A pilot study of mechanical chest compressions with the LUCAS™ device in cardiopulmonary resuscitation

被引:84
作者
Smekal, David [1 ]
Johansson, Jakob [1 ,2 ]
Huzevka, Tibor [1 ]
Rubertsson, Sten [1 ]
机构
[1] Uppsala Univ, Dept Surg Sci Anaesthesiol & Intens Care, SE-75185 Uppsala, Sweden
[2] Dept Emergency Med, SE-75185 Uppsala, Sweden
关键词
Cardiac arrest; Active compression-decompression (ACD); Cardiopulmonary resuscitation (CPR); Out-of-hospital CPR; Mechanical devices; Survival; Device; LUCAS; HOSPITAL CARDIAC-ARREST; TIDAL CARBON-DIOXIDE; BLOOD-FLOW; ACTIVE DECOMPRESSION; PERFUSION-PRESSURE; DEFIBRILLATION; QUALITY; OUTPUT; FORCE; CPR;
D O I
10.1016/j.resuscitation.2011.01.032
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Aim: The LUCASTM device has been shown to improve organ perfusion during cardiac arrest in experimental studies. In this pilot study the aim was to compare short-term survival between cardiopulmonary resuscitation (CPR) performed with mechanical chest compressions using the LUCASTM device and CPR performed with manual chest compressions. The intention was to use the results for power calculation in a larger randomised multicentre trial. Methods: In a prospective pilot study, from February 1, 2005, to April 1, 2007, 149 patients with out-of hospital cardiac arrest in two Swedish cities were randomised to mechanical chest compressions or standard CPR with manual chest compressions. Results: After exclusion, the LUCAS and the manual groups contained 75 and 73 patients, respectively. In the LUCAS and manual groups, spontaneous circulation with a palpable pulse returned in 30 and 23 patients (p = 0.30), spontaneous circulation with blood pressure above 80/50 mmHg remained for at least 5 min in 23 and 19 patients (p = 0.59), the number of patients hospitalised alive >4 h were 18 and 15 (p = 0.69), and the number discharged, alive 6 and 7 (p = 0.78), respectively. Conclusions: In this pilot study of out-of-hospital cardiac arrest patients we found no difference in early survival between CPR performed with mechanical chest compression with the LUCASTM device and CPR with manual chest compressions. Data have been used for power calculation in a forthcoming multicentre trial. (C) 2011 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:702 / 706
页数:5
相关论文
共 34 条
[1]   Mechanical active compression-decompression cardiopulmonary resuscitation (ACD-CPR) versus manual CPR according to pressure of end tidal carbon dioxide (PETCO2) during CPR in out-of-hospital cardiac arrest (OHCA) [J].
Axelsson, C. ;
Karlsson, T. ;
Axelsson, A. B. ;
Herlitz, J. .
RESUSCITATION, 2009, 80 (10) :1099-1103
[2]   Clinical consequences of the introduction of mechanical chest compression in the EMS system for treatment of out-of-hospital cardiac arrest -: A pilot study [J].
Axelsson, Christer ;
Nestin, Johan ;
Svensson, Leif ;
Axelsson, Asa B. ;
Herlitz, Johan .
RESUSCITATION, 2006, 71 (01) :47-55
[3]   Adverse hemodynamic effects of interrupting chest compressions for rescue breathing during cardiopulmonary resuscitation for ventricular fibrillation cardiac arrest [J].
Berg, RA ;
Sanders, AB ;
Kern, KB ;
Hilwig, RW ;
Heidenreich, JW ;
Porter, ME ;
Ewy, GA .
CIRCULATION, 2001, 104 (20) :2465-2470
[4]   Effects of interrupting precordial compressions on the calculated probability of defibrillation success during out-of-hospital cardiac arrest [J].
Eftestol, T ;
Sunde, K ;
Steen, PA .
CIRCULATION, 2002, 105 (19) :2270-2273
[5]   Active compression-decompression CPR necessitates follow-up post mortem [J].
Englund, E ;
Kongstad, PC .
RESUSCITATION, 2006, 68 (01) :161-162
[6]   Cardiopulmonary resuscitation - some physiological considerations [J].
Frenneaux, M .
RESUSCITATION, 2003, 58 (03) :259-265
[7]   Utstein style analysis of out-of-hospital cardiac arrest - Bystander CPR and end expired carbon dioxide [J].
Grmec, Stefek ;
Krizmaric, Mijenko ;
Mally, Stefan ;
Kozelj, Anton ;
Spindler, Mateja ;
Lesnik, Bojan .
RESUSCITATION, 2007, 72 (03) :404-414
[8]   EFFECT OF MEAN CIRCULATORY FILLING PRESSURE AND OTHER PERIPHERAL CIRCULATORY FACTORS ON CARDIAC OUTPUT [J].
GUYTON, AC ;
LINDSEY, AW ;
KAUFMANN, BN .
AMERICAN JOURNAL OF PHYSIOLOGY, 1955, 180 (03) :463-468
[9]   Manual chest compression vs use of an automated chest compression device during resuscitation following out-of-hospital cardiac arrest - A randomized trial [J].
Hallstrom, Al ;
Rea, Thomas D. ;
Sayre, Michael R. ;
Christenson, James ;
Anton, Andy R. ;
Mosesso, Vince N., Jr. ;
Van Ottingham, Lois ;
Olsufka, Michele ;
Pennington, Sarah ;
White, Lynn J. ;
Yahn, Stephen ;
Husar, James ;
Morris, Mary F. ;
Cobb, Leonard A. .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 295 (22) :2620-2628
[10]  
HERLITZ J, SWEDISH NATL REGISTR