Biomechanical Testing of Additive Manufactured Proximal Humerus Fracture Fixation Plates

被引:19
作者
Tilton, Maryam [1 ]
Armstrong, April [2 ]
Sanville, Jennifer [3 ]
Chin, Matthew [3 ]
Hast, Michael W. [3 ]
Lewis, Gregory S. [2 ]
Manogharan, Guha P. [1 ]
机构
[1] Penn State Univ, Dept Mech Engn, 232 Reber Bldg, University Pk, PA 16802 USA
[2] Penn State Univ, Coll Med, Dept Orthopaed & Rehabil, Hershey, PA USA
[3] Univ Penn, Dept Orthopaed Surg, Biedermann Lab Orthopaed Res, Philadelphia, PA 19104 USA
关键词
Proximal humerus fracture; Additive manufacturing; 3D printing; Locking fixation plate; Biomechanics; Varus collapse; POWDER-BED FUSION; LOCKING PLATE; SHOULDER ARTHROPLASTY; MECHANICAL-PROPERTIES; CANCELLOUS BONE; MEDIAL SUPPORT; IMPLANTS; SCREW; REVERSE; DESIGN;
D O I
10.1007/s10439-019-02365-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Achieving satisfactory fracture fixation in osteoporotic patients with unstable proximal humerus fractures remains a major clinical challenge. Varus collapse is one of the more prominent complications that may lead to screw cutout. This aim of this study was to compare the fixation provided by conventional locking plates with novel design concepts that are only feasible through additive manufacturing (AM) techniques. In addition to reversed engineered implants, two novel implant designs with integrated struts were included in the study to provide medial support to humeral head. The medial strut was either solid or included a porous lattice structure intended to promote bone ingrowth. Biomechanical tests were performed using low density synthetic bones with simulated 3-part comminuted fractures. Nondestructive torsion and compression were performed, followed by increasing cyclic loading. The relative displacements between the bone fragments were determined using a 3D motion capture system. The AM manufactured implants with medial strut showed significant reduction of varus displacement during the increasing cyclic loading when compared to conventional designs. AM reversed-engineered locking plates showed similar mechanical behavior to conventional plates with identical geometry. This study demonstrates the feasibility and potential of employing alternative design via AM for fixation of unstable comminuted proximal humerus fractures to reduce fragment displacement.
引用
收藏
页码:463 / 476
页数:14
相关论文
共 73 条
[1]  
[Anonymous], J BONE JOINT SURG AM
[2]  
[Anonymous], 2013, PRACT ASSESSM RES EV, DOI DOI 10.7275/E4R6-DJ05
[3]  
[Anonymous], ORTHOP J SPORTS MED
[4]  
[Anonymous], SCI TECHNOLOGY
[5]  
[Anonymous], 2007, NOTES
[6]  
[Anonymous], TEST MAT COMP BON BI
[7]   Fully Porous 3D Printed Titanium Femoral Stem to Reduce Stress-Shielding Following Total Hip Arthroplasty [J].
Arabnejad, Sajad ;
Johnston, Burnett ;
Tanzer, Michael ;
Pasini, Damiano .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2017, 35 (08) :1774-1783
[8]   The biomechanical performance of locking plate fixation with intramedullary fibular strut graft augmentation in the treatment of unstable fractures of the proximal humerus [J].
Bae, J. -H. ;
Oh, J. -K. ;
Chon, C. -S. ;
Oh, C. -W. ;
Hwang, J. -H. ;
Yoon, Y. -C. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (07) :937-941
[9]   Which Non-Clinical Factors Influence the Use of Innovative Implants? The Example of Drug-Releasing Coronary Stents in Patients with Acute Myocardial Infarction: A Multilevel Regression Analysis [J].
Baeumler, M. .
GESUNDHEITSWESEN, 2013, 75 (12) :822-831
[10]   In vivo glenohumeral contact forces -: Measurements in the first patient 7 months postoperatively [J].
Bergmann, G. ;
Graichen, F. ;
Bender, A. ;
Kaab, M. ;
Rohlmann, A. ;
Westerhoff, P. .
JOURNAL OF BIOMECHANICS, 2007, 40 (10) :2139-2149