Testing of Flow-Based Microfluidic Biochips: Fault Modeling, Test Generation, and Experimental Demonstration

被引:68
作者
Hu, Kai [1 ]
Yu, Feiqiao [2 ]
Ho, Tsung-Yi [3 ]
Chakrabarty, Krishnendu [1 ]
机构
[1] Duke Univ, Dept Elect & Comp Engn, Durham, NC 27708 USA
[2] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[3] Natl Chiao Tung Univ, Dept Comp Sci, Hsinchu 300, Taiwan
关键词
Automatic test pattern generation (ATPG); defects; fault modeling; lab-on-chip; microfluidics; testing; LITHOGRAPHY;
D O I
10.1109/TCAD.2014.2336215
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
Recent advances in flow-based microfluidics have led to the emergence of biochemistry-on-a-chip as a new paradigm in clinical diagnostics and biomolecular recognition. However, a potential roadblock in the deployment of microfluidic biochips is the lack of test techniques to screen defective devices before they are used for biochemical analysis. Defective chips lead to repetition of experiments, which is undesirable due to high reagent cost and limited availability of samples. Prior work on fault detection in biochips has been limited to digital ("droplet") microfluidics and other electrode-based technology platforms. The paper proposes the first approach for automated testing of flow-based microfluidic biochips that are designed using membrane-based valves for flow control. The proposed test technique is based on a behavioral abstraction of physical defects in microchannels and valves. The flow paths and flow control in the microfluidic device are modeled as a logic circuit composed of Boolean gates, which allows test generation to be carried out using standard automatic test pattern generation tools. The tests derived using the logic circuit model are then mapped to fluidic operations involving pumps and pressure sensors in the biochip. Feedback from pressure sensors can be compared to expected responses based on the logic circuit model, whereby the types and positions of defects are identified. We show how a fabricated biochip can be tested using the proposed method, and demonstrate experimental results for two additional fabricated chips.
引用
收藏
页码:1463 / 1475
页数:13
相关论文
共 29 条
[1]   An Oscillation-Based Technique for Degradation Monitoring of Sensing and Actuation Electrodes Within Microfluidic Systems [J].
Al-Gayem, Qais ;
Richardson, Andrew ;
Liu, Hongyuan ;
Burd, Nick .
JOURNAL OF ELECTRONIC TESTING-THEORY AND APPLICATIONS, 2011, 27 (03) :375-387
[2]  
[Anonymous], P INT C SYST BIOL AU
[3]  
[Anonymous], P INT TEST C SEPT
[4]  
[Anonymous], STANFORD MICROFLUIDI
[5]  
[Anonymous], CHIPS TIPS LAB ON CH
[6]  
[Anonymous], 2005, INTRO MICROFLUIDICS
[7]   Digital MDA for enumeration of total nucleic acid contamination [J].
Blainey, Paul C. ;
Quake, Stephen R. .
NUCLEIC ACIDS RESEARCH, 2011, 39 (04) :e19
[8]  
Bushnell M., 2000, ESSENTIALS ELECT TES
[9]  
Chakrabarty K., 2006, DIGITAL MICROFLUIDIC
[10]   Rapid prototyping of microfluidic systems in poly(dimethylsiloxane) [J].
Duffy, DC ;
McDonald, JC ;
Schueller, OJA ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4974-4984