Two quality control procedures on radiotherapy beam calibration and treatment planning system implementation

被引:1
|
作者
Piermattei, A
Azario, L
delle Canne, S
Fidanzio, A
Palazzoni, G
机构
[1] Univ Cattolica Sacro Cuore, Ist Fis, I-00168 Rome, Italy
[2] Univ Cattolica Sacro Cuore, Ist Radiol, Div Radioterapia, I-00168 Rome, Italy
关键词
dosimetry; quality control; radiotherapy;
D O I
10.1177/030089169808400210
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
New challenging dosimetric approaches, such as narrow beams and 3D algorithms, are being used in radiotherapy. In this paper two quality control (QC) procedures are reported, The first one concerns the QC of the dosimetry of small x-ray beams, generally carried out by using silicon detectors. The comparison of dose values obtained by a silicon diode, a dia mend detector, and radiochromic films shows that for x-ray beams of high energy, the silicon diode can give an overestimation of the output factors in phantom, up to 4%, This is due to the higher than unit density silicon diode and the surrounding envelope that restore the lateral electron equilibrium. About the 3D algorithms for breast treatment planning, a quality control test has been adopted to verify the accuracy of the computed dosimetry when "loss of scatter" occurs. The results show a sensible agreement (within 1.5%) between computed and experimental data.
引用
收藏
页码:150 / 154
页数:5
相关论文
共 50 条
  • [1] Quality control procedures of a Treatment Planning System
    Milano, F
    Casamassima, F
    Masi, L
    PATRAS MEDICAL PHYSICS 99, 1999, : 9 - 16
  • [2] First Quality Assurance Implementation of Treatment Planning System for Pencil Beam Scanning Proton Radiotherapy
    Xiong, W.
    Yu, G.
    Kang, M.
    Huang, S.
    Lin, H.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [3] Anatomical Information Quality Control of a Radiotherapy Treatment Planning System
    Elaziz R.A.
    Azzam A.A.
    Awad I.A.
    Oraby A.H.
    Iranian Journal of Medical Physics, 2023, 20 (04) : 201 - 206
  • [4] Quality assurance and quality control in operable breast carcinoma external beam radiotherapy treatment planning
    Stankovic, J
    Saric, M
    Borojevic, N
    RadosevicJelic, L
    BALKAN CONGRESS OF ONCOLOGY, 1996, : 177 - 183
  • [5] QUALITY-CONTROL OF RADIOTHERAPY CENTERS IN EUROPE - BEAM CALIBRATION
    DUTREIX, A
    DERREUMAUX, S
    CHAVAUDRA, J
    VANDERSCHUEREN, E
    RADIOTHERAPY AND ONCOLOGY, 1994, 32 (03) : 256 - 264
  • [6] Electron density calibration for radiotherapy treatment planning
    Herrera-Martinez, F.
    Rodriguez-Villafuerte, M.
    Martinez-Davalos, A.
    Ruiz-Trejo, C.
    Celis-Lopez, M. A.
    Larraga-Gutierrez, J. M.
    Garcia-Garduno, A.
    MEDICAL PHYSICS, 2006, 854 : 104 - +
  • [7] Design and implementation of an electronic data recording and processing system for physics quality control checks in external beam radiotherapy
    Patel, I.
    Kirby, M. C.
    BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (950): : 126 - 131
  • [8] Commissioning and quality assurance-of the Pinnacle3 radiotherapy treatment planning system for external beam photons
    Bedford, JL
    Childs, PJ
    Hansen, VN
    Mosleh-Shirazi, MA
    Verhaegen, F
    Warrington, AP
    BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (903): : 163 - 176
  • [9] New automatic quality control methods for geometrical treatment planning system tools in external conformal radiotherapy
    Denis, Eloise
    Beaumont, Stephane
    Guedon, JeanPierre
    Torfeh, Tarraf
    Normand, Nicolas
    Ailleres, Norbert
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [10] The calibration of CT Hounsfield units for radiotherapy treatment planning
    Schneider, U
    Pedroni, E
    Lomax, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01): : 111 - 124