Measurement of meniscofemoral contact pressure after repair of bucket-handle tears with biodegradable implants

被引:67
作者
Becker, R
Wirz, D
Wolf, C
Göpfert, B
Nebelung, W
Friederich, N
机构
[1] Univ Magdeburg, Dept Orthopaed Surg, D-39120 Magdeburg, Germany
[2] Univ Basel, Orthopaed Biomech Lab, Basel, Switzerland
[3] Dist Hosp, Dept Orthopaed Surg, Muhlhausen, Germany
[4] Marien Hosp, Dept Orthopaed Surg, Dusseldorf, Germany
[5] Kantonsspital, Dept Orthopaed Surg, Bruderholz, Switzerland
关键词
biodegradable implant; meniscus repair; meniscofemoral pressure;
D O I
10.1007/s00402-004-0739-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction Biodegradable implants are frequently used for meniscus repair. Articular cartilage damage has been reported recently after meniscus repair with biodegradable implants. The aim of the study was to investigate the meniscofemoral contact pressure at the posterior horn of the medial and lateral meniscus after repair of bucket-handle lacerations. Materials and methods Specimens were mounted in a materials testing machine (Bionix 858, MTS) which was equipped with a load cell. The quadriceps tendon was attached to a hydraulic cylinder, and knee motion was controlled via tension of the quadriceps tendon. A piezo-resistive system (Tekscan, Boston, MA, USA) measured the meniscofemoral contact pressure. Five different types of biodegradable implants ( Arrow, Dart, Fastener, Stinger and Meniscal Screw) and horizontal suture ( no. 2 Ethibond) were tested. The knee was extended from 90 degrees of flexion to 0 degrees under a constant load of 350 N due to adjustment of the tension force of the quadriceps tendon. The femorotibial pressure and contact area were recorded at 0 degrees, 30 degrees, 60 degrees and 90 degrees of flexion. Results The meniscofemoral pressure did not increase after meniscus repair with biodegradable implants or sutures. The meniscofemoral peak pressure at the posterior horn was 1.46 +/- 1.54 MPa in the medial compartment and 1.08 +/- 1.17 MPa in the lateral compartment at full knee extension. The meniscofemoral pressure increased significantly in both compartments with knee flexion from 0 degrees to 90 degrees. Conclusion Biodegradable implants for meniscus repair do not affect the meniscofemoral pressure. However, there remains a risk of damage to the cartilage when barbed implants are used. If the implant is not entirely advanced into the meniscus, the sharp head or some of the barbs at the column of the implant may come into direct contact with the articular cartilage of the femoral condyle or tibial plateau. The authors presume that incorrect positioning of the implant seems to be the major reason for cartilage damage.
引用
收藏
页码:254 / 260
页数:7
相关论文
共 42 条
[11]  
FAVENESI JA, 1983, ORTHOP T, V8, P174
[12]   THE PREVALENCE OF KNEE OSTEOARTHRITIS IN THE ELDERLY - THE FRAMINGHAM OSTEOARTHRITIS STUDY [J].
FELSON, DT ;
NAIMARK, A ;
ANDERSON, J ;
KAZIS, L ;
CASTELLI, W ;
MEENAN, RF .
ARTHRITIS AND RHEUMATISM, 1987, 30 (08) :914-918
[13]  
FITHIAN DC, 1990, CLIN ORTHOP RELAT R, P19
[14]   THE CONTACT AREA AND PRESSURE DISTRIBUTION PATTERN OF THE KNEE - A STUDY OF NORMAL AND OSTEOARTHROTIC KNEE JOINTS [J].
FUKUBAYASHI, T ;
KUROSAWA, H .
ACTA ORTHOPAEDICA SCANDINAVICA, 1980, 51 (06) :871-879
[15]   Methodology and apparatus to determine material properties of the knee joint meniscus [J].
Goertzen, DJ ;
Budney, DR ;
Cinats, JG .
MEDICAL ENGINEERING & PHYSICS, 1997, 19 (05) :412-419
[16]   An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: a comparison of K-scan sensor and Fuji film [J].
Harris, ML ;
Morberg, P ;
Bruce, WJM ;
Walsh, WR .
JOURNAL OF BIOMECHANICS, 1999, 32 (09) :951-958
[17]  
HIGUCHI H, 2000, CLIN ORTHOP RELAT R, V377, P161
[18]   STABILIZING MECHANISMS OF LOADED AND UNLOADED KNEE-JOINT [J].
HSIEH, HH ;
WALKER, PS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1976, 58 (01) :87-93
[19]  
KOHN D, 1989, Arthroscopy, V5, P324, DOI 10.1016/0749-8063(89)90150-3
[20]   MATERIAL PROPERTIES OF MENISCUS TISSUE AND INFLUENCE OF THE MENISCI ON KNEE COMPRESSION [J].
KOHN, D ;
MUSSACK, T ;
PLITZ, W .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1993, 131 (05) :397-403