Two quality control procedures on radiotherapy beam calibration and treatment planning system implementation

被引:1
|
作者
Piermattei, A
Azario, L
delle Canne, S
Fidanzio, A
Palazzoni, G
机构
[1] Univ Cattolica Sacro Cuore, Ist Fis, I-00168 Rome, Italy
[2] Univ Cattolica Sacro Cuore, Ist Radiol, Div Radioterapia, I-00168 Rome, Italy
关键词
dosimetry; quality control; radiotherapy;
D O I
10.1177/030089169808400210
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
New challenging dosimetric approaches, such as narrow beams and 3D algorithms, are being used in radiotherapy. In this paper two quality control (QC) procedures are reported, The first one concerns the QC of the dosimetry of small x-ray beams, generally carried out by using silicon detectors. The comparison of dose values obtained by a silicon diode, a dia mend detector, and radiochromic films shows that for x-ray beams of high energy, the silicon diode can give an overestimation of the output factors in phantom, up to 4%, This is due to the higher than unit density silicon diode and the surrounding envelope that restore the lateral electron equilibrium. About the 3D algorithms for breast treatment planning, a quality control test has been adopted to verify the accuracy of the computed dosimetry when "loss of scatter" occurs. The results show a sensible agreement (within 1.5%) between computed and experimental data.
引用
收藏
页码:150 / 154
页数:5
相关论文
共 50 条
  • [31] Treatment Planning System Beam Modeling Parameters Exhibit High Variation Among Radiotherapy Institutions
    Glenn, M.
    Followill, D.
    Howell, R.
    Pollard-Larkin, J.
    Zhou, S.
    Kry, S.
    MEDICAL PHYSICS, 2018, 45 (06) : E653 - E653
  • [32] A System for Proton Beam Control during Radiotherapy
    Budyashov, Yu. G.
    Karpunin, V. O.
    Kolonuto, P. E.
    Mitsyn, G. V.
    Molokanov, A. G.
    Shvidky, S. V.
    PHYSICS OF PARTICLES AND NUCLEI LETTERS, 2006, 3 (01) : 59 - 64
  • [33] Quality control system for radiotherapy services
    Allam, Abdelkrim
    Oussalah, Slimane
    Khiter, Belkacem
    Ferhat, Sabrina
    Afiane, M'hammed
    Hocini, Brahim
    2013 10TH IEEE INTERNATIONAL CONFERENCE ON NETWORKING, SENSING AND CONTROL (ICNSC), 2013, : 7 - 10
  • [34] A new methodological approach for PET implementation in radiotherapy treatment planning
    Bellan, Elena
    Ferretti, Alice
    Capirci, Carlo
    Grassetto, Gaia
    Gava, Marcello
    Chondrogiannis, Sotirios
    Virdis, Graziella
    Marzola, Maria Cristina
    Massaro, Arianna
    Rubello, Domenico
    Nibale, Otello
    NUCLEAR MEDICINE COMMUNICATIONS, 2012, 33 (05) : 516 - 520
  • [35] Relative electron density calibration of CT scanners for radiotherapy treatment planning
    Thomas, SJ
    BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (860): : 781 - 786
  • [36] Treatment planning for heavy ion radiotherapy:: clinical implementation and application
    Jäkel, O
    Krämer, M
    Karger, CP
    Debus, J
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (04): : 1101 - 1116
  • [37] Treatment Planning Automation in Prostate Radiotherapy: Overcoming Implementation Challenges
    Sepulveda, Esteban
    Desbiens, Maxime
    Turgeon, Vincent
    Truchon, Dany
    Wakil, Georges
    Hervieux, Yannick
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3715 - S3716
  • [38] Testing and Implementation of Atlas Based Segmentation for Radiotherapy Treatment Planning
    Jameson, M.
    Hellyer, J.
    Choong, C.
    Dinsdale, G.
    MEDICAL PHYSICS, 2018, 45 (06) : E398 - E398
  • [39] Overview of recent advances in treatment planning for ion beam radiotherapy
    Kraemer, Michael
    Scifoni, Emanuele
    Schmitz, Frederike
    Sokol, Olga
    Durante, Marco
    EUROPEAN PHYSICAL JOURNAL D, 2014, 68 (10):
  • [40] Overview of recent advances in treatment planning for ion beam radiotherapy
    Michael Krämer
    Emanuele Scifoni
    Frederike Schmitz
    Olga Sokol
    Marco Durante
    The European Physical Journal D, 2014, 68