Amperometric Tyrosinase Biosensor Based on Carbon Nanotube-Doped Sol-Gel-Derived Zinc Oxide-Nafion Composite Films

被引:27
|
作者
Lee, Joong Min [1 ,2 ]
Xu, Guang-Ri [3 ]
Kim, Byung Kun [1 ,2 ]
Choi, Han Nim [1 ,2 ]
Lee, Won-Yong [1 ,2 ]
机构
[1] Yonsei Univ, Dept Chem, Seoul 120749, South Korea
[2] Yonsei Univ, Ctr Bioact Mol Hybrids, Seoul 120749, South Korea
[3] Henan Inst Sci & Technol, Xinxiang 453003, Peoples R China
关键词
Biosensor; Phenol; Tyrosinase; Carbon nanotube; ZnO-Nafion composite; PHENOLIC-COMPOUNDS; MATRIX; SENSOR; CHEMILUMINESCENCE; WATER;
D O I
10.1002/elan.201000556
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A highly sensitive amperometric tyrosinase biosensor has been developed based on mesoporous composite films of carbon nanotube (CNT)-ZnO-Nafion composite film encapsulating tyrosinase on a glassy carbon electrode. Due to the mesoporous nature (diameter: 3.17 nm) of the composite film, the biosensor exhibits very fast response time of 2 s. The present biosensor shows an excellent sensitivity of 766 mA/M and a detection limit of 4.7 x 10(-8) M (S/N=3) for phenol. The biosensor retains 88% of its initial activity after 2 weeks of storage in 50 mM phosphate buffer at pH 7.0. The present biosensor exhibits good recovery for the spiked phenol in an industrial waste water sample.
引用
收藏
页码:962 / 970
页数:9
相关论文
共 50 条
  • [1] Amperometric glucose biosensor based on sol-gel-derived metal oxide/Nafion composite films
    Choi, HN
    Kim, MA
    Lee, WY
    ANALYTICA CHIMICA ACTA, 2005, 537 (1-2) : 179 - 187
  • [2] Amperometric ethanol biosensor based on carbon nanotubes dispersed in sol-gel-derived titania - nafion composite film
    Choi, Han Nim
    Lyu, Young-Ku
    Han, Jee Hoon
    Lee, Won-Yong
    ELECTROANALYSIS, 2007, 19 (14) : 1524 - 1530
  • [3] A Highly Sensitive Amperometric Galactose Biosensor Based on Graphene-doped Sol-gel-derived Titania-Nafion Composite Films
    Lee, Sumin
    Jeong, Daeho
    Lee, Sangjung
    Nam, Sungju
    Lee, Won-Yong
    ELECTROANALYSIS, 2022, 34 (04) : 629 - 639
  • [4] Amperometric glucose biosensor based on sol-gel-derived zirconia/Nafion composite film as encapsulation matrix
    Kim, HJ
    Yoon, SH
    Choi, HN
    Lyu, YK
    Lee, WY
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2006, 27 (01): : 65 - 70
  • [5] Amperometric tyrosinase biosensor based on a sol-gel-derived titanium oxide-copolymer composite matrix for detection of phenolic compounds
    Chen, X
    Cheng, GJ
    Dong, SJ
    ANALYST, 2001, 126 (10) : 1728 - 1732
  • [6] Amperometric tyrosinase biosensor based on carbon nanotube-Titania-Nafion composite film
    Lee, You-Jin
    Lyu, Young-Ku
    Choi, Han Nim
    Lee, Won-Yong
    ELECTROANALYSIS, 2007, 19 (10) : 1048 - 1054
  • [7] Sol-gel-derived amperometric biosensor for hydrogen peroxide based on methylene green incorporated in Nafion film
    Wang, BQ
    Dong, SJ
    TALANTA, 2000, 51 (03) : 565 - 572
  • [8] Carbon nanotube-doped composite sol–gel coatings for float glass
    A. A. Stepko
    A. S. Chainikova
    E. G. Vinokurov
    L. A. Orlova
    Inorganic Materials, 2016, 52 : 207 - 212
  • [9] Sol-gel-derived titanium oxide/copolymer composite based glucose biosensor
    Chen, X
    Dong, SJ
    BIOSENSORS & BIOELECTRONICS, 2003, 18 (08): : 999 - 1004
  • [10] Carbon nanotube-doped composite sol-gel coatings for float glass
    Stepko, A. A.
    Chainikova, A. S.
    Vinokurov, E. G.
    Orlova, L. A.
    INORGANIC MATERIALS, 2016, 52 (02) : 207 - 212