In vitro characterization of postheat-treated plasma-sprayed hydroxyapatite coatings

被引:39
作者
Lee, YP
Wang, CK
Huang, TH
Chen, CC
Kao, CT
Ding, SJ [1 ]
机构
[1] Chung Shan Med Univ, Inst Oral Mat Sci, Taichung 402, Taiwan
[2] Chung Shan Med Univ Hosp, Dept Dent, Taichung 402, Taiwan
[3] Kaohsiung Med Univ, Dept Med & Appl Chem, Kaohsiung 807, Taiwan
关键词
hydroxyapatite; plasma spraying; heat treatment; immersion test; electrochemical test;
D O I
10.1016/j.surfcoat.2004.09.025
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Plasma-sprayed hydroxyapatite (HA)-coated devices on metallic substrates have been widely used as dental and orthopedic implants. The in vitro stability of HA coatings has been concerned, serving as a long-term application. The purpose of this work was to use heat treatment to increase the crystallinity of HA coating and promote corrosion resistance, after which a comparative study in a simulated body fluid (SBF) was conducted through electrochemical techniques and immersion tests. The analyses of phase composition showed that the postdeposition heat treatment of the coatings led to recrystallization of amorphous calcium phosphate and conversion of nonapatite phases into apatite with a reduction of the plasma spray-induced "flake-type" structures. The result of bond strength test indicated no significant difference between the as-sprayed and heat-treated coatings. When immersed in SBF for 30 days, the heat-treated coatings did not have an apatite precipitation layer in contrast to untreated samples. However, the in vitro electrochemical test finding was that the heat-treated coatings showed a higher corrosion resistance, significantly reducing the corrosion rate and enhancing the corrosion potential. The increased crystalline HA content and dense structure of the heat-treated coatings could give rise to a substantial increase of in vitro stability. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:367 / 374
页数:8
相关论文
共 46 条
[1]   A study on corrosion resistance characteristics of PVD Cr-N coated steels by electrochemical method [J].
Ahn, SH ;
Choi, YS ;
Kim, JG ;
Han, JG .
SURFACE & COATINGS TECHNOLOGY, 2002, 150 (2-3) :319-326
[2]   Evaluation of degradation of thermal barrier coatings using impedance spectroscopy [J].
Ali, MS ;
Song, SH ;
Xiao, P .
JOURNAL OF THE EUROPEAN CERAMIC SOCIETY, 2002, 22 (01) :101-107
[3]   Surface analysis of PECVD organosilicon films for corrosion protection of steel substrates [J].
Angelini, E ;
d'Agostino, R ;
Fracassi, F ;
Grassini, S ;
Rosalbino, F .
SURFACE AND INTERFACE ANALYSIS, 2002, 34 (01) :155-159
[4]   A CALCIUM HYDROXYAPATITE PRECIPITATED FROM AN AQUEOUS-SOLUTION - AN INTERNATIONAL MULTIMETHOD ANALYSIS [J].
ARENDS, J ;
CHRISTOFFERSEN, J ;
CHRISTOFFERSEN, MR ;
ECKERT, H ;
FOWLER, BO ;
HEUGHEBAERT, JC ;
NANCOLLAS, GH ;
YESINOWSKI, JP ;
ZAWACKI, SJ .
JOURNAL OF CRYSTAL GROWTH, 1987, 84 (03) :515-532
[5]   Highly crystalline Mp-1™ hydroxylapatite coating Part I:: In vitro characterization and comparison to other plasma-sprayed hydroxylapatite coatings [J].
Burgess, AV ;
Story, BJ ;
La, D ;
Wagner, WR ;
LeGeros, JP .
CLINICAL ORAL IMPLANTS RESEARCH, 1999, 10 (04) :245-256
[6]   Highly crystalline Mp-1™ hydroxylapatite coating Part II:: In vivo performance on endosseous root implants in dogs [J].
Burgess, AV ;
Story, BJ ;
Wagner, WR ;
Trisi, P ;
Pikos, MA ;
Guttenberg, SA .
CLINICAL ORAL IMPLANTS RESEARCH, 1999, 10 (04) :257-266
[7]   EFFECT OF CALCIUM-PHOSPHATE (CA-P) COATINGS ON TRABECULAR BONE RESPONSE - A HISTOLOGICAL STUDY [J].
CAULIER, H ;
VANDERWAERDEN, JPCM ;
PAQUAY, YCGJ ;
WOLKE, JGC ;
KALK, W ;
NAERT, I ;
JANSEN, JA .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (09) :1061-1069
[8]  
Chen JY, 1997, J BIOMED MATER RES, V34, P15, DOI 10.1002/(SICI)1097-4636(199701)34:1<15::AID-JBM3>3.0.CO
[9]  
2-Q
[10]  
Claes L., 1990, HDB BIOACTIVE CERAMI, VII, P77