Preclinical Bench Testing on a Novel Posterior Dynamic Deformity Correction Device for Scoliosis

被引:4
作者
Arnin, U. [1 ]
El-Hawary, R. [2 ]
Betz, R. R. [3 ]
Lonner, B. S. [4 ]
Floman, Y. [1 ]
机构
[1] Assuta Hosp, Israel Spine Ctr, Tel Aviv, Israel
[2] IWK Hlth Ctr, Halifax, NS B3K 6R8, Canada
[3] Inst Spine & Scoliosis Lawrenceville, Lawrenceville, NJ USA
[4] Mt Sinai Hosp, New York, NY 10029 USA
关键词
Scoliosis; Fusionless; Posterior dynamic deformity correction; AIS;
D O I
10.1016/j.jspd.2018.08.010
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study DesignBiomechanical test.ObjectiveTo summarize the preclinical tests performed to assess the durability of a novel fusionless dynamic device for the treatment of adolescent idiopathic scoliosis (AIS).Summary of Background DataThe minimal invasive deformity correction (MID-C) system is a distractible posterior dynamic deformity correction device designed to reduce scoliosis for AIS patients, to maintain curve correction, and to preserve spinal motion. To overcome the challenges of wear and fatigue of this procedure, the system has two unique features: polyaxial joints at the rod-screw interface and a ceramic coating of the moving parts.MethodsFive biomechanical tests were performed: Static compression to failure, fatigue loading per ASTM F 1717 with 5.5-mm screws for 10 million cycles (MC) at 5 Hz, wear assessment, wear test of the polyaxial joint under 100 N load for 10 MC, and wear particle implantation in rabbits.ResultsThe system failed through buckling of the rod with loads over 3000 N (400% of human body weight). Dynamically, the system maintained 700 N for 10 MC with 5.5 mm screws. The maximum total steady-state wear rate was 0.074 mg/MC (0.03 per polyaxial joint and 0.014 mg/MC for the ratchet mechanism). Histologic evaluation of the particle injection sites indicated no difference in the local tissue response between the control and test articles. At 3 and 6 months postinjection, there were neither adverse local effects nor systemic effects observed.ConclusionsThe unique design features of the MID-C system, based on polyaxial joints and ceramic coating, resulted in favorable static, fatigue, and wear resistance properties. Wear properties were superior to those published for artificial spinal discs. Long-term outcomes from clinical use will be required to correlate these bench tests to the in vivo reality of clinical use.Level of EvidenceLevel V.
引用
收藏
页码:203 / 212
页数:10
相关论文
共 21 条
  • [1] Appelt K., 2012, THESIS GERMANY U ULM THESIS GERMANY U ULM
  • [2] Appelt K., 2012, BIOMECHANICS THORACI
  • [3] ADOLESCENT IDIOPATHIC SCOLIOSIS - LONG-TERM EFFECT OF INSTRUMENTATION EXTENDING TO THE LUMBAR SPINE
    CONNOLLY, PJ
    VONSCHROEDER, HP
    JOHNSON, GE
    KOSTUIK, JP
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (08) : 1210 - 1216
  • [4] Reoperation after primary posterior instrumentation and fusion for idiopathic scoliosis - Toward defining late operative site pain of unknown cause
    Cook, S
    Asher, M
    Lai, SM
    Shobe, J
    [J]. SPINE, 2000, 25 (04) : 463 - 468
  • [5] Duke H., 1999, SERIES STUDIES HLTH, P109
  • [6] Ellipse Technologies Inc, 2014, PREMARKET NOTIFICATI
  • [7] CLINICAL REVIEW OF PATIENTS WITH BROKEN HARRINGTON RODS
    ERWIN, WD
    DICKSON, JH
    HARRINGTON, PR
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (08) : 1302 - 1307
  • [8] Floman Y, 2018, INT M ADV SPIN TECHN INT M ADV SPIN TECHN
  • [9] Hallab Nadim James, 2009, SAS J, V3, P143, DOI 10.1016/j.esas.2009.11.005
  • [10] A Novel Spinal Implant for Fusionless Scoliosis Correction: A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device
    Holewijn, Roderick M.
    de Kleuver, Marinus
    van der Veen, Albert J.
    Emanuel, Kaj S.
    Bisschop, Arno
    Stadhouder, Agnita
    van Royen, Barend J.
    Kingma, Idsart
    [J]. GLOBAL SPINE JOURNAL, 2017, 7 (05) : 400 - 409