Flap Monitoring using Thermal Imaging Camera: A Contactless Method

被引:13
|
作者
Rabbani, Muhammad Jibran [1 ]
Bhatti, Asif Zubair [1 ]
Shahzad, Ahmed [1 ]
机构
[1] Shalamar Hosp, Dept Burn & Plast Surg, Lahore, Pakistan
来源
JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN | 2021年 / 31卷 / 06期
关键词
Flir one; Flap monitoring; Thermal imaging camera; INFRARED THERMOGRAPHY; SALVAGE;
D O I
10.29271/jcpsp.2021.06.703
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To determine the accuracy of the thermal imaging camera for the detection of vascular insult before the onset of permanent tissue damage. Study Design: Descriptive study. Place and Duration of Study: Department of Burn and Plastic Surgery, from July to December 2019. Methodology: The thermal image was used as an adjunct to clinical assessment in 84 flaps. Both pedicled and free flaps were performed in the study with a frequency of 68 (81%) and 16 (19%), respectively. Flap monitoring was performed regularly for five days by same team of surgeons and nurses. Flap vascularity was assessed by temperature gradient colour coding of thermal image taking the clinical assessment like colour, temperature, capillary refill, as a gold standard. Sensitivity and specificity of the thermal camera were calculated. Results: The age range of patients was from 5 years to 55 years. The diagnostic accuracy of thermal imaging camera was calculated to be 96.43% with a sensitivity of 98.7%, specificity of 75%. The positive predictive value was found to be 97.4%, negative predictive value was 85.7%, the false-negative value was 14.3%, and the false-positive value was 2.6%. Conclusion: Thermal imaging camera is an effective and useful adjunct for clinical assessment, as it is an easy and contactless technique.
引用
收藏
页码:703 / 706
页数:4
相关论文
共 50 条
  • [1] Transcutaneous oxygen measurement using ratiometric fluorescence imaging as a valid method for monitoring free flap transplants
    Schiltz, Daniel
    Taeger, Christian D.
    Biermann, Niklas
    Ranieri, Marco
    Klein, Silvan
    Prantl, Lukas
    Geis, Sebastian
    CLINICAL HEMORHEOLOGY AND MICROCIRCULATION, 2019, 73 (01) : 113 - 123
  • [2] Accuracy of Thermal Imaging Camera in Identification of Perforators
    Rabbani, Muhammad Jibran
    Ilyas, Asma
    Rabbani, Ammara
    ul Abidin, Zain
    Tarar, Moazzam Nazeer
    JCPSP-JOURNAL OF THE COLLEGE OF PHYSICIANS AND SURGEONS PAKISTAN, 2020, 30 (05): : 512 - 515
  • [3] Contactless Patient Care Using Hospital IoT: CCTV-Camera-Based Physiological Monitoring in ICU
    Wang, Haowen
    Huang, Jia
    Wang, Guowei
    Lu, Hongzhou
    Wang, Wenjin
    IEEE INTERNET OF THINGS JOURNAL, 2024, 11 (04): : 5781 - 5797
  • [4] Applications of thermal imaging camera for assessing structural integrity
    Tantele, Elia A.
    Votsis, Renos A.
    Kyriakides, Nicholas
    SIXTH INTERNATIONAL CONFERENCE ON REMOTE SENSING AND GEOINFORMATION OF THE ENVIRONMENT (RSCY2018), 2018, 10773
  • [5] Quality Control of Liquid Aluminium Alloy Using Thermal Imaging Camera
    Wladysiak, R.
    ARCHIVES OF FOUNDRY ENGINEERING, 2024, 24 (01) : 135 - 140
  • [6] Remote Welfare Monitoring of Rodents Using Thermal Imaging
    Pereira C.B.
    Kunczik J.
    Zieglowski L.
    Tolba R.
    Abdelrahman A.
    Zechner D.
    Vollmar B.
    Janssen H.
    Thum T.
    Czaplik M.
    Pereira, Carina Barbosa (cbarbosapere@ukaachen.de), 2018, MDPI (18):
  • [7] Remote Welfare Monitoring of Rodents Using Thermal Imaging
    Pereira, Carina Barbosa
    Kunczik, Janosch
    Zieglowski, Leonie
    Tolba, Rene
    Abdelrahman, Ahmed
    Zechner, Dietmar
    Vollmar, Brigitte
    Janssen, Heike
    Thum, Thomas
    Czaplik, Michael
    SENSORS, 2018, 18 (11)
  • [8] The Use of Thermal Imaging Techniques as a Method of Monitoring the New Born
    Luca, Catalina
    Corciova, C.
    Andritoi, D.
    Ciorap, R.
    6TH INTERNATIONAL CONFERENCE ON ADVANCEMENTS OF MEDICINE AND HEALTH CARE THROUGH TECHNOLOGY, MEDITECH 2018, 2019, 71 : 35 - 39
  • [9] Flap Monitoring Using Interstitial Fluid Glucose Measurements
    Kiuchi, Tomoki
    Ishii, Naohiro
    Uno, Takahiro
    Uoya, Yuichiro
    Sakai, Shigeki
    Matsuzaki, Kyoichi
    Kishi, Kazuo
    PLASTIC SURGERY, 2024, 32 (02) : 299 - 304
  • [10] Compensation method for the influence of angle of view on animal temperature measurement using thermal imaging camera combined with depth image
    Jiao, Leizi
    Dong, Daming
    Zhao, Xiande
    Han, Pengcheng
    JOURNAL OF THERMAL BIOLOGY, 2016, 62 : 15 - 19