Surface-based registration of liver in ultrasound and CT

被引:0
|
作者
Dehghan, Ehsan [1 ]
Lu, Kongkuo [1 ]
Yan, Pingkun [1 ]
Tahmasebi, Amir [1 ]
Xu, Sheng [2 ,3 ]
Wood, Bradford J. [2 ,3 ]
Abi-Jaoudeh, Nadine [2 ,3 ]
Venkatesan, Aradhana [2 ,3 ]
Kruecker, Jochen [1 ]
机构
[1] Philips Res North Amer, Briarcliff Manor, NY 10510 USA
[2] NIH, Dept Radiol, Bethesda, MD 20892 USA
[3] NIH, Imaging Sci Clin Ctr, Bethesda, MD 20892 USA
关键词
Multi-modality fusion; image registration; liver-diaphragm surface; image-guided intervention; ultrasound; computed tomography; electromagnetic tracking;
D O I
10.1117/12.2082160
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Ultrasound imaging is an attractive modality for real-time image-guided interventions. Fusion of US imaging with a diagnostic imaging modality such as CT shows great potential in minimally invasive applications such as liver biopsy and ablation. However, significantly different representation of liver in US and CT turns this image fusion into a challenging task, in particular if some of the CT scans may be obtained without contrast agents. The liver surface, including the diaphragm immediately adjacent to it, typically appears as a hyper-echoic region in the ultrasound image if the proper imaging window and depth setting are used. The liver surface is also well visualized in both contrast and non-contrast CT scans, thus making the diaphragm or liver surface one of the few attractive common features for registration of US and non-contrast CT. We propose a fusion method based on point-to-volume registration of liver surface segmented in CT to a processed electromagnetically (EM) tracked US volume. In this approach, first, the US image is pre-processed in order to enhance the liver surface features. In addition, non-imaging information from the EM-tracking system is used to initialize and constrain the registration process. We tested our algorithm in comparison with a manually corrected vessel-based registration method using 8 pairs of tracked US and contrast CT volumes. The registration method was able to achieve an average deviation of 12.8mm from the ground truth measured as the root mean square Euclidean distance for control points distributed throughout the US volume. Our results show that if the US image acquisition is optimized for imaging of the diaphragm, high registration success rates are achievable.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Improved Surface-Based Registration of CT and Intraoperative 3D Ultrasound of Bones
    Fanti, Zian
    Torres, Fabian
    Hazan-Lasri, Eric
    Gastelum-Strozzi, Alfonso
    Ruiz-Huerta, Leopoldo
    Caballero-Ruiz, Alberto
    Arambula Cosio, F.
    JOURNAL OF HEALTHCARE ENGINEERING, 2018, 2018
  • [2] Surface-Based CT-TEE Registration of the Aortic Root
    Lang, Pencilla
    Chu, Michael W. A.
    Bainbridge, Dan
    Guiraudon, Gerard M.
    Jones, Douglas L.
    Peters, Terry M.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (12) : 3382 - 3390
  • [3] Surface-based registration of physical space with CT images using A-mode ultrasound localization of the skull
    Bass, WA
    Galloway, RL
    Maurer, CR
    IMAGE DISPLAY - MEDICAL IMAGING 1998, 1998, 3335 : 228 - 238
  • [4] Surface-based registration with a particle filter
    Ma, B
    Ellis, RE
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2004, PT 1, PROCEEDINGS, 2004, 3216 : 566 - 573
  • [5] 3D surface-based registration of ultrasound and histology in prostate cancer imaging
    Schalk, Stefan G.
    Postema, Arnoud
    Saidov, Tamerlan A.
    Demi, Libertario
    Smeenge, Martijn
    de la Rosette, Jean J. M. C. H.
    Wijkstra, Hessel
    Mischi, Massimo
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2016, 47 : 29 - 39
  • [6] Surface-based registration accuracy of CT-based image-guided spine surgery
    Tamura, Y
    Sugano, N
    Sasama, T
    Sato, Y
    Tamura, S
    Yonenobu, K
    Yoshikawa, H
    Ochi, T
    EUROPEAN SPINE JOURNAL, 2005, 14 (03) : 291 - 297
  • [7] Surface-based registration accuracy of CT-based image-guided spine surgery
    Yuichi Tamura
    Nobuhiko Sugano
    Toshihiko Sasama
    Yoshinobu Sato
    Shinichi Tamura
    Kazuo Yonenobu
    Hideki Yoshikawa
    Takahiro Ochi
    European Spine Journal, 2005, 14 : 291 - 297
  • [8] Surface-based prostate registration with biomechanical regularization
    van de Ven, Wendy J. M.
    Hu, Yipeng
    Barentsz, Jelle O.
    Karssemeijer, Nico
    Barratt, Dean
    Huisman, Henkjan J.
    MEDICAL IMAGING 2013: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2013, 8671
  • [9] A surface-based metric for registration error quantification
    Nanayakkara, Nuwan D.
    Chiu, Bernard
    Fenster, Aaron
    2009 INTERNATIONAL CONFERENCE ON INDUSTRIAL AND INFORMATION SYSTEMS, 2009, : 349 - +
  • [10] Surface-based preoperative CT/MRI to intraoperative face scan registration: a clinical study
    Shamir, R. R.
    Freiman, M.
    Joskowicz, L.
    Zehavi, E.
    Spektor, S.
    Shoshan, Y.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2007, 2 : S208 - S210