Salvaging Pull-Out Strength in a Previously Stripped Screw Site: A Comparison of Three Rescue Techniques

被引:5
作者
Addevico, Francesco [1 ]
Solitro, Giovanni F. [2 ]
Morandi, Massimo Max [2 ]
机构
[1] Osped Niguarda Ca Granda, Dept Orthoped & Traumatol, I-20162 Milan, Italy
[2] Louisiana State Univ Hlth Shreveport, Dept Orthopaed Surg, Shreveport, LA 71103 USA
关键词
screw; stripping; rescue; thread; CORTICAL BONE; FIXATION; INSERTION; FRACTURE; TORQUE;
D O I
10.3390/jfmk6030071
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Screw stripping during bone fixation is a common occurrence during operations that results in decreased holding capacity and bone healing. We aimed to evaluate the rescue of the stripped screw site using screws of different dimensions. Five screw configurations were tested on cadaveric specimens for pull-out strength (POS). The configurations included a control screw tightened without stripping, a configuration voluntarily stripped and left in place, and three more configurations in which the stripped screws were replaced by a different screw with either increased overall length, diameter, or thread length. Each configuration was tested five times, with each screw tested once. The POS of the control screw, measured to be 153.6 +/- 27 N, was higher than the POS measured after stripping and leaving the screw in place (57.1 +/- 18 N, p = 0.001). The replacement of the stripped screw resulted in a POS of 158.4 +/- 64 N for the screw of larger diameter, while the screws of the same diameter but increased length or those with extended thread length yielded POS values of 138.4 +/- 42 and 185.7 +/- 48 N, respectively. Screw stripping is a frequent intraoperative complication that, according to our findings, cannot be addressed by leaving the screw in place. The holding capacity of a stripped screw implanted in cancellous bone can successfully be restored with a different screw of either larger diameter, longer length, or extended thread length.
引用
收藏
页数:8
相关论文
共 31 条
[1]   Screw insertion torque as parameter to judge the fixation. Assessment of torque and pull-out strength in different bone densities and screw-pitches [J].
Addevico, Francesco ;
Morandi , Massimo ;
Scaglione, Michelangelo ;
Solitro, Giovanni F. .
CLINICAL BIOMECHANICS, 2020, 72 :130-135
[2]   Subchondral bone microarchitecture analysis in the proximal tibia at 7-T MRI [J].
Agten, Christoph A. ;
Honig, Stephen ;
Saha, Punam K. ;
Regatte, Ravinder ;
Chang, Gregory .
ACTA RADIOLOGICA, 2018, 59 (06) :716-722
[3]   The proximal tibia metaphysis: A reliable donor site for bone grafting? [J].
Alt, V ;
Meeder, PJ ;
Seligson, D ;
Schad, A ;
Atienza, C .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (414) :315-321
[4]   Sacroiliac screw fixation: A mini review of surgical technique [J].
Alvis-Miranda, Hernando Raphael ;
Farid-Escorcia, Hector ;
Alcala-Cerra, Gabriel ;
Castellar-Leones, Sandra Milena ;
Moscote-Salazar, Luis Rafael .
JOURNAL OF CRANIOVERTEBRAL JUNCTION AND SPINE, 2014, 5 (03) :110-113
[5]   Use of a synthetic bone void filler to augment screws in osteopenic ankle fracture fixation [J].
Andreassen, GS ;
Hoiness, PR ;
Skraamm, I ;
Granlund, O ;
Engebretsen, L .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2004, 124 (03) :161-165
[6]   Mechanical Metrics of the Proximal Tibia are Precise and Differentiate Osteoarthritic and Normal Knees: A Finite Element Study [J].
Arjmand, Hanieh ;
Nazemi, Majid ;
Kontulainen, Saija A. ;
McLennan, Christine E. ;
Hunter, David J. ;
Wilson, David R. ;
Johnston, James D. .
SCIENTIFIC REPORTS, 2018, 8
[7]  
Au Brigham, 2017, J Orthop Trauma, V31, P210, DOI 10.1097/BOT.0000000000000769
[8]   Biomechanical measurements of stopping and stripping torques during screw insertion in five types of human and artificial humeri [J].
Aziz, Mina S. R. ;
Tsuji, Matthew R. S. ;
Nicayenzi, Bruce ;
Crookshank, Meghan C. ;
Bougherara, Habiba ;
Schemitsch, Emil H. ;
Zdero, Radovan .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (05) :446-455
[9]  
Bastian JD, 2012, ACTA ORTHOP BELG, V78, P100
[10]   Factors affecting the pullout strength of cancellous bone screws [J].
Chapman, JR ;
Harrington, RM ;
Lee, KM ;
Anderson, PA ;
Tencer, AF ;
Kowalski, D .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :391-398