Single cycle to failure in torsion of three standard and five locking plate constructs

被引:37
作者
Cabassu, J. B. [1 ]
Kowaleski, M. P. [1 ]
Shorinko, J. K. [2 ]
Blake, C. A. [1 ]
Gaudette, G. R. [2 ]
Boudrieau, R. J. [1 ]
机构
[1] Tufts Univ, Cummings Sch Vet Med, Dept Clin Sci, North Grafton, MA 01536 USA
[2] Worcester Polytech Inst, Worcester, MA 01609 USA
关键词
Torsion; bone plates; locking plates; in vitro; plate failure; DYNAMIC COMPRESSION PLATE; LIMITED CONTACT; INTERNAL-FIXATION; SHEEP TIBIA; MM LOCKING; IN-VIVO; IMPLANTS; TITANIUM; FRACTURES; DESIGN;
D O I
10.3415/VCOT-11-04-0050
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objectives: The biomechanical properties of standard plates and recently designed locking plates were compared in torsion. We hypothesized that titanium (Ti) constructs would have the greatest deformation, and String of Pearls (SOP) constructs the greatest strength and stiffness. Methods: Dynamic compression plates (DCP), stainless steel (SS) limited contact (LC)-DCP, Ti LC-DCP, locking compression plate (LCP), 10 mm and 11 mm Advanced Locking Plate System (ALPS) 10 and 11, SOP and Fixin plates were applied to a validated bone model simulating a bridging osteosynthesis. Yield torque (strength), yield angle (deformation) and stiffness were compared using one-way ANOVA with post hoc Tukey (p <0.05). Results: The ALPS 11 constructs had significantly greater elastic deformation than all constructs except for the ALPS 10. There were not any differences in strength observed except for the ALPS 10 constructs, which was less than that for the SOP, LCP, DCP and ALPS 11 constructs. No differences in construct torsional stiffness were observed with the SS LC-DCP, DCP, LCP and SOP constructs; however all had greater stiffness than all remaining constructs. The ALPS 10 construct had lower stiffness than all constructs. Clinical significance: Modulus of elasticity of Ti explains the higher deformation and lower stiffness of these systems, with similar results for the Fixin due to its lower section modulus compared to all other plates. The SOP and standard constructs had surprisingly similar biomechanical properties in torsion. The rationale for selecting these implants for fracture repair likely needs to be based upon their differing biomechanical properties inherent to the diverse implant systems.
引用
收藏
页码:418 / 425
页数:8
相关论文
共 30 条
[1]   Mechanical evaluation of a new minimum-contact plate for internal fracture fixation [J].
Abel, EW ;
Sun, J .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1998, 12 (06) :382-386
[2]  
Acker M, 2010, P 19 ANN SCI M EUR C, P150
[3]  
Aguila AZ, 2005, VET COMP ORTHOPAED, V18, P220
[4]   Influence of materials for fixation implants on local infection - An experimental study of steel versus titanium DCP in rabbits [J].
Arens, S ;
Schlegel, U ;
Printzen, G ;
Ziegler, WJ ;
Perren, SM ;
Hansis, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (04) :647-651
[5]  
*ASTM, 1994, 1994 ANN BOOK ASTM S, P57
[6]   Single cycle to failure in bending of three standard and five locking plates and plate constructs [J].
Blake, C. A. ;
Boudrieau, R. J. ;
Torrance, B. S. ;
Tacvorian, E. K. ;
Cabassu, J. B. ;
Gaudette, G. R. ;
Kowaleski, M. P. .
VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2011, 24 (06) :408-417
[7]   Mechanical testing of 3.5 mm locking and non-locking bone plates [J].
DeTora, M. ;
Kraus, K. .
VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2008, 21 (04) :318-322
[8]   A biomechanical comparison of 3.5 locking compression plate fixation to 3.5 limited contact dynamic compression plate fixation in a canine cadaveric distal humeral metaphyseal gap model [J].
Filipowicz, D. ;
Lanz, O. ;
McLaughlin, R. ;
Elder, S. ;
Werre, S. .
VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2009, 22 (04) :270-277
[9]   Strain distribution in plated and unplated sheep tibia - An in vivo experiment [J].
Gautier, E ;
Perren, SM ;
Cordey, J .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 :37-44
[10]   Prospective Evaluation of Minimally Invasive Plate Osteosynthesis in 36 Nonarticular Tibial Fractures in Dogs and Cats [J].
Guiot, Laurent P. ;
Dejardin, Loic M. .
VETERINARY SURGERY, 2011, 40 (02) :171-182