Use of normalized relative line intensities for qualitative and semi-quantitative analysis in inductively coupled plasma atomic emission spectrometry using a custom segmented-array charge-coupled device detector .2. Applications to qualitative analysis with line-rich matrices

被引:11
作者
Rivier, C [1 ]
Mermet, JM [1 ]
机构
[1] UNIV LYON 1,SCI ANALYT LAB,URA CNRS 435,F-69622 VILLEURBANNE,FRANCE
关键词
inductively coupled plasma; emission spectrometry; qualitative analysis; normalized line intensity ratio; multichannel detection; steel and waspaloy analysis;
D O I
10.1366/0003702963905402
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In a previously published work, a procedure was described to conduct qualitative analysis in inductively coupled plasma atomic emission spectrometry, keeping in mind the possibility of spectral interferences. This procedure is based on the use of both line correlation and normalized relative line intensities of given elements. When spectral interferences due to a single or two major elements are observed for an analyte, use of multiple linear regression of the normalized relative line intensities of both the analyte and the major element provides information about the certainty of the presence of the analyte. This procedure has been used with an echelle grating-based dispersive system equipped with a custom segmented-array charge-coupled device detector and evaluated with the use of line-rich matrices such as Fe, Cr, and Ni. Satisfactory results were obtained even when several lines of the analyte suffered from interference. The only limitation was the number of available lines for a given analyte.
引用
收藏
页码:959 / 964
页数:6
相关论文
共 10 条
[1]   SOLID-STATE DETECTOR FOR ICP-OES [J].
BARNARD, TW ;
CROCKETT, MI ;
IVALDI, JC ;
LUNDBERG, PL ;
YATES, DA ;
LEVINE, PA ;
SAUER, DJ .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1231-1239
[2]   DESIGN AND EVALUATION OF AN ECHELLE GRATING OPTICAL-SYSTEM FOR ICP-OES [J].
BARNARD, TW ;
CROCKETT, MI ;
IVALDI, JC ;
LUNDBERG, PL .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1225-1230
[3]  
BILHORN RB, 1987, APPL SPECTROSC, V41, P114
[4]  
IVALDI JC, COMMUNICATION
[5]   CORRELATION-BASED DATA-PROCESSING FOR AN INDUCTIVELY COUPLED PLASMA FOURIER-TRANSFORM SPECTROMETER SYSTEM [J].
NG, RCL ;
HORLICK, G .
APPLIED SPECTROSCOPY, 1985, 39 (05) :834-840
[6]   INFORMATION-BASED EXPERT SYSTEMS FOR ATOMIC EMISSION-SPECTROSCOPY [J].
POMEROY, RS ;
KOLCZYNSKI, JD ;
DENTON, MB .
APPLIED SPECTROSCOPY, 1991, 45 (07) :1111-1119
[7]   SIMPLE EXPERIMENTS FOR THE CONTROL, THE EVALUATION AND THE DIAGNOSIS OF INDUCTIVELY-COUPLED PLASMA SEQUENTIAL SYSTEMS [J].
POUSSEL, E ;
MERMET, JM ;
SAMUEL, O .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (6-7) :743-755
[8]   USE OF NORMALIZED RELATIVE LINE-INTENSITIES FOR QUALITATIVE AND SEMIQUANTITATIVE ANALYSIS IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY USING A CUSTOM SEGMENTED-ARRAY CHARGE-COUPLED-DEVICE DETECTOR .1. PRINCIPLE AND FEASIBILITY [J].
SOUDIER, L ;
MERMET, JM .
APPLIED SPECTROSCOPY, 1995, 49 (10) :1478-1484
[9]   QUALITATIVE AND SEMIQUANTITATIVE ICAP-AES ANALYSIS USING A COMPUTER-CONTROLLED MONOCHROMATOR [J].
TAYLOR, PDP ;
DEDONDER, J .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (08) :783-796