Chemically tapered silver halide fibers: An approach for increasing the sensitivity of mid-infrared evanescent wave sensors

被引:18
作者
Karlowatz, M
Kraft, M
Eitenberger, E
Mizaikoff, B
Katzir, A
机构
[1] Vienna Univ Technol, Inst Analyt Chem, A-1060 Vienna, Austria
[2] Tel Aviv Univ, Sch Phys & Astron, Raymond & Beverly Sackler Fac Exact Sci, IL-69978 Tel Aviv, Israel
关键词
IR fiber-optic sensor; silver halide fibers; chemical tapering; on-line monitoring; chlorinated hydrocarbons;
D O I
10.1366/0003702001948655
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this work an innovative etching technique for tapering silver halide fibers is introduced. As silver halides form soluble complexes with thiosulfate in aqueous solution, the fiber can be chemically tapered by an etching process, which also warrants a high quality of the fiber surface. The evanescent field sensitivity of thus obtained tapered fibers was raised by more than one order of magnitude, demonstrated by calibration curves of tetrachloroethylene in hexane recorded with a tapered sensor fiber coupled to a Fourier transform infrared (FT-IR) spertrometer.
引用
收藏
页码:1629 / 1633
页数:5
相关论文
共 18 条
[1]  
BURGER RJ, 1991, P SOC PHOTO-OPT INS, V1591, P246
[2]  
DREXHAGE MG, 1988, SCI AM, V11, P110
[3]  
HARRINGTON JA, 1990, SPIE, V1228
[4]  
HARRINGTON JA, 1991, SPIE, V1591
[5]  
HARRINGTON JA, 1989, SPIE, V1048
[6]   Optimized sensitive coatings for MIR fiber optic sensors [J].
Jakusch, M ;
Mizaikoff, B ;
Kellner, R ;
Katzir, A .
CHEMICAL, BIOCHEMICAL, AND ENVIRONMENTAL FIBER SENSORS IX, 1997, 3105 :283-288
[7]  
KRSKA R, 1993, APPL PHYS LETT, V63, P1
[8]   Mid-infrared fiberoptic sensors - Potential and perspectives [J].
Mizaikoff, B .
INFRARED OPTICAL FIBERS AND THEIR APPLICATIONS, 1999, 3849 :7-18
[9]   INFRARED FIBEROPTIC CHEMICAL SENSORS WITH REACTIVE SURFACE-COATINGS [J].
MIZAIKOFF, B ;
GOBEL, R ;
KRSKA, R ;
TAGA, K ;
KELLNER, R ;
TACKE, M ;
KATZIR, A .
SENSORS AND ACTUATORS B-CHEMICAL, 1995, 29 (1-3) :58-63
[10]   Vibrational spectra of individual millimeter-size membrane patches using miniature infrared waveguides [J].
Plunkett, SE ;
Jonas, RE ;
Braiman, MS .