Temperature monitoring in radiofrequency catheter ablation of atrial flutter using the linear ablation technique

被引:12
|
作者
Wen, ZC
Chen, SA
Tai, CT
Chiang, CE
Lee, SH
Chen, YJ
Yu, WC
Huang, JL
Chang, MS
机构
[1] VET GEN HOSP,DIV CARDIOL,TAIPEI,TAIWAN
[2] NATL YANG MING UNIV,SCH MED,DEPT MED,DIV CARDIOL,TAIPEI 112,TAIWAN
[3] SHIN KONG MEM HOSP,DEPT MED,TAIPEI,TAIWAN
[4] VET GEN HOSP,DEPT MED,DIV CARDIOL,TAICHUNG,TAIWAN
关键词
atrial flutter; temperature control; radiofrequency; catheter ablation;
D O I
10.1111/j.1540-8167.1996.tb00480.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Information about temperature and impedance monitoring during radiofrequency catheter linear ablation of atrial flutter has not been reported. We proposed that a radiofrequency catheter ablation system using a closed-loop temperature control model could decrease the incidence of coagulum formation and shorten the radiation exposure and procedure times compared with those found in a power control model. Methods and Results: Forty patients (8 women and 32 men; mean age 64 +/- 7 years) with atrial flutter were referred for radiofrequency ablation. The patients were randomized into two groups: group I patients underwent radiofrequency catheter linear ablation of atrial flutter using a power control of energy output model; and group II patients underwent the closed-loop temperature control model with a target electrode temperature of 70 degrees C. As compared with group II, group I patients had a higher incidence of coagulum formation (12% vs 2%, P < 0.05), temperature shutdown (11% vs 0%, P < 0.01), and impedance shutdown (16% vs 3%, P < 0.01), more radiofrequency applications (7 +/- 3 vs 4 +/- 2, P < 0.01), and longer procedure time (100 +/- 25 vs 75 +/- 23 minutes, P < 0.05) and radiation exposure time (31 +/- 10 vs 20 +/- 7 minutes, P < 0.05) required for successful ablation. Larger deviations of temperature (9.0 degrees +/- 2.4 degrees C vs 5.0 degrees +/- 1.2 degrees C, P < 0.0001) and impedance (9.2 +/- 2.6 Ohm vs 5.3 +/- 1.6 Ohm, P < 0.0001) were also found in group I patients compared with those in group II. Conclusions: This study demonstrated that a closed-loop temperature control model could facilitate the effects of radiofrequency catheter ablation of the atrial flutter circuit by decreasing coagulum formation, temperature and impedance shutdown, and procedure and radiation exposure times.
引用
收藏
页码:1050 / 1057
页数:8
相关论文
共 50 条
  • [31] Radiofrequency Catheter Ablation of Common Atrial Flutter: Role of the Eustachian Valve
    Frank Halimi
    Françoise Hidden-Lucet
    Joelci Tonet
    Guy Fontaine
    Robert Frank
    Journal of Interventional Cardiac Electrophysiology, 1999, 3 : 169 - 172
  • [32] Radiofrequency catheter ablation of atrial flutter after orthotopic heart transplantation
    Li, YG
    Gronefeld, G
    Hohnloser, SH
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (11) : 1086 - 1090
  • [33] Comparative characteristics of different methods of radiofrequency catheter ablation in atrial flutter
    Yakushev, A., V
    Paratsii, O. Z.
    ZAPOROZHYE MEDICAL JOURNAL, 2022, 24 (05) : 516 - 520
  • [34] Radiofrequency catheter ablation of common atrial flutter: Role of the Eustachian valve
    Halimi, F
    Hidden-Lucet, F
    Tonet, J
    Fontaine, G
    Frank, R
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 1999, 3 (02) : 169 - 172
  • [35] Radiofrequency ablation for cure of atrial flutter
    Kalman, JM
    Vohra, JK
    Jayaprakash, S
    Sparks, PB
    AUSTRALIAN AND NEW ZEALAND JOURNAL OF MEDICINE, 1997, 27 (06): : 653 - 657
  • [36] First Human Experience of the Linear Ablation Catheter for the Ablation of Right Atrial Flutter (Linear LAAF-141 Study): First Results
    Grimaldi, Massimo
    Katsouras, Grigorios
    Quadrini, Federico
    Langialonga, Tommaso
    CIRCULATION, 2012, 126 (21)
  • [37] Electrocardiographic patterns and results of radiofrequency catheter ablation of clockwise type I atrial flutter
    Saoudi, N
    Nair, M
    Abdelazziz, A
    Poty, H
    Daou, A
    Anselme, F
    Letac, B
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (10) : 931 - 942
  • [38] Apparent Bidirectional Conduction Block Following Radiofrequency Catheter Ablation of Typical Atrial Flutter
    Robert F. Quintos
    Tawfik Barakat
    Andrew Mecca
    Brian Olshansky
    Journal of Interventional Cardiac Electrophysiology, 2001, 5 : 109 - 118
  • [39] Sinus node dysfunction after radiofrequency catheter ablation of atrial flutter: A case report
    Li, Haiwei
    Wang, Zefeng
    Wu, Yongquan
    ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2023, 28 (02)
  • [40] Successful radiofrequency catheter ablation of common atrial flutter at the bottom of a coronary sinus diverticulum
    Igarashi, Miyako
    Tada, Hiroshi
    Sekiguchi, Yukio
    Aonuma, Kazutaka
    HEART AND VESSELS, 2012, 27 (06) : 643 - 647