Long-term follow-up of the pCONus device for the treatment of wide-neck bifurcation aneurysms

被引:3
作者
Guenego, Adrien [1 ]
Mine, Benjamin [1 ]
Bonnet, Thomas [1 ]
Elens, Stephanie [1 ]
Vazquez Suarez, Juan [1 ]
Jodaitis, Lise [2 ]
Ligot, Noemie [2 ]
Naeije, Gilles [2 ]
Lubicz, Boris [1 ]
机构
[1] Erasme Univ Hosp, Dept Intervent Neuroradiol, Route De Lennik 808, B-1070 Brussels, Belgium
[2] Erasme Univ Hosp, Dept Neurol, Anderlecht, Belgium
关键词
Aneurysm; coil; device; wide-neck; intervention; STENT-ASSISTED COILING; ENDOVASCULAR TREATMENT; INTRACRANIAL ANEURYSMS; WEB DEVICE; EMBOLIZATION; RECURRENCES;
D O I
10.1177/15910199211040279
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose Wide-neck bifurcation aneurysms remain challenging for the neurointerventionist and/or neurosurgeon despite many recent advances. The pCONus (Phenox, Bochum, Germany) is an emerging device for endovascular neck protection, we report the first long-term results of this device. Methods We performed a retrospective analysis of all consecutive intracranial wide-neck bifurcation aneurysms treated with the pCONus. Patients' characteristics were reviewed, procedural complications, angiographic (Roy-Raymond scale) and clinical outcomes were documented. Results Between January 2016 and September 2019, 43 patients (74% female, median age 56 [49-66] years) with 43 wide-neck bifurcation aneurysms (mean width of 6.8 +/- 2.1 mm, dome/neck ratio of 1.3 +/- 0.2 and neck of 5.2 +/- 1.3 mm) were included. A procedural angiographic complication was reported in five patients (12%), no patient presented a post-operative neurological deficit or long-term complication, mortality rate was 0%. At last follow-up (median of 46.5 months [38.3-51.7]), an adequate occlusion (complete and neck remnant) was observed in 37/43 patients (86%) and an aneurysm remnant in 6/43 (14%). Four patients (9%) needed retreatment. No in-stent stenosis or branch occlusion was depicted. Conclusion pCONus device provides a safe and efficient alternative for endovascular wide-neck bifurcation aneurysms management, with long-term stability.
引用
收藏
页码:455 / 462
页数:8
相关论文
共 40 条
[1]   Coil Occlusion of Wide-Neck Bifurcation Aneurysms Assisted by a Novel Intra- to Extra-Aneurysmatic Neck-Bridging Device (pCONus): Initial Experience [J].
Aguilar-Perez, M. ;
Kurre, W. ;
Fischer, S. ;
Baezner, H. ;
Henkes, H. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (05) :965-971
[2]  
Arat A, 2005, AM J NEURORADIOL, V26, P1768
[3]   Endovascular Treatment of Intracranial Aneurysms with the WEB Device: A Systematic Review of Clinical Outcomes [J].
Armoiry, X. ;
Turjman, F. ;
Hartmann, D. J. ;
Sivan-Hoffmann, R. ;
Riva, R. ;
Labeyrie, P. E. ;
Aulagner, G. ;
Gory, B. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2016, 37 (05) :868-872
[4]   "Y" and "X" Stent-Assisted Coiling of Complex and Wide-Neck Intracranial Bifurcation Aneurysms [J].
Bartolini, B. ;
Blanc, R. ;
Pistocchi, S. ;
Redjem, H. ;
Piotin, M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (11) :2153-2158
[5]   Stent-Assisted Coiling Versus Balloon Remodeling of Wide-Neck Aneurysms: Comparison of Angiographic Outcomes [J].
Chalouhi, N. ;
Starke, R. M. ;
Koltz, M. T. ;
Jabbour, P. M. ;
Tjoumakaris, S. I. ;
Dumont, A. S. ;
Rosenwasser, R. H. ;
Singhal, S. ;
Gonzalez, L. F. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (10) :1987-1992
[6]   Remnants and Recurrences After the Use of the WEB Intrasaccular Device in Large-Neck Bifurcation Aneurysms [J].
Cognard, Christophe ;
Januel, Anne Christine .
NEUROSURGERY, 2015, 76 (05) :522-530
[7]   Results of Embolization Used as the First Treatment Choice in a Consecutive Nonselected Population of Ruptured Aneurysms: Clinical Results of the Clarity GDC Study [J].
Cognard, Christophe ;
Pierot, Laurent ;
Anxionnat, Rene ;
Ricolfi, Frederic .
NEUROSURGERY, 2011, 69 (04) :837-841
[8]   How safe and effective are existing treatments for wide-necked bifurcation aneurysms? Literature-based objective performance criteria for safety and effectiveness [J].
Fiorella, David ;
Arthur, Adam S. ;
Chiacchierini, Richard ;
Emery, Evelyne ;
Molyneux, Andy ;
Pierot, Laurent .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2017, 9 (12) :1197-1201
[9]   pCONus Device for the Endovascular Treatment of Wide-Neck Middle Cerebral Artery Aneurysms [J].
Gory, B. ;
Aguilar-Perez, M. ;
Pomero, E. ;
Turjman, F. ;
Weber, W. ;
Fischer, S. ;
Henkes, H. ;
Biondi, A. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (09) :1735-1740
[10]   One-year Angiographic Results After pCONus Stent-Assisted Coiling of 40Wide-Neck Middle Cerebral Artery Aneurysms [J].
Gory, Benjamin ;
Aguilar-Perez, Marta ;
Pomero, Elisa ;
Turjman, Francis ;
Weber, Werner ;
Fischer, Sebastian ;
Henkes, Hans ;
Biondi, Alessandra .
NEUROSURGERY, 2017, 80 (06) :925-933