A study on beams passing through hip prosthesis for pelvic radiation treatment

被引:64
作者
Ding, GX [1 ]
Yu, CW [1 ]
机构
[1] British Columbia Canc Agcy, Fraser Valley Canc Ctr, Clin Phys Dept, Dept Med Phys, Surrey, BC V3V 1Z2, Canada
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2001年 / 51卷 / 04期
关键词
pelvic radiotherapy; hip prostheses; dose at interface; treatment planning; inhomogeneity corrections;
D O I
10.1016/S0360-3016(01)02592-5
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To study the dose distributions at the interface due to the presence of a metal implant; to show the dose distributions in combined fields in the presence of hip prostheses; and to demonstrate the capabilities and limitations of a conventional system. Methods and Materials: Perturbations in the dose distribution caused by a hip prosthesis can result in unacceptable dose inhomogeneities within the target volume and in regions where tissues interface with implant. The Monte Carlo technique and a conventional treatment planning system are used to calculate the dose distributions. Results: Dose increases of 15% in tissue are seen at the interface between metal implant and tissue. Dose reductions of 5-25% or 10-45% are observed in the shadow of the hip prosthesis made of 0.5-3-cm-thick titanium or steel alloy respectively. We compared predicted dose distribution between the Monte Carlo simulation and a commercial treatment planning system (CADPLAN). We found that CADPLAN underestimated the attenuation of hip prostheses. This has led to overestimation of the target dose by 14% for a typical four-field box technique. Conclusions: An acceptable dose distribution can be achieved with a proper lateral beam weighting and compensation using an eight-field technique. The beam compensation may be applied to achieve an adequate target dose. (C) 2001 Elsevier Science Inc.
引用
收藏
页码:1167 / 1175
页数:9
相关论文
共 12 条
  • [1] Alecu R, 1999, Med Dosim, V24, P33, DOI 10.1016/S0958-3947(98)00044-2
  • [2] EFFECT OF A FEMORAL-HEAD PROSTHESIS ON MEGAVOLTAGE BEAM RADIOTHERAPY
    BIGGS, PJ
    RUSSELL, MD
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (03): : 581 - 586
  • [3] Carolan M, 2000, Australas Radiol, V44, P290, DOI 10.1046/j.1440-1673.2000.00816.x
  • [4] Ding G, 1995, PIRS, V439
  • [5] Evaluation of a commercial three-dimensional electron beam treatment planning system
    Ding, GX
    Cygler, JE
    Zhang, GG
    Yu, MK
    [J]. MEDICAL PHYSICS, 1999, 26 (12) : 2571 - 2580
  • [6] HIP PROSTHESES DURING PELVIC IRRADIATION - EFFECTS AND CORRECTIONS
    HAZUKA, MB
    IBBOTT, GS
    KINZIE, JJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (06): : 1311 - 1317
  • [7] JOHNS HE, 1983, PHYSICS RADIOLOGY
  • [8] BEAM - A MONTE-CARLO CODE TO SIMULATE RADIOTHERAPY TREATMENT UNITS
    ROGERS, DWO
    FADDEGON, BA
    DING, GX
    MA, CM
    WE, J
    MACKIE, TR
    [J]. MEDICAL PHYSICS, 1995, 22 (05) : 503 - 524
  • [9] Schild S E, 1992, Med Dosim, V17, P83
  • [10] Comparison of measured and Monte Carlo calculated dose distributions from the NRC linac
    Sheikh-Bagheri, D
    Rogers, DWO
    Ross, CK
    Seuntjens, JP
    [J]. MEDICAL PHYSICS, 2000, 27 (10) : 2256 - 2266