Phosphorylcholine surface modified flow diverter associated with reduced intimal hyperplasia

被引:37
作者
Caroff, Jildaz [1 ,2 ]
Tamura, Takamitsu [1 ]
King, Robert M. [1 ]
Lylyk, Pedro N. [1 ]
Langan, Erin T. [1 ]
Brooks, Olivia W. [1 ]
Clarencon, Frederic [3 ]
Wainwright, John Michael [4 ]
Spelle, Laurent [2 ]
Marosfoi, Miklos [1 ]
Gounis, Matthew J. [1 ]
Puri, Ajit S. [1 ]
机构
[1] Univ Massachusetts, Sch Med, Dept Radiol, New England Ctr Stroke Res, Worcester, MA 01655 USA
[2] Bicetre Hosp, Dept Intervent Neuroradiol, NEURI Ctr, Le Kremlin Bicetre, France
[3] Hop La Pitie Salpetriere, Dept Intervent Neuroradiol, Paris, France
[4] Medtron Neurovasc, Res & Dev, Irvine, CA USA
关键词
Intracranial aneurysm; flow diverter; optical coherence tomography; OPTICAL COHERENCE TOMOGRAPHY; INTRACRANIAL ANEURYSMS; STENT; ARTERY; MODEL; COMPLICATIONS; BRANCHES; ASPIRIN; RABBITS; GRAFTS;
D O I
10.1136/neurintsurg-2018-013776
中图分类号
R445 [影像诊断学];
学科分类号
100207 ;
摘要
Background Optical coherence tomography (OCT) is a high-resolution, intra-vascular diagnostic technique widely used for the characterization of vascular pathologies and optimization of stent implantation during percutaneous coronary intervention. OCT was used to investigate the in vivo vascular response to a new phosphorylcholine surface modified flow diverter (sPED). Methods In an in vivo rabbit aneurysmal model, we used two different types of flow diverters (classic Pipeline - cPED; and sPED) with or without dual antiplatelet therapy (four groups, n=10 per group). OCT cross-sectional area measurements were compared with histology in all animals. Neointimal hyperplasia (NIH) ratio was compared between OCT and histology at five different levels for each stent. The severity of NIH was also compared between the different stents, antiplatelet protocols, and vessel locations. Results OCT was used to calculate in-stent hyperplasia in 227 different locations corresponding to histology sections. OCT measurement strongly correlated with gold standard histology (r(2)=0.83; slope=0.988; P<0.0001). sPED had significantly less in-stent NIH than non-treated flow diverters (mean percent of lumen reduction 5.7% for sPED versus 8.9% for cPED; P<0.0001). The NIH ratio was slightly higher with dual antiplatelet therapy (DAPT) (NIH ratio=7.9% with DAPT versus 6.8% without DAPT; P<0.05). Complete and near complete occlusion rates of the aneurysms were not different with the cPED or sPED. Conclusion OCT is a promising technique for immediate and long-term evaluation of flow diverter stent treatments. In an animal model, phosphorylcholine surface modified flow diverters induces less NIH after stent implant without reducing aneurysm occlusion rates.
引用
收藏
页码:1097 / 1101
页数:6
相关论文
共 31 条
[1]   Flow Diverters for Intracranial Aneurysms [J].
Alderazi, Yazan J. ;
Shastri, Darshan ;
Kass-Hout, Tareq ;
Prestigiacomo, Charles J. ;
Gandhi, Chirag D. .
STROKE RESEARCH AND TREATMENT, 2014, 2014
[2]   Creation of saccular aneurysms in the rabbit: A model suitable for testing endovascular devices [J].
Altes, TA ;
Cloft, HJ ;
Short, JG ;
DeGast, A ;
Do, HM ;
Helm, GA ;
Kallmes, DF .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) :349-354
[3]   ELASTASE-INDUCED EXPERIMENTAL ANEURYSMS IN RATS [J].
ANIDJAR, S ;
SALZMANN, JL ;
GENTRIC, D ;
LAGNEAU, P ;
CAMILLERI, JP ;
MICHEL, JB .
CIRCULATION, 1990, 82 (03) :973-981
[4]   Flow diverter implantation in a rat model of sidewall aneurysm: a feasibility study [J].
Aquarius, Rene ;
Smits, Debby ;
Gounis, Matthew J. ;
Leenders, William P. J. ;
de Vries, Joost .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2018, 10 (01) :88-92
[5]   Intracoronary Optical Coherence Tomography: A Comprehensive Review Clinical and Research Applications [J].
Bezerra, Hiram G. ;
Costa, Marco A. ;
Guagliumi, Giulio ;
Rollins, Andrew M. ;
Simon, Daniel I. .
JACC-CARDIOVASCULAR INTERVENTIONS, 2009, 2 (11) :1035-1046
[6]  
Bledsoe SL, 2005, VASCULAR, V13, P43
[7]   Flow-Diverter Stents for the Treatment of Saccular Middle Cerebral Artery Bifurcation Aneurysms [J].
Caroff, J. ;
Neki, H. ;
Mihalea, C. ;
D'Argento, F. ;
Abdel Khalek, H. ;
Ikka, L. ;
Moret, J. ;
Spelle, L. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2016, 37 (02) :279-284
[8]   Phosphorylcholine coating of ePTFE grafts reduces neointimal hyperplasia in canine model [J].
Chen, CY ;
Lumsden, AB ;
Ofenloch, JC ;
Noe, B ;
Campbell, EJ ;
Stratford, PW ;
Yianni, YP ;
Taylor, AS ;
Hanson, SR .
ANNALS OF VASCULAR SURGERY, 1997, 11 (01) :74-79
[9]   Flow Diverters Can Occlude Aneurysms and Preserve Arterial Branches: A New Experimental Model [J].
Darsaut, T. E. ;
Bing, F. ;
Salazkin, I. ;
Gevry, G. ;
Raymond, J. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (10) :2004-2009
[10]   Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia [J].
Flood, Thomas F. ;
van der Bom, Imramsjah M. J. ;
Strittmatter, Lara ;
Puri, Ajit S. ;
Hendricks, Gregory M. ;
Wakhloo, Ajay K. ;
Gounis, Matthew J. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2015, 7 (02) :118-125