Sparsity enhanced spatial resolution and depth localization in diffuse optical tomography

被引:54
作者
Kavuri, Venkaiah C. [1 ]
Lin, Zi-Jing [1 ]
Tian, Fenghua [1 ]
Liu, Hanli [1 ]
机构
[1] Univ Texas Arlington, Joint Grad Program Univ Texas Arlington & Univ Te, Dept Bioengn, Arlington, TX 76019 USA
基金
美国国家卫生研究院;
关键词
STEADY-STATE; REFLECTANCE; QUANTIFICATION; REGULARIZATION; ACTIVATION; DENSITY;
D O I
10.1364/BOE.3.000943
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In diffuse optical tomography (DOT), researchers often face challenges to accurately recover the depth and size of the reconstructed objects. Recent development of the Depth Compensation Algorithm (DCA) solves the depth localization problem, but the reconstructed images commonly exhibit over-smoothed boundaries, leading to fuzzy images with low spatial resolution. While conventional DOT solves a linear inverse model by minimizing least squares errors using L2 norm regularization, L1 regularization promotes sparse solutions. The latter may be used to reduce the over-smoothing effect on reconstructed images. In this study, we combined DCA with L1 regularization, and also with L2 regularization, to examine which combined approach provided us with an improved spatial resolution and depth localization for DOT. Laboratory tissue phantoms were utilized for the measurement with a fiber-based and a camera-based DOT imaging system. The results from both systems showed that L1 regularization clearly outperformed L2 regularization in both spatial resolution and depth localization of DOT. An example of functional brain imaging taken from human in vivo measurements was further obtained to support the conclusion of the study. (c) 2012 Optical Society of America
引用
收藏
页码:943 / 957
页数:15
相关论文
共 42 条
[31]   TIME RESOLVED REFLECTANCE AND TRANSMITTANCE FOR THE NONINVASIVE MEASUREMENT OF TISSUE OPTICAL-PROPERTIES [J].
PATTERSON, MS ;
CHANCE, B ;
WILSON, BC .
APPLIED OPTICS, 1989, 28 (12) :2331-2336
[32]   Spatially variant regularization improves diffuse optical tomography [J].
Pogue, BW ;
McBride, TO ;
Prewitt, J ;
Österberg, UL ;
Paulsen, KD .
APPLIED OPTICS, 1999, 38 (13) :2950-2961
[33]   Instrumentation for fast functional optical tomography [J].
Schmitz, CH ;
Löcker, M ;
Lasker, JM ;
Hielscher, AH ;
Barbour, RL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (02) :429-439
[34]   Automated region detection based on the contrast-to-noise ratio in near-infrared tomography [J].
Song, XM ;
Pogue, BW ;
Jiang, SD ;
Doyley, MM ;
Dehghani, H ;
Tosteson, TD ;
Paulsen, KD .
APPLIED OPTICS, 2004, 43 (05) :1053-1062
[35]   Compressed sensing in diffuse optical tomography [J].
Suezen, Mehmet ;
Giannoula, Alexia ;
Durduran, Turgut .
OPTICS EXPRESS, 2010, 18 (23) :23676-23690
[36]   Algorithmic depth compensation improves quantification and noise suppression in functional diffuse optical tomography [J].
Tian, Fenghua ;
Niu, Haijing ;
Khadka, Sabin ;
Lin, Zi-Jing ;
Liu, Hanli .
BIOMEDICAL OPTICS EXPRESS, 2010, 1 (02) :441-452
[37]   Quantification of functional near infrared spectroscopy to assess cortical reorganization in children with cerebral palsy [J].
Tian, Fenghua ;
Delgado, Mauricio R. ;
Dhamne, Sameer C. ;
Khan, Bilal ;
Alexandrakis, George ;
Romero, Mario I. ;
Smith, Linsley ;
Reid, Dahlia ;
Clegg, Nancy J. ;
Liu, Hanli .
OPTICS EXPRESS, 2010, 18 (25) :25973-25986
[38]   Optimization of probe geometry for diffuse optical brain imaging based on measurement density and distribution [J].
Tian, Fenghua ;
Alexandrakis, George ;
Liu, Hanli .
APPLIED OPTICS, 2009, 48 (13) :2496-2504
[39]  
Wang L. V., 2012, Biomedical Optics: Principles and Imaging
[40]   Retinotopic mapping of adult human visual cortex with high-density diffuse optical tomography [J].
Zeff, Benjamin W. ;
White, Brian R. ;
Dehghani, Hamid ;
Schlaggar, Bradley L. ;
Culver, Joseph P. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2007, 104 (29) :12169-12174