Three hip prostheses of cobalt-chromium, stainless steel, and titanium were embedded in agarose gel together with a rectilinear grid. Coronal MR images of the prostheses were acquired on a 1.5T scanner. Three pulse sequences were evaluated; TSE: a high-bandwidth turbo spin echo; VAT: TSE with view angle tilting, SEMAC: TSE with both VAT and slice distortion correction (6, 10 or 16 z-phase-encoding steps). Through-plane distortions were assessed as the length of visible gridlines, in-plane artifacts as the artifact area, and total artifacts by subtraction of an ideal, undistorted image from the actual image. Results: VAT reduced in-plane artifacts by up to 50% compared to TSE, but did not reduce through-plane artifacts. SEMAC reduced through-plane artifacts by 60-80% compared to TSE and VAT. SEMAC in-plane artifacts were from 20% higher (6 encoding steps) to 50% lower (16 steps) than VAT. Total artifacts were reduced by 60-80% in the best sequence (SEMAC, 16 steps) compared to the worst (TSE). The titanium prosthesis produced 3-4 times lower artifact scores than the other prostheses. Conclusions: A rectilinear grid phantom is useful for qualitative and quantitative evaluation of artifacts provoked by different MRI protocols and prosthesis models. VAT and SEMAC were superior to TSE with high bandwidth. A proper number of z-encoding steps in SEMAC was critical. The titanium prosthesis caused least artifacts. (C) 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:173 / 178
页数:6
相关论文
共 17 条
[1]   SEMAC-VAT and MSVAT-SPACE Sequence Strategies for Metal Artifact Reduction in 1.5T Magnetic Resonance Imaging [J].
Ai, Tao ;
Padua, Abraham ;
Goerner, Frank ;
Nittka, Mathias ;
Gugala, Zbigniew ;
Jadhav, Siddharth ;
Trelles, Miguel ;
Johnson, Raleigh F. ;
Lindsey, Ronald W. ;
Li, Xiaoming ;
Runge, Val M. .
INVESTIGATIVE RADIOLOGY, 2012, 47 (05) :267-276
[2]  
[Anonymous], 2007, ANN BOOK ASTM STAND
[3]   High rate of revision and a high incidence of radiolucent lines around Metasul metal-on-metal total hip replacements RESULTS FROM A RANDOMISED CONTROLLED TRIAL OF THREE BEARINGS AFTER SEVEN YEARS [J].
Bjorgul, K. ;
Novicoff, W. N. ;
Andersen, S. T. ;
Ahlund, O. R. ;
Bunes, A. ;
Wiig, M. ;
Brevig, K. .
BONE & JOINT JOURNAL, 2013, 95B (07) :881-886
[4]   New MR Imaging Methods for Metallic Implants in the Knee: Artifact Correction and Clinical Impact [J].
Chen, Christina A. ;
Chen, Weitian ;
Goodman, Stuart B. ;
Hargreaves, Brian A. ;
Koch, Kevin M. ;
Lu, Wenmaio ;
Brau, Anja C. ;
Draper, Christine E. ;
Delp, Scott L. ;
Gold, Garry E. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 33 (05) :1121-1127
[5]   TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11
[6]   Total hip arthroplasty with use of the metasul metal-on-metal articulation - Four to seven-year results [J].
Dorr, LD ;
Wang, ZI ;
Longjohn, DB ;
Dubois, B ;
Murken, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (06) :789-798
[7]   Accelerated Slice Encoding for Metal Artifact Correction [J].
Hargreaves, Brian A. ;
Chen, Weitian ;
Lu, Wenmiao ;
Alley, Marcus T. ;
Gold, Garry E. ;
Brau, Anja C. S. ;
Pauly, John M. ;
Pauly, Kim Butts .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 31 (04) :987-996
[8]   MRI After Arthroplasty: Comparison of MAVRIC and Conventional Fast Spin-Echo Techniques [J].
Hayter, Catherine L. ;
Koff, Matthew F. ;
Shah, Parina ;
Koch, Kevin M. ;
Miller, Theodore T. ;
Potter, Hollis G. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (03) :W405-W411
[9]   A Multispectral Three-Dimensional Acquisition Technique for Imaging Near Metal Implants [J].
Koch, Kevin M. ;
Lorbiecki, John. E. ;
Hinks, R. Scott ;
King, Kevin F. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (02) :381-390
[10]   Quantifying image distortion of orthopedic materials in magnetic resonance imaging [J].
Koff, Matthew F. ;
Shah, Parina ;
Koch, Kevin M. ;
Potter, Hollis G. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (03) :610-618