Evaluation of intra-aneurysmal hemodynamics after flow diverter placement in a patient-specific aneurysm model

被引:9
|
作者
Balasso, Andrea [1 ]
Bauer, Jan S. [1 ]
Liebig, Thomas [2 ]
Dorn, Franziska [2 ]
Zimmer, Claus [1 ]
Liepsch, Dieter [3 ]
Prothmann, Sascha [1 ]
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Dept Neuroradiol, D-80290 Munich, Germany
[2] Univ Cologne, Dept Neuroradiol, D-50931 Cologne, Germany
[3] Munich Univ Appl Sci, Munich, Germany
关键词
Stent; flow diversion; endovascular; non-Newtonian fluid; laser Doppler anemometry; PIPELINE EMBOLIZATION DEVICE; INTRACRANIAL ANEURYSMS; ENDOVASCULAR TREATMENT; SACCULAR ANEURYSMS; STENT; ARTERY; RECONSTRUCTION; THROMBOSIS; RUPTURE; STRESS;
D O I
10.3233/BIR-14019
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
BACKGROUND: The growth and rupture of cerebral aneurysms is intrinsically related to the hemodynamics prevailing in the diseased area. Therefore, a better understanding of intra-aneurysmal hemodynamics is essential for developing effective treatment methods. OBJECTIVE: The intention of this study was to evaluate the intra-aneurysmal flow and flow reduction induced by flow diverters in a true-to-scale elastic aneurysm model, obtained from real patient data. METHODS: Based on the computed tomography angiography (CTA) data of a fusiform aneurysm of a 34 year old patient, an elastic silicon rubber model of the aneurysm was produced. A physiologic pulsatile flow was created with a circulatory experimental set-up, and a non-Newtonian perfusion fluid was used as a substitute for human blood. Hemodynamics were measured by LDA before and after flow diverter implantation. RESULTS: Implantation of a flow diverter device resulted in a reduction of intra-aneurysmal maximum flow velocities of 97.8% at the inflow zone, 89.1% in the dome and 89.3% at the outflow zone, when compared to the native model. A significant reduction of 94% in the mean intra-aneurysmal velocity was found. CONCLUSIONS: This promising methodology can optimize patient treatment and will correlate with computational simulations to evaluate their reliability.
引用
收藏
页码:341 / 354
页数:14
相关论文
共 50 条
  • [41] Fusiform Aneurysm on the Basilar Artery Trunk Treated with Intra-Aneurysmal Embolization with Parent Vessel Occlusion after Complete Preoperative Occlusion Test
    Jung, Young-Jin
    Kim, Min-Soo
    Choi, Byung-Yon
    Chang, Chul-Hoon
    JOURNAL OF KOREAN NEUROSURGICAL SOCIETY, 2013, 53 (04) : 235 - 240
  • [42] Towards Predicting Patient-Specific Flow-Diverter Treatment Outcomes for Bifurcation Aneurysms: From Implantation Rehearsal to Virtual Angiograms
    Peach, T. W.
    Spranger, K.
    Ventikos, Y.
    ANNALS OF BIOMEDICAL ENGINEERING, 2016, 44 (01) : 99 - 111
  • [43] Patient-specific computer modeling of blood flow in cerebral arteries with aneurysm and stent
    Takizawa, Kenji
    Schjodt, Kathleen
    Puntel, Anthony
    Kostov, Nikolay
    Tezduyar, Tayfun E.
    COMPUTATIONAL MECHANICS, 2012, 50 (06) : 675 - 686
  • [44] Patient-specific computer modeling of blood flow in cerebral arteries with aneurysm and stent
    Kenji Takizawa
    Kathleen Schjodt
    Anthony Puntel
    Nikolay Kostov
    Tayfun E. Tezduyar
    Computational Mechanics, 2012, 50 : 675 - 686
  • [45] The Role of Hemodynamics in Intracranial Bifurcation Arteries after Aneurysm Treatment with Flow-Diverter Stents
    Narata, A. P.
    de Moura, F. S.
    Larrabide, I.
    Perrault, C. M.
    Patat, F.
    Bibi, R.
    Velasco, S.
    Januel, A. -C.
    Cognard, C.
    Chapot, R.
    Bouakaz, A.
    Sennoga, C. A.
    Marzo, A.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (02) : 323 - 330
  • [46] Thin-Film Patient-Specific Flow Diverter Stents for the Treatment of Intracranial Aneurysms
    Velvaluri, Prasanth
    Pravdivtseva, Mariya S.
    Berg, Philipp
    Wodarg, Fritz
    de Miranda, Rodrigo Lima
    Hoevener, Jan-Bernd
    Jansen, Olav
    Quandt, Eckhard
    ADVANCED MATERIALS TECHNOLOGIES, 2021, 6 (09):
  • [47] Leveraging Patient-Specific Simulated Angiograms to Characterize Cerebral Aneurysm Hemodynamics using Computational Fluid Dynamics
    Chivukula, V
    White, R.
    Shields, A.
    Davies, J.
    Mokin, M.
    Bednarek, D. R.
    Rudin, S.
    Ionita, C.
    MEDICAL IMAGING 2022: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2022, 12036
  • [48] Influence of Clip Locations on Intraaneurysmal Flow Dynamics in Patient-specific Anterior Communicating Aneurysm Models with Different Aneurysmal Angle
    Mu, Lizhong
    Chi, Qingzhuo
    Ji, Changjin
    He, Ying
    Gao, Ge
    CMES-COMPUTER MODELING IN ENGINEERING & SCIENCES, 2018, 116 (02): : 175 - 197
  • [49] Understanding the effect of effective metal surface area of flow diverter stent's on the patient-specific intracranial aneurysm numerical model using Lagrangian coherent structures
    Mutlu, Onur
    Olcay, Ali Bahadir
    Bilgin, Cem
    Hakyemez, Bahattin
    JOURNAL OF CLINICAL NEUROSCIENCE, 2020, 80 : 298 - 309
  • [50] EFFECT OF COIL SURFACE AREA ON THE HEMODYNAMICS OF A PATIENT-SPECIFIC INTRACRANIAL ANEURYSM: A COMPUTATIONAL STUDY
    Aguilar, M. L.
    Morales, H. G.
    Larrabide, I.
    Macho, J. M.
    San Roman, L.
    Frangi, A. F.
    2012 9TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), 2012, : 1180 - 1183